Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch

被引:2
作者
Lapic, Ivana [1 ]
Rade, Anamarija [2 ]
Kraljevic, Andela [3 ]
Milos, Marija [4 ]
Herak, Desiree Coen [3 ]
Daskijevic, Lucija [3 ]
Cerovac, Paula [3 ]
Rogic, Dunja [3 ,5 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Kispaticeva 12, Zagreb 10000, Croatia
[2] Gen Hosp Varazdin, Lab Med Biochem, Varazhdin, Croatia
[3] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Zagreb, Croatia
[4] Univ Mostar, Fac Pharm, Mostar, Bosnia & Herceg
[5] Univ Zagreb, Fac Pharm & Biochem, Zagreb, Croatia
关键词
automation; erythrocyte sedimentation rate; validation study; Westergren method; SYSTEM;
D O I
10.1515/cclm-2023-0033
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Analytical validation of automated erythrocyte sedimentation rate (ESR) analyzers is necessary prior to their implementation into routine practice. Our aim was to perform the analytical validation of the modified Westergren method applied on the CUBE 30 touch analyzer (Diesse, Siena, Italy). Methods: Validation included determination of within-run and between-run precision following the Clinical and Laboratory Standards Institute EP15-A3 protocol, comparison with the reference Westergren method, sample stability assessment at both room temperature and 4 degrees C, after 4, 8 and 24-h storage, and checking the extent of hemolysis and lipemia interference. Results: Coefficients of variation (CVs) for within-run precision were 5.2% for the normal and 2.6% for the abnormal range, while between-run CVs were 9.4 and 2.2%, respectively. Comparison with the Westergren method (n=191) yielded Spearman's correlation coefficient of 0.93, no constant nor proportional difference [y=0.4 (95% CI: -1.7-1.0) + 1.06 (95% CI: 1.00-1.14)x] and a non-significant mean absolute bias of -2.6 mm (95% CI: -5.3-0.2). Lower comparability was evidenced with increasing ESR values, with both constant and proportional differences for ESR values between 40 and 80 mm, and above 80 mm. Sample stability was not compromised up to 8-h storage both at room temperature (p=0.054) and 4 degrees C (p=0.421). Hemolysis did not affect ESR measurement up to 1.0 g/L of free hemoglobin (p=0.089), while lipemia index above 5.0 g/L affects the ESR result (p=0.004). Conclusions: This study proved that CUBE 30 touch provides reliable ESR measurement and satisfactory comparability with the reference Westergren methods, with minor variation related to methodological differences.
引用
收藏
页码:1463 / 1469
页数:7
相关论文
共 18 条
[1]   Diagnosis and Management of Rheumatoid Arthritis A Review [J].
Aletaha, Daniel ;
Smolen, Josef S. .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2018, 320 (13) :1360-1372
[2]   General Assembly, Diagnosis, Laboratory Test: Proceedings of International Consensus on Orthopedic Infections [J].
Barrack, Robert ;
Bhimani, Samrath ;
Blevins, Jason L. ;
Blevins, Kier ;
Demetres, Michelle ;
Figgie, Mark ;
Fillingham, Yale ;
Goodman, Susan ;
Huddleston, James ;
Kahlenberg, Cynthia ;
Lautenbach, Charles ;
Lin, Jianhao ;
Lonner, Jess ;
Lynch, Maureen ;
Malkani, Arthur ;
Martin, Lily ;
Mirza, Serene ;
Najjad, Muhammad Kazim Rahim ;
Penna, Sreeram ;
Richardson, Shawn ;
Sculco, Peter ;
Shahi, AliSina ;
Szymonifka, Jackie ;
Wang, Qiaojie .
JOURNAL OF ARTHROPLASTY, 2019, 34 (02) :S187-S195
[3]   Comparison of iSED and Ves-Matic Cube 200 Erythrocyte Sedimentation Rate Measurements With Westergren Method [J].
Bogdaycioglu, Nihal ;
Yilmaz, Fatma Meric ;
Sezer, Sevilay ;
Oguz, Esra .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2015, 29 (05) :397-404
[4]   Evaluation of the TEST 1 erythrocyte sedimentation rate system and intra- and inter-laboratory quality control using new latex control materials [J].
Cha, Choong-Hwan ;
Cha, Young Joo ;
Park, Chan-Jeoung ;
Kim, Hyun Kyung ;
Cha, Eun-Jong ;
Kim, Duck Hee ;
Honghoon ;
Jung, Jae-Seol ;
Kim, Mi-Jung ;
Jang, Seongsoo ;
Chi, Hyun-Sook ;
Lee, Dong Soon .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (07) :1043-1048
[5]   Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate [J].
Cicak, Helena ;
Sonjic, Pavica ;
Simundic, Ana-Maria .
BIOCHEMIA MEDICA, 2022, 32 (01) :98-105
[6]  
CLSI, 2014, EP15A3 CLSI
[7]   Hodgkin lymphoma: ESMO Clinical Practice Guidelines for diagnosis, treatment and follow-up [J].
Eichenauer, D. A. ;
Aleman, B. M. P. ;
Andre, M. ;
Federico, M. ;
Hutchings, M. ;
Illidge, T. ;
Engert, A. ;
Ladetto, M. .
ANNALS OF ONCOLOGY, 2018, 29 :19-29
[8]   The Top 25 Laboratory Tests by Volume and Revenue in Five Different Countries [J].
Horton, Susan ;
Fleming, Kenneth A. ;
Kuti, Modupe ;
Looi, Lai-Meng ;
Pai, Sanjay A. ;
Sayed, Shahin ;
Wilson, Michael L. .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2019, 151 (05) :446-451
[9]  
Hrvatska komora medicinskih biokemiara, 2007, HARM LAB NAL PODR OP
[10]   ICSH recommendations for modified and alternate methods measuring the erythrocyte sedimentation rate [J].
Kratz, A. ;
Plebani, M. ;
Peng, M. ;
Lee, Y. K. ;
McCafferty, R. ;
Machin, S. J. .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (05) :448-457