Development of sweep frequency ultrasonic interferometer for high precision velocity measurement in liquids

被引:0
作者
Dhiman, Nitin [1 ,2 ]
Sharma, Sahil [1 ,2 ]
Piyush [1 ,2 ]
Kumar, Bishan [1 ,2 ]
Yadav, Sanjay [1 ,2 ]
Dubey, P. K. [1 ,2 ]
机构
[1] CSIR Natl Phys Lab, Div Physico Mech Metrol, Pressure Vacuum & Ultrason Metrol, Dr K S Krishnan Marg, New Delhi 110012, India
[2] Acad Sci & Innovat Res AcSIR, Ghaziabad 201002, India
关键词
SOUND-VELOCITY; ATTENUATION; ACCURATE;
D O I
10.1063/5.0151763
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An ultrasonic interferometer with variable separation between the transducer and reflector is widely used for the measurement of ultrasonic propagation velocity in liquids. The inherent limitation of such an interferometer is due to the mechanical movement of its reflector for ultrasonic wavelength measurement in a liquid medium. It is observed that the ultrasonic velocity measurement precision is adversely affected at higher frequencies compared to lower ones. For instance, in our experimentation, a standard deviation of +/- 21.5 m/s (+/- 1.43%) was obtained for velocity measurement at 1.84 MHz with the consideration of two consecutive maxima, which increases drastically to +/- 76.8 m/s (+/- 5.12%) at 9.4 MHz. These measurements can significantly be improved by considering many maxima and averaging for wavelength estimation. However, it still requires design attention and improvement, particularly for higher frequencies. In this article, a sweep-frequency based ultrasonic interferometer design with a fixed separation for liquid characterization is proposed and described. This technique overcomes the limitations of mechanical movement systems and also provides a better and uniform precision for lower as well as higher frequencies. The functionality of the developed sweep frequency method was tested in water, carbon tetrachloride, ethylene glycol, and glycerol, which shows good agreement with literature values. The velocity measurement in double distilled water by the developed technique at 1 Hz sweep resolution has shown an improved standard deviation of +/- 0.74 m/s (+/- 0.05%) at 9.4 MHz.
引用
收藏
页数:10
相关论文
共 19 条
[1]   Comparison of theoretical and experimental ultrasonic velocity in three-component mixtures of methyl ethyl ketone and toluene with n-alkanols and estimation of acoustic nonlinearity parameter (B/A) [J].
Abraham, R ;
Khadar, MA ;
Jugan, J .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1999, 106 (03) :1284-1288
[2]  
[Anonymous], 2006, IEC Standard (61161)
[3]   Ultrasonic interferometric sensor for rheological changes of fluids [J].
Balasubramaniam, Krishnan ;
Sethuraman, S. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2006, 77 (08)
[4]   Sound attenuation, shear viscosity, and mutual diffusivity behavior in the nitroethane-cyclohexane critical mixture [J].
Behrends, R ;
Iwanowski, I ;
Kosmowska, M ;
Szala, A ;
Kaatze, U .
JOURNAL OF CHEMICAL PHYSICS, 2004, 121 (12) :5929-5934
[5]   Ultrasonic interferometry for liquid media [J].
Fox, FE .
PHYSICAL REVIEW, 1937, 52 (09) :0973-0981
[6]   ACCURATE MEASUREMENTS OF THE SOUND-VELOCITY IN PURE WATER BY COMBINING A COHERENT PHASE-DETECTION TECHNIQUE AND A VARIABLE PATH-LENGTH INTERFEROMETER [J].
FUJII, K ;
MASUI, R .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1993, 93 (01) :276-282
[7]   A digital, constant-frequency pulsed phase-locked-loop instrument for real-time, absolute ultrasonic phase measurements [J].
Haldren, H. A. ;
Perey, D. F. ;
Yost, W. T. ;
Cramer, K. E. ;
Gupta, M. C. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2018, 89 (05)
[8]   Swept-frequency ultrasonic phase evaluation of adhesive bonding in tri-layer structures [J].
Haldren, Harold A. ;
Perey, Daniel F. ;
Yost, William T. ;
Cramer, K. Elliott ;
Gupta, Mool C. .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2019, 145 (03) :1609-1618
[9]   Ultrasonic velocity measurements in liquids with high resolution - techniques, selected applications and perspectives [J].
Kaatze, Udo ;
Eggers, Frieder ;
Lautscham, Karl .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2008, 19 (06)
[10]   Contact and non-contact ultrasonic measurement in the food industry: a review [J].
Khairi, Mohd Taufiq Mohd ;
Ibrahim, Sallehuddin ;
Yunus, Mohd Amri Md ;
Faramarzi, Mahdi .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2016, 27 (01)