The value of using a deep learning image reconstruction algorithm of thinner slice thickness to balance the image noise and spatial resolution in low-dose abdominal CT

被引:3
作者
Wang, Huan [1 ]
Li, Xinyu [2 ]
Wang, Tianze
Li, Jianying [2 ,3 ]
Sun, Tianze [1 ]
Chen, Lihong [1 ]
Cheng, Yannan [1 ]
Jia, Xiaoqian [1 ]
Niu, Xinyi [1 ]
Guo, Jianxin [1 ]
机构
[1] Xi An Jiao Tong Univ, Affiliated Hosp 1, Dept Radiol, Xian, Peoples R China
[2] Xi An Jiao Tong Univ, Sch Med, Dept Neurosurg, Xian, Peoples R China
[3] Computed Tomog Res Ctr, GE Healthcare, Beijing, Peoples R China
关键词
Deep learning; image reconstruction; different layer thickness; radiation dose; ITERATIVE RECONSTRUCTION; QUALITY; ANGIOGRAPHY; REDUCTION;
D O I
10.21037/qims-22-353
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Traditional reconstruction techniques have certain limitations in balancing image quality and reducing radiation dose. The deep learning image reconstruction (DLIR) algorithm opens the door to a new era of medical image reconstruction. The purpose of the study was to evaluate the DLIR images at 1.25 mm thickness in balancing image noise and spatial resolution in low-dose abdominal computed tomography (CT) in comparison with the conventional adaptive statistical iterative reconstruction-V at 40% strength (ASIR-V40%) at 5 and 1.25 mm. Methods: This retrospective study included 89 patients who underwent low-dose abdominal CT. Five sets of images were generated using ASIR-V40% at a 5 mm slice thickness and 1.25 mm (high-resolution) with DLIR at 1.25 mm using 3 strengths: low (DLIR-L), medium (DLIR-M), and high (DLIR-H). Qualitative evaluation was performed for image noise, artifacts, and visualization of small structures, while quantitative evaluation was performed for standard deviation (SD), signal-to-noise ratio (SNR), and spatial resolution (defined as the edge rising slope). Results: At 1.25 mm, DLIR-M and DLIR-H images had significantly lower noise (SD in fat: 14.29 +/- 3.37 and 9.65 +/- 3.44 HU, respectively), higher SNR for liver (3.70 +/- 0.78 and 5.64 +/- 1.20, respectively), and higher overall image quality (4.30 +/- 0.44 and 4.67 +/- 0.40, respectively) than did the respective values in ASIR-V40% images (20.60 +/- 4.04 HU, 2.60 +/- 0.63, and 3.77 +/- 0.43; all P values <0.05). Compared with the 5 mm ASIR-V40% images, the 1.25 mm DLIR-H images had lower noise (SD: 9.65 +/- 3.44 vs. 13.63 +/- 10.03 HU), higher SNR (5.64 +/- 1.20 vs. 4.69 +/- 1.28), and higher overall image quality scores (4.67 +/- 0.40 vs. 3.94 +/- 0.46) (all P values <0.001). In addition, DLIR-L, DLIR-M, and DLIR-H images had a significantly higher spatial resolution in terms of edge rising slope (59.66 +/- 21.46, 58.52 +/- 17.48, and 59.26 +/- 13.33, respectively, vs. 33.79 +/- 9.23) and significantly higher image quality scores in the visualization of fine structures (4.43 +/- 0.50, 4.41 +/- 0.49, and 4.38 +/- 0.49, respectively vs. 2.62 +/- 0.49) than did the 5 mm ASIR-V40 images. Conclusions: The 1.25 mm DLIR-M and DLIR-H images had significantly reduced image noise and improved SNR and overall image quality compared to the 1.25 mm ASIR-V40% images, and theyhad significantly improved the spatial resolution , visualization of fine structures compared to the 5 mm ASIR-V40% images. DLIR-H images had further reduced image noise compared with the 5 mm ASIR-V40% images , DLIR-H was the most effective technique at balancing the image noise and spatial resolution in low-dose abdominal CT.
引用
收藏
页码:1814 / 1824
页数:11
相关论文
共 27 条
  • [1] Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT
    Akagi, Motonori
    Nakamura, Yuko
    Higaki, Toru
    Narita, Keigo
    Honda, Yukiko
    Zhou, Jian
    Yu, Zhou
    Akino, Naruomi
    Awai, Kazuo
    [J]. EUROPEAN RADIOLOGY, 2019, 29 (11) : 6163 - 6171
  • [2] Validation of deep-learning image reconstruction for coronary computed tomography angiography: Impact on noise, image quality and diagnostic accuracy
    Benz, Dominik C.
    Benetos, Georgios
    Rampidis, Georgios
    von Felten, Elia
    Bakula, Adam
    Sustar, Aleksandra
    Kudura, Ken
    Messerli, Michael
    Fuchs, Tobias A.
    Gebhard, Catherine
    Pazhenkottil, Aju P.
    Kaufmann, Philipp A.
    Buechel, Ronny R.
    [J]. JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (05) : 444 - 451
  • [3] A study of using a deep learning image reconstruction to improve the image quality of extremely low-dose contrast-enhanced abdominal CT for patients with hepatic lesions
    Cao, Le
    Liu, Xiang
    Li, Jianying
    Qu, Tingting
    Chen, Lihong
    Cheng, Yannan
    Hu, Jieliang
    Sun, Jingtao
    Guo, Jianxin
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1118)
  • [4] CT Radiation Dose: What Can You Do Right Now in Your Practice?
    Coakley, Fergus V.
    Gould, Robert
    Yeh, Benjamin M.
    Arenson, Ronald L.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (03) : 619 - 625
  • [5] Low-dose CT of the abdomen: Initial experience on a novel photon-counting detector CT and comparison with energy-integrating detector CT
    Decker, Josua A.
    Bette, Stefanie
    Lubina, Nora
    Rippel, Katharina
    Braun, Franziska
    Risch, Franka
    Wollny, Claudia
    Scheurig-Muenkler, Christian
    Kroencke, Thomas J.
    Schwarz, Florian
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2022, 148
  • [6] Adaptive Statistical Iterative Reconstruction-V Versus Adaptive Statistical Iterative Reconstruction: Impact on Dose Reduction and Image Quality in Body Computed Tomography
    Gatti, Marco
    Marchisio, Filippo
    Fronda, Marco
    Rampado, Osvaldo
    Faletti, Riccardo
    Bergamasco, Laura
    Ropolo, Roberto
    Fonio, Paolo
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (02) : 191 - 196
  • [7] Deep Learning Reconstruction at CT: Phantom Study of the Image Characteristics
    Higaki, Toru
    Nakamura, Yuko
    Zhou, Jian
    Yu, Zhou
    Nemoto, Takuya
    Tatsugami, Fuminari
    Awai, Kazuo
    [J]. ACADEMIC RADIOLOGY, 2020, 27 (01) : 82 - 87
  • [8] Association of Exposure to Diagnostic Low-Dose Ionizing Radiation With Risk of Cancer Among Youths in South Korea
    Hong, Jae-Young
    Han, Kyungdo
    Jung, Jin-Hyung
    Kim, Jung Sun
    [J]. JAMA NETWORK OPEN, 2019, 2 (09)
  • [9] Incremental Image Noise Reduction in Coronary CT Angiography Using a Deep Learning-Based Technique with Iterative Reconstruction
    Hong, Jung Hee
    Park, Eun-Ah
    Lee, Whal
    Ahn, Chulkyun
    Kim, Jong-Hyo
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2020, 21 (10) : 1165 - 1177
  • [10] Image Quality Assessment of Abdominal CT by Use of New Deep Learning Image Reconstruction: Initial Experience
    Jensen, Corey T.
    Liu, Xinming
    Tamm, Eric P.
    Chandler, Adam G.
    Sun, Jia
    Morani, Ajaykumar C.
    Javadi, Sanaz
    Wagner-Bartak, Nicolaus A.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2020, 215 (01) : 50 - 57