Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch

被引:1
作者
Lapic, Ivana [1 ]
Rade, Anamarija [2 ]
Kraljevic, Andela [3 ]
Milos, Marija [4 ]
Herak, Desiree Coen [3 ]
Daskijevic, Lucija [3 ]
Cerovac, Paula [3 ]
Rogic, Dunja [3 ,5 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Kispaticeva 12, Zagreb 10000, Croatia
[2] Gen Hosp Varazdin, Lab Med Biochem, Varazhdin, Croatia
[3] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Zagreb, Croatia
[4] Univ Mostar, Fac Pharm, Mostar, Bosnia & Herceg
[5] Univ Zagreb, Fac Pharm & Biochem, Zagreb, Croatia
关键词
automation; erythrocyte sedimentation rate; validation study; Westergren method; SYSTEM;
D O I
10.1515/cclm-2023-0033
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Analytical validation of automated erythrocyte sedimentation rate (ESR) analyzers is necessary prior to their implementation into routine practice. Our aim was to perform the analytical validation of the modified Westergren method applied on the CUBE 30 touch analyzer (Diesse, Siena, Italy). Methods: Validation included determination of within-run and between-run precision following the Clinical and Laboratory Standards Institute EP15-A3 protocol, comparison with the reference Westergren method, sample stability assessment at both room temperature and 4 degrees C, after 4, 8 and 24-h storage, and checking the extent of hemolysis and lipemia interference. Results: Coefficients of variation (CVs) for within-run precision were 5.2% for the normal and 2.6% for the abnormal range, while between-run CVs were 9.4 and 2.2%, respectively. Comparison with the Westergren method (n=191) yielded Spearman's correlation coefficient of 0.93, no constant nor proportional difference [y=0.4 (95% CI: -1.7-1.0) + 1.06 (95% CI: 1.00-1.14)x] and a non-significant mean absolute bias of -2.6 mm (95% CI: -5.3-0.2). Lower comparability was evidenced with increasing ESR values, with both constant and proportional differences for ESR values between 40 and 80 mm, and above 80 mm. Sample stability was not compromised up to 8-h storage both at room temperature (p=0.054) and 4 degrees C (p=0.421). Hemolysis did not affect ESR measurement up to 1.0 g/L of free hemoglobin (p=0.089), while lipemia index above 5.0 g/L affects the ESR result (p=0.004). Conclusions: This study proved that CUBE 30 touch provides reliable ESR measurement and satisfactory comparability with the reference Westergren methods, with minor variation related to methodological differences.
引用
收藏
页码:1463 / 1469
页数:7
相关论文
共 28 条
  • [21] Evaluation of Ves-Matic Cube 200-an automated system for the measurement of the erythrocyte sedimentation rate
    Perovic, E.
    Bakovic, L.
    Valcic, A.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2010, 32 (01) : 88 - 94
  • [22] Erythrocyte Sedimentation Rate Measurements by TEST 1 Better Reflect Inflammation Than Do Those by the Westergren Method in Patients With Malignancy, Autoimmune Disease, or Infection
    Cha, Choong-Hwan
    Park, Chan-Jeoung
    Cha, Young Joo
    Kim, Hyun Kyung
    Kim, Duck Hee
    Honghoon
    Bae, Jae Hoon
    Jung, Jae-Seol
    Jang, Seongsoo
    Chi, Hyun-Sook
    Lee, Dong Soon
    Cho, Han-Ik
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 131 (02) : 189 - 194
  • [23] Erythrocyte sedimentation rate measurements using MIX-RATE® X20 and VISION A automated analyzers: Method validation and comparison study
    Prompetchara, Eakachai
    Nowaratsopon, Sunudda
    Wongkamchai, Suppakorn
    Srieakpanit, Jittra
    Ketloy, Chutitorn
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (05) : 846 - 853
  • [24] The comparison of erythrocyte sedimentation rate (ESR) modify Westergren Caretium Xc-A30 and Westergren Manual in Clinical Pathology Laboratory, Sanglah General Hospital, Denpasar, Bali
    Dewi, Made Minarti Witarini
    Herawati, Sianny
    Mulyantari, Ni Kadek
    Prabawa, I. Putu Yuda
    BALI MEDICAL JOURNAL, 2019, 8 (02) : 396 - 399
  • [25] Accuracy study of a novel alternate method measuring erythrocyte sedimentation rate for prototype hematology analyzer Celltac α
    Maki, Ryosuke
    Tabe, Yoko
    Yamamoto, Takamasa
    Takemura, Hiroyuki
    Sawada, Tomohiro
    Esaki, Tatsuhiro
    Higuchi, Makoto
    Misawa, Shigeki
    Miyake, Kazunori
    Miida, Takashi
    Nagai, Yutaka
    Ohsaka, Akimichi
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (04) : 588 - 596
  • [26] Erythrocyte Sedimentation Rate (ESR) Measured by the Streck ESR-Auto Plus Is Higher Than With the Sediplast Westergren Method
    Vennapusa, Bharathi
    De La Cruz, Laura
    Shah, Hemangini
    Michalski, Veronica
    Zhang, Qian-Yun
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2011, 135 (03) : 386 - 390
  • [27] The TEST 1 automated system - A new method for measuring the erythrocyte sedimentation rate
    Plebani, M
    De Toni, S
    Sanzari, MC
    Bernardi, D
    Stockreiter, E
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1998, 110 (03) : 334 - 340
  • [28] Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice
    Cennamo, Michele
    Giuliano, Loredana
    Arrigoni, Gloria
    Fardone, Valentina
    Russo, Roberta
    De Tomasi, Luca Maria
    Bertani, Fabio
    Cammarota, Gaetano
    Brunetti, Giovanni
    Del Vecchio, Lucia
    Partenope, Michelarcangelo
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (03)