Volumetric absorptive microsampling-LC-MS/MS assays for quantitation of giredestrant in dried human whole blood

被引:3
|
作者
Johnson, Ryan [1 ]
Rehmani, Imran [2 ]
Leahy, Mark [2 ]
Walker, Abigail [2 ]
Liang, Xiaorong [1 ]
Dean, Brian [1 ]
Wang, Jianshuang [1 ]
机构
[1] Genentech Inc, Dept Drug Metab & Pharmacokinet, San Francisco, CA 94080 USA
[2] Labcorp Early Drug Dev, Dept Bioanalyt Chem, Madison, WI USA
关键词
anticoagulant; hematocrit; LC-MS/MS; microsampling; Mitra; Tasso-M20; validation; VAMS; whole blood; QUANTIFICATION; HEMATOCRIT; SPOTS; VALIDATION;
D O I
10.4155/bio-2022-0189
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Volumetric absorption microsampling devices offer minimally invasive and user-friendly collection of capillary blood in volumes as low as 10 mu l. Herein we describe the assay validation for determination of the selective estrogen receptor degrader giredestrant (GDC-9545) in dried human whole blood collected using the Mitra((R)) and Tasso-M20 devices. Both LC-MS/MS assays met validation acceptance criteria for the linear range 1-1000 ng/ml giredestrant. Mitra and Tasso-M20 samples were stable for 84 and 28 days at ambient conditions, respectively, and for 7-9 days at 40 and -70 degrees C. Blood hematocrit, hyperlipidemia and anticoagulant did not impact quantitation of giredestrant. These validated assays are suitable for the determination of giredestrant in dried blood samples collected using Mitra and Tasso-M20 microsampling devices. This methodology publication explores the bioanalytical considerations of the assays used to measure giredestrant in human whole blood samples collected using the Mitra and Tasso-M20 volumetric absorption microsampling devices. #Microsampling
引用
收藏
页码:1377 / 1389
页数:14
相关论文
共 50 条
  • [1] Volumetric absorptive microsampling coupled with hybridization LC-MS/MS for quantitation of antisense oligonucleotides
    Chen, Ming-Luan
    Mekhssian, Kevork
    Dutt, Muskaan
    Plomley, Jeff
    Keyhani, Anahita
    BIOANALYSIS, 2023, : 1115 - 1128
  • [2] Bioanalysis of emixustat (ACU-4429) in whole blood collected with volumetric absorptive microsampling by LC-MS/MS
    Miao, Zhixin
    Farnham, James G.
    Hanson, Glenn
    Podoll, Terry
    Reid, Michael J.
    BIOANALYSIS, 2015, 7 (16) : 2071 - 2083
  • [3] A volumetric absorptive microsampling LC-MS/MS method for five immunosuppressants and their hematocrit effects
    Koster, Remco A.
    Niemeijer, Pascal
    Veenhof, Herman
    van Hateren, Kai
    Alffenaar, Jan-Willem C.
    Touw, Daan J.
    BIOANALYSIS, 2019, 11 (06) : 495 - 508
  • [4] Development of LC-MS/MS method for quantification of Lurasidone using volumetric absorptive microsampling (VAMS); a comparative study between dried blood and plasma samples
    Rajadhyaksha, Madhura
    Londhe, Vaishali
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2024, 1244
  • [5] Volumetric absorptive MicroSampling vs. other blood sampling materials in LC-MS-based protein analysis - preliminary investigations
    Andersen, Ida Kristine Lysgaard
    Rosting, Cecilie
    Gjelstad, Astrid
    Halvorsen, Trine Gronhaug
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 156 : 239 - 246
  • [6] Blood and Plasma Volumetric Absorptive Microsampling (VAMS) Coupled to LC-MS/MS for the Forensic Assessment of Cocaine Consumption
    Mandrioli, Roberto
    Mercolini, Laura
    Protti, Michele
    MOLECULES, 2020, 25 (05):
  • [7] LC-MS/MS measurement of endogenous steroid hormones and phase II metabolites in blood volumetric absorptive microsampling (VAMS) for doping control purposes
    Ponzetto, Federico
    Parasiliti-Caprino, Mirko
    Leoni, Laura
    Marinelli, Lorenzo
    Nonnato, Antonello
    Nicoli, Raul
    Kuuranne, Tiia
    Ghigo, Ezio
    Mengozzi, Giulio
    Settanni, Fabio
    CLINICA CHIMICA ACTA, 2024, 557
  • [8] Volumetric absorptive microsampling as an effective microsampling technique for LC-MS/MS bioanalysis of biomarkers in drug discovery
    Kapadnis, Unnati
    Locuson, Chuck
    Okamura, Heidi
    Rienzo, Gianluca De
    Cotter, Cece
    Zhu, Dongwei
    Narayanaswami, Rohini
    Castro-Perez, Jose
    Marathe, Punit
    Yang, Wen-Chu
    BIOANALYSIS, 2023, 15 (14) : 845 - 859
  • [9] Volumetric absorptive microsampling (VAMS) as an alternative to conventional dried blood spots in the quantification of miltefosine in dried blood samples
    Kip, A. E.
    Kiers, K. C.
    Rosing, H.
    Schellens, J. H. M.
    Beijnen, J. H.
    Dorlo, T. P. C.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2017, 135 : 160 - 166
  • [10] Quantitation of phosphatidylethanol in dried blood after volumetric absorptive microsampling
    Van Uytfanghe, Katleen
    Fernandez, Maria del Mar Ramirez
    De Vos, Aurelie
    Wille, Sarah M. R.
    Stove, Christophe Pol
    TALANTA, 2021, 223