Sensitivity and specificity of Varian Halcyon's portal dosimetry for plan-specific pre-treatment QA

被引:2
作者
Razinskas, Gary [1 ]
Schindhelm, Robert [1 ]
Sauer, Otto A. [1 ]
Wegener, Sonja [1 ]
机构
[1] Univ Hosp Wurzburg, Dept Radiat Oncol, Wurzburg, Germany
关键词
error detection sensitivity and specificity; plan-specific quality assurance; portal dosimetry; STATISTICAL PROCESS-CONTROL; VOLUMETRIC MODULATED ARC; QUALITY-ASSURANCE; DOSE VERIFICATION; EPID DOSIMETRY; IMRT; SYSTEM; THERAPY; FILM;
D O I
10.1002/acm2.14001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeDeveloped as a plan-specific pre-treatment QA tool, Varian portal dosimetry promises a fast, high-resolution, and integrated QA solution. In this study, the agreement between predicted fluence and measured cumulative portal dose was determined for the first 140 patient plans at our Halcyon linear accelerator. Furthermore, the capability of portal dosimetry to detect incorrect plan delivery was compared to that of a common QA phantom. Finally, tolerance criteria for verification of VMAT plan delivery with Varian portal dosimetry were derived. MethodsAll patient plans and the corresponding verification plans were generated within the Eclipse treatment planning system. Four representative plans of different treatment sites (prostate, prostate with lymphatic drainage, rectum, and head & neck) were intentionally altered to model incorrect plan delivery. Investigated errors included both systematic and random errors. Gamma analysis was conducted on both portal dose (criteria gamma(2%/2 mm), gamma(2%/1 mm), and gamma(1%/1 mm)) and ArcCHECK measurements (criteria gamma(3%/3 mm), gamma(3%/2 mm), and gamma(2%/2 mm)) with a 10% low-dose threshold. Performance assessment of various acceptance criteria for plan-specific treatment QA utilized receiver operating characteristic (ROC) analysis. ResultsPredicted and acquired portal dosimetry fluences demonstrated a high agreement evident by average gamma passing rates for the clinical patient plans of 99.90%, 96.64%, and 91.87% for gamma(2%/2 mm), gamma(2%/1 mm), and gamma(1%/1 mm), respectively. The ROC analysis demonstrated a very high capability of detecting erroneous plan delivery for portal dosimetry (area under curve (AUC) > 0.98) and in this regard outperforms QA with the ArcCHECK phantom (AUC approximate to 0.82). With the suggested optimum decision thresholds excellent sensitivity and specificity is simultaneously possible. ConclusionsOwing to the high achievable spatial resolution, portal dosimetry at the Halcyon can reliably be deployed as plan-specific pre-treatment QA tool to screen for errors. It is recommended to support the fluence integrated portal dosimetry QA by independent phantom-based measurements of a random sample survey of treatment plans.
引用
收藏
页数:11
相关论文
共 43 条
[1]   EPID dosimetry for pretreatment quality assurance with two commercial systems [J].
Bailey, Daniel W. ;
Kumaraswamy, Lalith ;
Bakhtiari, Mohammad ;
Malhotra, Harish K. ;
Podgorsak, Matthew B. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (04) :82-99
[2]   Using an EPID for patient-specific VMAT quality assurance [J].
Bakhtiari, M. ;
Kumaraswamy, L. ;
Bailey, D. W. ;
de Boer, S. ;
Malhotra, H. K. ;
Podgorsak, M. B. .
MEDICAL PHYSICS, 2011, 38 (03) :1366-1373
[3]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[4]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[5]   ROC analysis in patient specific quality assurance [J].
Carlone, Marco ;
Cruje, Charmainne ;
Rangel, Alejandra ;
McCabe, Ryan ;
Nielsen, Michelle ;
MacPherson, Miller .
MEDICAL PHYSICS, 2013, 40 (04)
[6]   Validation and IMRT/VMAT delivery quality of a preconfigured fast-rotating O-ring linac system [J].
De Roover, Robin ;
Crijns, Wouter ;
Poels, Kenneth ;
Michiels, Steven ;
Nulens, An ;
Vanstraelen, Bianca ;
Petillion, Saskia ;
De Brabandere, Marisol ;
Haustermans, Karin ;
Depuydt, Tom .
MEDICAL PHYSICS, 2019, 46 (01) :328-339
[7]   Assessment of a commercial EPID dosimetry system to detect radiotherapy treatment errors [J].
Doolan, Paul ;
Nikolaou, Maria ;
Ferentinos, Konstantinos ;
Anagnostopoulos, Georgios .
BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (04)
[8]   Dosimetric impact and detectability of multi-leaf collimator positioning errors on Varian Halcyon [J].
Gay, Skylar S. ;
Netherton, Tucker J. ;
Cardenas, Carlos E. ;
Ger, Rachel B. ;
Balter, Peter A. ;
Dong, Lei ;
Mihailidis, Dimitris ;
Court, Laurence E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (08) :47-55
[9]   A comprehensive analysis of the IMRT dose delivery process using statistical process control (SPC) [J].
Gerard, Karine ;
Grandhaye, Jean-Pierre ;
Marchesi, Vincent ;
Kafrouni, Hanna ;
Husson, Francois ;
Aletti, Pierre .
MEDICAL PHYSICS, 2009, 36 (04) :1275-1285
[10]   Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy [J].
Greer, PB ;
Popescu, CC .
MEDICAL PHYSICS, 2003, 30 (07) :1618-1627