Quantification of Mechanical Characteristics of Conventional Steerable Ablation Catheters for Treatment of Atrial Fibrillation Using a Heart Phantom

被引:2
作者
Miller, Jacob [1 ,2 ]
Sewani, Alykhan [1 ,2 ]
Rezazada, Jeffrey [1 ,2 ]
Alawneh, Yara [1 ,2 ]
Kazemian, Pedram [3 ]
Terricabras, Maria [2 ]
Wright, Graham [2 ]
Tavallaei, M. Ali [1 ,2 ]
机构
[1] Toronto Metropolitan Univ, Toronto, ON, Canada
[2] Sunnybrook Res Inst, Toronto, ON, Canada
[3] Deborah Heart & Lung Ctr, Browns Mills, NJ USA
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Catheter ablation; Pulmonary vein isolation (PVI); Catheter orientation; Catheter; Contact force; Orientation; PULMONARY VEIN ISOLATION; CONTACT FORCE; NAVIGATION; ACCURACY;
D O I
10.1007/s13239-023-00662-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeAccurate and reliable catheter navigation is important in formation of adequate lesions during radiofrequency cardiac catheter ablation. To inform future device design efforts and to characterize the limitations of conventional devices, the focus of this study is to assess and quantify the mechanical performance of manual ablation catheters for pulmonary vein isolation procedures within a phantom heart model.MethodsWe measured three important metrics: accuracy of catheter tip navigation to target anatomical landmarks at the pulmonary veins (PVs), orientation of the catheter relative to the tissue at the targets, and the delivered force values and their stability and variations at those targets. A stereovision system was used for navigational guidance and to measure the catheter's tip position and orientation relative to the targets. To measure force, piezoelectric sensors were used which were integrated at the targets, whereby operators were instructed to stabilize the catheter to achieve a chosen reference force value.ResultsAn overall positioning accuracy of 1.57 +/- 1.71 mm was achieved for all targets. No statistical significance was observed in position accuracy between the right and left PVs (p = 0.5138). The orientation of the catheter relative to tissue surface was 41 degrees +/- 21 degrees with no statistical significance between targets. The overall force stability was 41 +/- 6 g with higher difficulty in force stabilization in the right compared to the left PV (40 +/- 8 vs. 43 +/- 2 g, p < 0.0001).ConclusionBased on our results, future improvements to manual catheter navigation for ablation should focus on improving device performance in orientation control and improved force stability.
引用
收藏
页码:419 / 427
页数:9
相关论文
共 28 条
[1]  
[Anonymous], VIEW EXPLORE IMAGES
[2]  
[Anonymous], TRIANGULATION 2 D 3
[3]   A Percutaneous Catheter forIn VivoHyperspectral Imaging of Cardiac Tissue: Challenges, Solutions and Future Directions [J].
Armstrong, Kenneth ;
Larson, Cinnamon ;
Asfour, Huda ;
Ransbury, Terry ;
Sarvazyan, Narine .
CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2020, 11 (05) :560-575
[4]  
Bergman R., 2010, Cardiovasc Eng Technol, V1, P4
[5]   2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Patient Selection, Procedural Techniques, Patient Management and Follow-up, Definitions, Endpoints, and Research Trial Design [J].
Calkins, Hugh ;
Kuck, Karl Heinz ;
Cappato, Riccardo ;
Brugada, Josep ;
Camm, A. John ;
Chen, Shih-Ann ;
Crijns, Harry J. G. ;
Damiano, Ralph J., Jr. ;
Davies, D. Wyn ;
DiMarco, John ;
Edgerton, James ;
Ellenbogen, Kenneth ;
Ezekowitz, Michael D. ;
Haines, David E. ;
Haissaguerre, Michel ;
Hindricks, Gerhard ;
Iesaka, Yoshito ;
Jackman, Warren ;
Jalife, Jose ;
Jais, Pierre ;
Kalman, Jonathan ;
Keane, David ;
Kim, Young-Hoon ;
Kirchhof, Paulus ;
Klein, George ;
Kottkamp, Hans ;
Kumagai, Koichiro ;
Lindsay, Bruce D. ;
Mansour, Moussa ;
Marchlinski, Francis E. ;
McCarthy, Patrick M. ;
Mont, J. Lluis ;
Morady, Fred ;
Nademanee, Koonlawee ;
Nakagawa, Hiroshi ;
Natale, Andrea ;
Nattel, Stanley ;
Packer, Douglas L. ;
Pappone, Carlo ;
Prystowsky, Eric ;
Raviele, Antonio ;
Reddy, Vivek ;
Ruskin, Jeremy N. ;
Shemin, Richard J. ;
Tsao, Hsuan-Ming ;
Wilber, David ;
Ad, Niv ;
Cummings, Jennifer ;
Gillinov, A. Mark ;
Heidbuchel, Hein .
EUROPACE, 2012, 14 (04) :528-606
[6]   Ablation catheter orientation: In vitro effects on lesion size and in vivo analysis during PVI for atrial fibrillation [J].
Calzolari, Vittorio ;
De Mattia, Luca ;
Basso, Francesco ;
Crosato, Martino ;
Scalon, Alessandra ;
Squasi, Paolo Antonio Maria ;
Del Favero, Simone ;
Cernetti, Carlo .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2020, 43 (12) :1554-1563
[7]   A Review of Catheter Related Complications During Minimally Invasive Transcatheter Cardiovascular Intervention with Implications for Catheter Design [J].
Dellimore, Kiran H. ;
Franklin, Steve E. ;
Helyer, Annika R. .
CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2014, 5 (03) :217-232
[8]   The durability of pulmonary vein isolation using the visually guided laser balloon catheter: Multicenter results of pulmonary vein remapping studies [J].
Dukkipati, Srinivas R. ;
Neuzil, Petr ;
Kautzner, Josef ;
Petru, Jan ;
Wichterle, Dan ;
Skoda, Jan ;
Cihak, Robert ;
Peichl, Petr ;
Dello Russo, Antonio ;
Pelargonio, Gemma ;
Tondo, Claudio ;
Natale, Andrea ;
Reddy, Vivek Y. .
HEART RHYTHM, 2012, 9 (06) :919-925
[9]   Determinants of Acute and Late Pulmonary Vein Reconnection in Contact Force-Guided Pulmonary Vein Isolation Identifying the Weakest Link in the Ablation Chain [J].
El Haddad, Milad ;
Taghji, Philippe ;
Phlips, Thomas ;
Wolf, Michael ;
Demolder, Anthony ;
Choudhury, Rajin ;
Knecht, Sebastien ;
Vandekerckhove, Yves ;
Tavernier, Rene ;
Nakagawa, Hiroshi ;
Duytschaever, Mattias .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (04)
[10]   Stand-Alone Pulmonary Vein Isolation Versus Pulmonary Vein Isolation With Additional Substrate Modification as Index Ablation Procedures in Patients With Persistent and Long-Standing Persistent Atrial Fibrillation The Randomized Alster-Lost-AF Trial (Ablation at St. Georg Hospital for Long-Standing Persistent Atrial Fibrillation) [J].
Fink, Thomas ;
Schuelter, Michael ;
Heeger, Christian-Hendrik ;
Lemes, Christine ;
Maurer, Tilman ;
Reissmann, Bruno ;
Riedl, Johannes ;
Rottner, Laura ;
Santoro, Francesco ;
Schmidt, Boris ;
Wohlmuth, Peter ;
Mathew, Shibu ;
Sohns, Christian ;
Ouyang, Feifan ;
Metzner, Andreas ;
Kuck, Karl-Heinz .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (07)