Calibration method of laser scanning measurement system for large-scale component using a free-form surface reconstruction model

被引:7
作者
Gu, Rongrui [1 ]
Wang, Zhongyu [1 ]
Ma, Liqun [2 ]
Nie, Jing [1 ]
机构
[1] Beihang Univ, Sch Instrumentat & Optoelect Engn, Beijing 100191, Peoples R China
[2] Beijing Chang cheng Inst Metrol & Measurement, Beijing 100191, Peoples R China
关键词
Calibration method; Surface reconstruction; Uncertainty analysis; Large-scale components; Point cloud data; UNCERTAINTY; METROLOGY;
D O I
10.1016/j.measurement.2023.112511
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
For large dimensional scanning measurement, the calibration method is limited by the standard part with the scale in millimeters or centermeters. In this paper, a novel calibration method suitable for large-scale components in meters has been developed based on the free-form surface reconstruction model. The Non-Uniform Rational B -Splines (NURBS) surface is approximated from the discrete measurement points obtained by high-precision detection using the least-squares method. The deviation between the measurement points and reference model is defined by statistical results and represented by the standard deviation of the distances from the measurement points to the corresponding closest points on the surface. Furthermore, the uncertainty components introduced in the procedure of coordinate registration and surface approximation are considered to estimate the combined uncertainty of a reconstructed surface. Finally, a scanning measurement device can be calibrated by the reconstruction model consisting of a NURBS surface expression and its uncertainty prediction. To validate the proposed calibration method, the surface of a standard part with a 4-meter diameter is detected to acquire a set of measurement points and subsequently reconstructed to a parametric surface, the uncertainty of which is analyzed considering errors introduced by both device accuracy and data processing. For the reconstruction model with degree of 4, the result shows that the calibration accuracy is superior to 30 mu m with a tolerance error of less than 12 mu m.
引用
收藏
页数:12
相关论文
共 23 条
  • [1] Evaluation and Optimization of Task-oriented Measurement Uncertainty for Coordinate Measuring Machines Based on Geometrical Product Specifications
    Cheng, Yinbao
    Wang, Zhongyu
    Chen, Xiaohuai
    Li, Yaru
    Li, Hongyang
    Li, Hongli
    Wang, Hanbin
    [J]. APPLIED SCIENCES-BASEL, 2019, 9 (01):
  • [2] Metrological evaluation of laser scanner integrated with measuring arm using optical feature-based gauge
    Cuesta, E.
    Alvarez, B. J.
    Martinez-Pellitero, S.
    Barreiro, J.
    Patino, H.
    [J]. OPTICS AND LASERS IN ENGINEERING, 2019, 121 : 120 - 132
  • [3] Calibration procedure for a laser triangulation scanner with uncertainty evaluation
    Genta, Gianfranco
    Minetola, Paolo
    Barbato, Giulio
    [J]. OPTICS AND LASERS IN ENGINEERING, 2016, 86 : 11 - 19
  • [4] Error evaluation of free-form surface based on distance function of measured point to surface
    He, Gaiyun
    Zhang, Mei
    Song, Zhanjie
    [J]. COMPUTER-AIDED DESIGN, 2015, 65 : 11 - 17
  • [5] An efficient sensing localization algorithm for free-form surface digitization
    Huang, Yunbao
    Qian, Xiaoping
    [J]. JOURNAL OF COMPUTING AND INFORMATION SCIENCE IN ENGINEERING, 2008, 8 (02) : 0210081 - 02100810
  • [6] Iso I., 1995, GENEVA SWITZ, V122, P16
  • [7] Jiang X.J., 2020, Advanced Metrology: Freeform Surfaces
  • [8] Uncertainty of NURBS surface fit by Monte Carlo simulations
    Koch, Karl-Rudolf
    [J]. JOURNAL OF APPLIED GEODESY, 2009, 3 (04) : 239 - 247
  • [9] Li Yaru, 2022, OPT ENG, V61
  • [10] Method for determining the probing points for efficient measurement and reconstruction of freeform surfaces
    Li, YF
    Liu, ZG
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2003, 14 (08) : 1280 - 1288