Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy in localizing horizontal rectus muscle insertions

被引:1
|
作者
Duan, Rui [1 ,2 ]
Yang, Jiwen [1 ,3 ,4 ,5 ]
机构
[1] Cent South Univ, Aier Sch Ophthalmol, Changsha, Peoples R China
[2] Lanzhou Univ, Affiliated Hosp 1, Lanzhou, Gansu, Peoples R China
[3] Liaoning Aier Eye Hosp, Shenyang, Liaoning, Peoples R China
[4] Cent South Univ, Aier Sch Ophthalmol, 188 South Furong Rd, Changsha 410000, Hunan, Peoples R China
[5] Liaoning Aier Eye Hosp, 11 Shiyiwei Rd, Shenyang 110000, Liaoning, Peoples R China
关键词
Strabismus; rectus muscle insertion; anterior segment optical coherence tomography; ultrasound biomicroscopy; ACCURACY;
D O I
10.1177/11206721231202539
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate the viability and precision of measuring the distance from the limbus to extraocular muscle insertion using anterior segment optical coherence tomography (AS-OCT) and panoramic ultrasound biomicroscopy (UBM) before and after strabismus surgery.Methods: We recruited primary strabismus patients and measured the limbus-insertion distance by AS-OCT and UBM preoperatively, 2 weeks, and 1, 3, and 6 months postoperatively. Values were also measured using callipers intraoperatively before and after the planned procedures. Preoperative AS-OCT and UBM values were compared to intraoperative calliper measurements as the gold standard. Postoperative AS-OCT and UBM values were compared to the new postoperative limbus-insertion distance. The limit of agreement deemed clinically acceptable was defined as 1 mm.Results: A total of 85 horizontal muscles of 40 patients, including 48 lateral rectus muscles and 37 medial rectus muscles, were analysed. Primary muscles could be successfully detected by AS-OCT (95%) and UBM (100%). At 2 weeks and 1, 3, and 6 months postoperatively, the new rectus muscle attachment site detection rate by AS-OCT was 6%, 32%, 80%, and 89%, respectively, and that by UBM was 24%, 60%, 85%, and 93%, respectively. The Bland-Altman plots revealed better consistency in pairs of AS-OCT, UBM, and calliper measurements of primary muscles than postoperative muscles. For primary muscles, 89% of AS-OCT measurements fell within the permissible range of surgical measurements (1 mm), but this dropped to 67% at 6 months postoperatively (P < 0.001). The accuracy of UBM measurements of primary muscles was 81%, and this decreased to 59% at 6 months postoperatively (P = 0.001).Conclusions: AS-OCT and UBM performed well in terms of imaging primary horizontal rectus muscles, but showed decreased accuracy and reproducibility in postoperative muscle measures.
引用
收藏
页码:656 / 665
页数:10
相关论文
共 50 条
  • [1] Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment
    Dada, Tanuj
    Sihota, Ramanjit
    Gadia, Ritu
    Aggarwal, Anand
    Mandal, Subrata
    Gupta, Viney
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (05): : 837 - 840
  • [2] THE ACCURACY OF ANTERIOR SEGMENT OPTICAL COHERENCE TOMOGRAPHY (AS-OCT) IN LOCALIZING EXTRAOCULAR RECTUS MUSCLES INSERTIONS
    Paciuc-Beja, Miguel
    Salcedo-Villanueva, Guillermo
    Quiroz-Mercado, Hugo
    JOURNAL OF AAPOS, 2015, 19 (05): : 489 - 490
  • [3] The accuracy of anterior segment optical coherence tomography (AS-OCT) in localizing extraocular rectus muscles insertions
    Ngo, Cheryl S.
    Smith, David
    Kraft, Stephen P.
    JOURNAL OF AAPOS, 2015, 19 (03): : 233 - 236
  • [4] Anterior segment imaging: ultrasound biomicroscopy and anterior segment optical coherence tomography
    Nolan, Winifred
    CURRENT OPINION IN OPHTHALMOLOGY, 2008, 19 (02) : 115 - 121
  • [5] Imaging of Anterior Segment Tumours: A Comparison of Ultrasound Biomicroscopy Versus Anterior Segment Optical Coherence Tomography
    Kottaridou, Eleni
    Hatoum, Adam
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [6] Comparison of Ultrasound Biomicroscopy versus Anterior Segment Optical Coherence Tomography for Imaging of Anterior Segment Tumors
    Bianciotto, C. G.
    Romanelli, M.
    Guzman, M.
    Mazzuca, D.
    Green, R.
    Shields, C. L.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [7] THE ACCURACY OF ANTERIOR SEGMENT OPTICAL COHERENCE TOMOGRAPHY (AS-OCT) IN LOCALIZING EXTRAOCULAR RECTUS MUSCLES INSERTIONS REPLY
    Ngo, Cheryl
    Smith, David
    Kraft, Stephen P.
    JOURNAL OF AAPOS, 2015, 19 (05): : 490 - 491
  • [8] Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy in the Imaging of Anterior Segment Tumors
    Pavlin, Charles J.
    Vasquez, Luz Maria
    Lee, Richard
    Simpson, E. Rand
    Ahmed, Iqbal Ike K.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (02) : 214 - 219
  • [9] Comparison of Anterior Segment Time Domain Optical Coherence Tomography and High Frequency Ultrasound Biomicroscopy for Anterior Segment
    Cardascia, N.
    Guerriero, S.
    Ferreri, P.
    Furino, C.
    Boscia, F.
    Vetrugno, M.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [10] Anterior Segment Imaging: Optical Coherence Tomography Versus Ultrasound Biomicroscopy
    Garcia, Julian P. S., Jr.
    Rosen, Richard B.
    OPHTHALMIC SURGERY LASERS & IMAGING, 2008, 39 (06) : 476 - 484