Biomechanical Comparison of Conventional Plate and the C-Nail® System for the Treatment of Displaced Intra-Articular Calcaneal Fractures: A Finite Element Analysis

被引:5
作者
Pinzaru, Roxana Maria [1 ]
Paval, Silviu Dumitru [2 ]
Pertea, Mihaela [3 ]
Alexa, Ovidiu [1 ]
Sirbu, Paul Dan [1 ]
Filip, Alexandru [1 ]
Carp, Adrian Claudiu [1 ]
Savin, Liliana [1 ]
Forna, Norin [1 ]
Veliceasa, Bogdan [1 ]
机构
[1] Grigore T Popa Univ Med & Pharm, Fac Med, Dept Orthopaed & Traumatol Surg Sci 2, 16 Univ St, Iasi 700115, Romania
[2] Gheorghe Asachi Tech Univ, Dept Comp Sci & Engn, 27 Dimitrie Mangeron, Iasi 700050, Romania
[3] Grigore T Popa Univ Med & Pharm, Fac Med, Dept Plast Surg & Reconstruct Microsurg Surg Sci 1, 16, Univ St, Iasi 700115, Romania
来源
JOURNAL OF PERSONALIZED MEDICINE | 2023年 / 13卷 / 04期
关键词
calcaneal fracture; finite element analysis; biomechanics; interlocking nail; C-Nail; locking plate; INTERNAL-FIXATION; OPEN REDUCTION; LOCKING NAIL;
D O I
10.3390/jpm13040587
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
The C-Nail((R)) system is a novel intramedullary fixation method for displaced intra-articular calcaneal fractures. The aim of this study was to evaluate the biomechanical performance of the C-Nail((R)) system and compare it with conventional plate fixation for the treatment of displaced intra-articular calcaneal fractures using finite element analysis. The geometry of a Sanders type-IIB fracture was constructed using the computer-aided design software Ansys SpaceClaim. The C-Nail((R)) system (Medin, Nove Mesto n. Morave, Czech Republic) and the calcaneal locking plate (Auxein Inc., 35 Doral, Florida) and screws were designed according to the manufacturer specifications. Vertical loading of 350 N and 700 N were applied to the subtalar joint surfaces to simulate partial weight bearing and full weight bearing. Construct stiffness, total deformation, and von Mises stress were assessed. The maximum stress on the C-Nail((R)) system was lower compared with the plate (110 MPa vs. 360 MPa). At the bone level the stress was found to have higher values in the case of the plate compared to the C-Nail((R)) system. The study suggests that the C-Nail((R)) system can provide sufficient stability, making it a viable option for the treatment of displaced intra-articular calcaneal fractures.
引用
收藏
页数:12
相关论文
共 32 条
  • [1] Allegra Paul R, 2020, Foot Ankle Orthop, V5, p2473011420927334, DOI 10.1177/2473011420927334
  • [2] Wound infections following open reduction and internal fixation of calcaneal fractures with an extended lateral approach
    Backes, Manouk
    Schepers, Tim
    Beerekamp, M. Suzan H.
    Luitse, Jan S. K.
    Goslings, J. Carel
    Schep, Niels W. L.
    [J]. INTERNATIONAL ORTHOPAEDICS, 2014, 38 (04) : 767 - 773
  • [3] Use of intramedullary locking nail for displaced intraarticular fractures of the calcaneus: what is the evidence?
    Bernasconi, Alessio
    Iorio, Paolino
    Ghani, Yaser
    Argyropoulos, Miltiadis
    Patel, Shelain
    Barg, Alexej
    Smeraglia, Francesco
    Balato, Giovanni
    Welck, Matthew
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (08) : 1911 - 1922
  • [4] Biomechanical Evaluation of a Locking and Nonlocking Reconstruction Plate in an Osteoporotic Calcaneal Fracture Model
    Blake, Matthew H.
    Owen, John R.
    Sanford, Timothy S.
    Wayne, Jennifer S.
    Adelaar, Robert S.
    [J]. FOOT & ANKLE INTERNATIONAL, 2011, 32 (04) : 432 - 436
  • [5] Intra-Articular Calcaneal Fractures: Comparison between Mini-Invasive Approach and Kirschner Wires vs. Extensive Approach and Dedicated Plate-A Retrospective Evaluation at Long-Term Follow-Up
    Caravelli, Silvio
    Gardini, Giammarco
    Pungetti, Camilla
    Gentile, Paolo
    Perisano, Carlo
    Greco, Tommaso
    Rinaldi, Vito Gaetano
    Muccioli, Giulio Maria Marcheggiani
    Tigani, Domenico
    Mosca, Massimiliano
    [J]. JOURNAL OF CLINICAL MEDICINE, 2023, 12 (01)
  • [6] Predictive Factors of Poor Outcome in Sanders Type III and IV Calcaneal Fractures Treated with an Open Reduction and Internal Fixation with Plate: A Medium-Term Follow-Up
    Cianni, Luigi
    Vitiello, Raffaele
    Greco, Tommaso
    Sirgiovanni, Mattia
    Ragonesi, Giulia
    Maccauro, Giulio
    Perisano, Carlo
    [J]. JOURNAL OF CLINICAL MEDICINE, 2022, 11 (19)
  • [7] Coughlin M.J.S.C.L.A.R.B., 2014, Manns surgery of the foot and ankle
  • [8] Surgical treatment of displaced intraarticular calcaneal fractures by a minimally invasive technique using a locking nail: A preliminary study
    Fascione, Fabrizio
    Di Mauro, Michele
    Guelfi, Matteo
    Malagelada, Francesc
    Pantalone, Andrea
    Salini, Vincenzo
    [J]. FOOT AND ANKLE SURGERY, 2019, 25 (05) : 679 - 683
  • [9] Locked nailing for the treatment of displaced articular fractures of the calcaneus: description of a new procedure with calcanail®
    Goldzak, Mario
    Mittlmeier, Thomas
    Simon, Patrick
    [J]. EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2012, 22 (04) : 345 - 349
  • [10] Operative Versus Nonoperative Treatment for Displaced Intra-Articular Calcaneal Fractures: A Meta-Analysis of Randomized Controlled Trials
    Luo, Xiangping
    Li, Qi
    He, Shengmao
    He, Shunqing
    [J]. JOURNAL OF FOOT & ANKLE SURGERY, 2016, 55 (04) : 821 - 828