Intraobserver Repeatability Assessment of the S390L Firefly WDR Slitlamp in Patients With Dry Eye Disease: Objective, Automated, and Noninvasive Measures

被引:9
作者
Ballesteros-Sanchez, Antonio [1 ,2 ]
Gargallo-Martinez, Beatriz [2 ,3 ]
Gutierrez-Ortega, Ramon [2 ,3 ]
Sanchez-Gonzalez, Jose-Maria [1 ]
机构
[1] Univ Seville, Dept Phys Condensed Matter, Opt Area, Reina Mercedes St, Seville, Spain
[2] Clin Novovis, Dept Ophthalmol, Murcia, Spain
[3] Univ Murcia, Dept Ophthalmol Optometry Otorhinolaryngol & Anat, Murcia, Spain
来源
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE | 2023年 / 49卷 / 07期
关键词
Automatic objective measures; Dry eye diagnostic system; Intraobserver repeatability; Meibomian gland dysfunction; OCULAR SURFACE; TEAR FILM; INTERNATIONAL WORKSHOP; AMBIENT LEVELS; DIAGNOSIS; SYMPTOMS; HORMONES; SIGNS; TIME; SEX;
D O I
10.1097/ICL.0000000000001001
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objectives:To assess the intraobserver repeatability of automated, objective, and noninvasive measures obtained with the S390L Firefly WDR slitlamp.Methods:This cross-sectional study included 50 eyes of patients with dry eye disease with a mean age of 55.06 & PLUSMN;12.96 years. Three consecutively repeated measures of the following variables were obtained: first noninvasive break-up time (F-NIBUT), average noninvasive break-up time (A-NIBUT), tear meniscus height, tear meniscus area (TMA), nasal ciliary hyperemia (NCIH), temporal ciliary hyperemia (TCIH), nasal conjunctival hyperemia (NCOH), temporal conjunctival hyperemia (TCOH), upper loss area meibomian gland (U-LAMG), lower loss area meibomian gland (L-LAMG), upper meibomian gland dysfunction grade (U-MGD grade), and lower meibomian gland dysfunction grade (L-MGD grade). Intraobserver repeatability was estimated with coefficient of variation (CoV), intrasubject standard deviation (SD) (S-w), and Bland-Altman plots.Results:All variables showed no statistically significant differences in the repeated-measures analysis except for L-MGD grade (P=0.045). F-NIBUT and A-NIBUT obtained the highest CoV with an average value of 0.48 & PLUSMN;0.41 [0.02-1.00] and 0.34 & PLUSMN;0.25 [0.02-1.00], respectively. The remaining variables showed CoVs between 0.04 & PLUSMN;0.11 [0.00-0.43] and 0.18 & PLUSMN;0.16 [0.00-0.75]. A-NIBUT, TMA, NCOH, and L-LAMG obtained an S-w of 2.78s, 0.21 mm(2), <0.001, and 4.11%, respectively. Bland-Altman plots showed a high level of agreement between pairs of repeated measures.Conclusion:The S390L Firefly WDR slitlamp has moderate intraobserver repeatability for F-NIBUT and A-NIBUT, which suggests that F-NIBUT and A-NIBUT are tests with high variability. The remaining variables show satisfactory intraobserver repeatability.
引用
收藏
页码:283 / 291
页数:9
相关论文
共 50 条
[1]   Effect of Ambient Temperature on the Human Tear Film [J].
Abusharha, Ali A. ;
Pearce, E. Ian ;
Fagehi, Raied .
EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2016, 42 (05) :308-312
[2]   The Effect of Low Humidity on the Human Tear Film [J].
Abusharha, Ali A. ;
Pearce, E. Ian .
CORNEA, 2013, 32 (04) :429-434
[3]   Automated Measurement of Tear Meniscus Height with the Kowa DR-1α Tear Interferometer in Both Healthy Subjects and Dry Eye Patients [J].
Arita, Reiko ;
Yabusaki, Katsumi ;
Hirono, Taisuke ;
Yamauchi, Takanori ;
Ichihashi, Tadashi ;
Fukuoka, Shima ;
Morishige, Naoyuki .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (06) :2092-2101
[4]   Meibography: A Japanese Perspective [J].
Arita, Reiko .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (14) :DES48-DES55
[5]   Agreement and repeatability of four different devices to measure noninvasive tear breakup time (NIBUT) [J].
Bandlitz, Stefan ;
Peter, Barbara ;
Pflugi, Tanja ;
Jaeger, Kai ;
Anwar, Aaisha ;
Bikhu, Paramdeep ;
Nosch, Daniela S. ;
Wolffsohn, James S. .
CONTACT LENS & ANTERIOR EYE, 2020, 43 (05) :507-511
[6]   The relationship between habitual patient-reported symptoms and clinical signs among patients with dry eye of varying severity [J].
Begley, CG ;
Chalmers, RL ;
Abetz, L ;
Venkataraman, K ;
Mertzanis, P ;
Caffery, BA ;
Snyder, C ;
Edrington, T ;
Nelson, D ;
Simpson, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (11) :4753-4761
[7]   TFOS DEWS II pathophysiology report [J].
Bron, Anthony J. ;
de Paiva, Cintia S. ;
Chauhan, Sunil K. ;
Bonini, Stefano ;
Gabison, Eric E. ;
Jain, Sandeep ;
Knop, Erich ;
Markoulli, Maria ;
Ogawa, Yoko ;
Perez, Victor ;
Uchino, Yuichi ;
Yokoi, Norihiko ;
Zoukhri, Driss ;
Sullivan, David A. .
OCULAR SURFACE, 2017, 15 (03) :438-510
[8]   Repeatability of tear meniscus evaluation using spectral-domain Cirrus® HD-OCT and time-domain Visante® OCT [J].
Chan, Hiok Hong ;
Zhao, Yang ;
Tun, Tin A. ;
Tong, Louis .
CONTACT LENS & ANTERIOR EYE, 2015, 38 (05) :368-372
[9]   TFOS DEWS II Definition and Classification Report [J].
Craig, Jennifer P. ;
Nichols, Kelly K. ;
Akpek, Esen K. ;
Caffery, Barbara ;
Dua, Harminder S. ;
Joo, Choun-Ki ;
Liu, Zuguo ;
Nelson, J. Daniel ;
Nichols, Jason J. ;
Tsubota, Kazuo ;
Stapleton, Fiona .
OCULAR SURFACE, 2017, 15 (03) :276-283
[10]   Rasch Analysis of the Ocular Surface Disease Index (OSDI) [J].
Dougherty, Bradley E. ;
Nichols, Jason J. ;
Nichols, Kelly K. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (12) :8630-8635