A methodology for incorporating a photon-counting CT system into routine clinical use

被引:7
作者
Bache, Steven T. [1 ]
Samei, Ehsan [2 ]
机构
[1] Duke Univ, Dept Radiol, Med Ctr, Clin Imaging Phys Grp, Durham, NC 27708 USA
[2] Duke Univ, Carl E Ravin Adv Imaging Labs, Med Ctr, Durham, NC USA
关键词
CT image quality; CT protocols; photon-counting CT; IMAGE QUALITY; RESOLUTION; SCANNERS; NOISE;
D O I
10.1002/acm2.14069
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Photon-counting computed tomography (PCCT) systems are increasingly available in the U.S. following Food and Drug Administration (FDA) approval of the first clinical PCCT system in Fall 2021. Consequently, there will be a need to incorporate PCCTs into existing fleets of traditional CT systems. The commissioning process of a PCCT was devised by evaluating the degree of agreement between the performance of the PCCT and that of established clinical CT systems. A PCCT system (Siemens NAEOTOM Alpha) was evaluated using the American College of Radiology(ACR) CT phantom (Gammex 464). The phantom was scanned on the system and on a 3rd Generation EID CT system (Siemens Force) at three clinical dose levels. Images were reconstructed across the range of available reconstruction kernels and Iterative Reconstruction (IR) strengths. Two image quality metrics-spatial resolution and noise texture-were calculated using AAPM TG233 software (imQuest), as well as a dose metric to achieve target image noise magnitude of 10 HU. For each pair of EID-PCCT kernel/IR strengths, the difference in metrics were calculated, weighted, and multiplied over all metrics to determine the concordance between systems. IR performance was characterized by comparing relative noise texture and reference dose as a function of IR strength for each system. In general, as kernel "sharpness" increased for each system, spatial resolution, noise spatial frequency, and reference dose increased. For a given kernel, EID reconstruction showed higher spatial resolution compared to PCCT in standard resolution mode. PCCT implementation of IR better preserved noise texture across all strengths compared to the EID, demonstrated by respective 20 and 7% shifts in noise texture from IR "Off" to IR "Max." Overall, the closest match for a given EID reconstruction kernel/IR strength was identified as a PCCT kernel with "sharpness" increased by 1 step and IR strength increased by 1-2 steps. Substantial dose reduction potential of up to 70% was found when targeting a constant noise magnitude.
引用
收藏
页数:9
相关论文
共 25 条
[1]  
[Anonymous], 2021, FDA Clears First Major Imaging Device Advancement for Computed Tomography in Nearly a Decade
[2]   Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods [J].
Chen, Baiyu ;
Christianson, Olav ;
Wilson, Joshua M. ;
Samei, Ehsan .
MEDICAL PHYSICS, 2014, 41 (07)
[3]   Photon-counting x-ray detectors for CT [J].
Danielsson, Mats ;
Persson, Mats ;
Sjolin, Martin .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03)
[4]   Impact of improved spatial resolution on radiomic features using photon-counting-detector CT [J].
Dunning, Chelsea A. S. ;
Rajendran, Kishore ;
Fletcher, Joel G. ;
McCollough, Cynthia H. ;
Leng, Shuai .
MEDICAL IMAGING 2022: IMAGE PROCESSING, 2022, 12032
[5]   First-generation clinical dual-source photon-counting CT: ultra-low-dose quantitative spectral imaging [J].
Liu, Leening P. ;
Shapira, Nadav ;
Chen, Andrew A. ;
Shinohara, Russell T. ;
Sahbaee, Pooyan ;
Schnall, Mitchell ;
Litt, Harold, I ;
Noel, Peter B. .
EUROPEAN RADIOLOGY, 2022, 32 (12) :8579-8587
[6]  
Rajagopal JR, 2021, IEEE T RADIAT PLASMA, V5, P588, DOI [10.1109/TRPMS.2020.3019954, 10.1109/trpms.2020.3019954]
[7]   Comparison of Low Dose Performance of Photon-Counting and Energy Integrating CT [J].
Rajagopal, Jayasai R. ;
Farhadi, Faraz ;
Solomon, Justin ;
Sahbaee, Pooyan ;
Saboury, Babak ;
Pritchard, William F. ;
Jones, Elizabeth C. ;
Samei, Ehsan .
ACADEMIC RADIOLOGY, 2021, 28 (12) :1754-1760
[8]   First Clinical Photon-counting Detector CT System: Technical Evaluation [J].
Rajendran, Kishore ;
Petersilka, Martin ;
Henning, Andre ;
Shanblatt, Elisabeth R. ;
Schmidt, Bernhard ;
Flohr, Thomas G. ;
Ferrero, Andrea ;
Baffour, Francis ;
Diehn, Felix E. ;
Yu, Lifeng ;
Rajiah, Prabhakar ;
Fletcher, Joel G. ;
Leng, Shuai ;
McCollough, Cynthia H. .
RADIOLOGY, 2022, 303 (01) :130-138
[9]   Full field-of-view, high-resolution, photon-counting detector CT: technical assessment and initial patient experience [J].
Rajendran, Kishore ;
Petersilka, Martin ;
Henning, Andre ;
Shanblatt, Elisabeth ;
Marsh, Jeffrey, Jr. ;
Thorne, Jamison ;
Schmidt, Bernhard ;
Flohr, Thomas ;
Fletcher, Joel ;
McCollough, Cynthia ;
Leng, Shuai .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (20)
[10]   Towards task-based assessment of CT performance: System and object MTF across different reconstruction algorithms [J].
Richard, Samuel ;
Husarik, Daniela B. ;
Yadava, Girijesh ;
Murphy, Simon N. ;
Samei, Ehsan .
MEDICAL PHYSICS, 2012, 39 (07) :4115-4122