共 12 条
Conventional versus reverse insertion of i-gel® in overweight and obese patients - Interventional randomised controlled trial
被引:2
作者:
Ahuja, Sonali
[1
]
Kaur, Gurpreeti
[1
]
Garg, Kamakshi
[1
,2
]
Grewal, Anju
[1
]
机构:
[1] DMCH Ludhiana, Dept Anaesthesiol, Ludhiana, Punjab, India
[2] H 4,Barewal Rd, Ludhiana 141012, Punjab, India
关键词:
Airway management;
i-gel;
obese;
oropharyngeal leak pressure;
overweight;
supraglottic airway device;
LARYNGEAL MASK AIRWAY;
I-GEL;
SUPRAGLOTTIC AIRWAY;
PLACEMENT;
STANDARD;
D O I:
10.4103/ija.ija_749_22
中图分类号:
R614 [麻醉学];
学科分类号:
100217 ;
摘要:
Background and Aims:The supraglottic airway device, i-gel, is used in obese patients for short- to medium-duration surgical procedures. Insertion techniques have contributed to the successful and proper placement of i-gel in the first attempt. This study aims to compare two techniques for successfully inserting i-gel in the first attempt in overweight and obese patients as measured by oropharyngeal leak pressure (OLP).Methods:This interventional, randomised, controlled study was conducted after ethical approval, and trial registration in overweight and obese patients. Patients were randomised into two groups: In Group C, the conventional technique was used, while in Group R, the reverse technique was used to insert i-gel. OLP, successful placement, required manipulations, time taken for insertion, number of attempts, and intraoperative and postoperative complications were studied. The collected data were analysed statistically.Results:The mean OLP (30.46 & PLUSMN; 3.76 vs. 32.12 & PLUSMN; 3.10 mmHg, P = 0.018) and the mean time of insertion (16.42 & PLUSMN; 1.86 vs. 13.98 & PLUSMN; 1.97 s, P = 0.001) for conventional and reverse techniques, respectively, were statistically significant and favourable for Group R compared to Group C. Successful placement of i-gel at the first attempt, ease of insertion, number of attempts and all the manipulations except withdrawal and advancement were comparable in both the groups. No postoperative complications were noted.Conclusion:The reverse technique significantly favoured the actual OLP values and the mean insertion time. Successful placement of i-gel at the first attempt was observed with both conventional and reverse techniques.
引用
收藏
页码:708 / 713
页数:6