摘要
Objective: Electrically evoked auditory steady-state responses (EASSRs) can potentially be used as an objective measure to realize the automatic fitting of cochlear implants (CIs). They can be recorded using electroencephalography (EEG) and objectively detected at the modulation frequency of the stimulus. The main roadblock in using EASSRs is the presence of CI stimulation artifacts in the EEG recording. In this article, we present an improvement of a recently introduced system identification (SI) based artifact removal method. We evaluate its applicability for objective CI fitting on a larger dataset. Methods: The parameter estimation problem of the SI is solved using ordinary least squares (OLS), where an additional regularization term is added to the cost function. We compare EASSR latencies as determined by the commonly used linear interpolation artifact removal method and SI, to evaluate the artifact removal and EASSR detection quality on a dataset of 16 CI recipients and four different stimulation levels. Results: SI can fully remove stimulation artifacts and detect EASSRs, even for recordings from ipsilateral EEG channels, where all other artifact removal methods fail so far. Using OLS with regularization prevents false positive response detection. Conclusion: Using SI, EASSRs can reliably be detected in EEG recordings, even for ipsilateral recording channels and recordings with lower stimulation levels. As the recordings are obtained with clinically relevant settings of the CI, they reveal the potential impact of SI on the objective fitting of CIs. Significance: We argue, that SI enables therefore a big step towards automated CI fitting with EASSRs.
引用
收藏
页码:738 / 749
页数:12
相关论文
共 50 条
  • [1] Electrically evoked auditory steady state response detection in cochlear implant recipients using a system identification approach
    Schott, Julian
    Gransier, Robin
    Wouters, Jan
    Moonen, Marc
    2022 30TH EUROPEAN SIGNAL PROCESSING CONFERENCE (EUSIPCO 2022), 2022, : 1243 - 1247
  • [2] Characterization of cochlear implant artifacts in electrically evoked auditory steady-state responses
    Deprez, Hanne
    Gransier, Robin
    Hofmann, Michael
    van Wieringen, Astrid
    Wouters, Jan
    Moonen, Marc
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2017, 31 : 127 - 138
  • [3] Electrically Evoked Auditory Steady State Responses in Cochlear Implant Users
    Michael Hofmann
    Jan Wouters
    Journal of the Association for Research in Otolaryngology, 2010, 11 : 267 - 282
  • [4] Electrically Evoked Auditory Steady State Responses in Cochlear Implant Users
    Hofmann, Michael
    Wouters, Jan
    JARO-JOURNAL OF THE ASSOCIATION FOR RESEARCH IN OTOLARYNGOLOGY, 2010, 11 (02): : 267 - 282
  • [5] COCHLEAR IMPLANT ARTIFACT REJECTION IN ELECTRICALLY EVOKED AUDITORY STEADY STATE RESPONSES
    Deprez, Hanne
    Hofmann, Michael
    van Wieringen, Astrid
    Wouters, Jan
    Moonen, Marc
    2014 PROCEEDINGS OF THE 22ND EUROPEAN SIGNAL PROCESSING CONFERENCE (EUSIPCO), 2014, : 1995 - 1999
  • [6] Electrically-evoked auditory steady-state responses as neural correlates of loudness growth in cochlear implant users
    Van Eeckhoutte, Maaike
    Wouters, Jan
    Francart, Tom
    HEARING RESEARCH, 2018, 358 : 22 - 29
  • [7] Electrically evoked auditory steady-state responses in guinea pigs
    Jeng, Fuh-Cherng
    Abbas, Paul J.
    Brown, Carolyn J.
    Miller, Charles A.
    Nourski, Kirill V.
    Robinson, Barbara K.
    AUDIOLOGY AND NEURO-OTOLOGY, 2007, 12 (02) : 101 - 112
  • [8] Electrically Evoked Auditory Late Responses as a Novel Method for Cochlear Implant Programming
    Deniz, Risvan
    Kara, Eyyup
    Deniz, Burcu
    Kara, Halide Cetin
    Yener, Haydar Murat
    Atas, Ahmet
    OTOLOGY & NEUROTOLOGY, 2022, 43 (04) : E414 - E420
  • [9] Electrically evoked auditory brainstem responses in cochlear implant patients
    Kim, LS
    Kang, MK
    Park, HS
    Kim, SJ
    Heo, SD
    COCHLEAR IMPLANT AND RELATED SCIENCES UPDATE, 1997, 52 : 92 - 95
  • [10] Electrically evoked auditory brainstem responses in cochlear implant patients
    Ciprut, Ayca
    Akdas, Ferda
    MEDITERRANEAN JOURNAL OF OTOLOGY, 2007, 3 (01): : 6 - 11