Role of Matrix-Assisted Laser Desorption Ionization-Time-of-Flight Mass Spectrometry for Species Identification of Acinetobacter Strains

被引:1
|
作者
Krishnaraj, Anusha [1 ]
Gupta, Pratima [2 ,3 ]
Bhatia, Mohit [2 ]
Omar, Balram Ji [2 ]
机构
[1] St Johns Med Coll, Dept Microbiol, Bengaluru, Karnataka, India
[2] All India Inst Med Sci, Dept Microbiol, Deoghar, Jharkhand, India
[3] All India Inst Med Sci, Dept Microbiol, Deogarh 814142, Jharkhand, India
关键词
Acinetobacter species; antibiotic sensitivity; MALDI-TOF; MicroScan; COLISTIN SUSCEPTIBILITY; BAUMANNII; SYSTEM; RESISTANCE; COMPLEX; JUNII; GENE;
D O I
10.1055/s-0042-1760401
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction Acinetobacter species has become a leading cause of nosocomial infections in recent years.Objectives The aim of the study was to establish the usefulness of matrix-assisted laser desorption ionization-time-of-flight (MALDI-TOF) mass spectrometry (MS) for the identification of Acinetobacter species with respect to conventional biochemical methods and MicroScan WalkAway 96 Plus system and to compare the antibiotic susceptibility test results Kirby-Bauer disk diffusion method with MicroScan WalkAway 96 Plus automated identification and antimicrobial susceptibility testing system.Materials and Methods The study sample comprised 100 clinical isolates of Acinetobacter species. They were all identified using MALDI-TOF MS and compared with other two identification systems.Statistical Analysis Comparison of categorical variables by Fisher's exact test or Pearson's chi-square test was done. All statistical tools were two tailed, and a significant level p < 0.05 was used. All statistical tests were performed using SPSS v22.0 (Armonk IBM Corp., New York, United States). Cohen's kappa coefficients were also calculated and used as applicable.Results MALDI-TOF MS revealed 92 A. baumannii , 2 Acinetobacter nosocomialis , 3 Acinetobacter lwoffii , and 1 each was identified as Acinetobacter junii , Acinetobacter johnsonii , and Acinetobacter tandoii . There was moderate agreement between identification by MicroScan WalkAway and MALDI-TOF, and substantial agreement between conventional biochemical tests and MALDI-TOF. We found that there was a 100% categorical agreement with respect to susceptibility of aminoglycosides (amikacin, gentamicin, tobramycin) and cephalosporins (ceftazidime, cefepime, cefotaxime) between disk diffusion method and MicroScan WalkAway 96 Plus system. Total of 16 errors were observed.Conclusion Although MALDI-TOF MS could be useful to identify A. baumannii but not other species in the genus, it is a rapid, reliable method and can be routinely used in clinical laboratories.
引用
收藏
页码:336 / 343
页数:8
相关论文
共 50 条
  • [41] Typing of Methicillin-Resistant Staphylococcus aureus by Matrix-Assisted Laser Desorption Ionization-Time-of-Flight Mass Spectrometry
    Wolters, M.
    Christner, M.
    Aepfelbacher, M.
    Rohde, H.
    INTERNATIONAL JOURNAL OF MEDICAL MICROBIOLOGY, 2009, 299 : 3 - 3
  • [42] Matrix-assisted laser desorption ionization time-of-flight mass spectrometry: a powerful tool for identification of Corynebacterium species
    Bao, Rong
    Gao, Xiaodong
    Hu, Bijie
    Zhou, Zhaoyan
    JOURNAL OF THORACIC DISEASE, 2017, 9 (09) : 3239 - 3245
  • [43] Species Identification of Clinical Prevotella Isolates by Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry
    Wybo, Ingrid
    Soetens, Oriane
    De Bel, Annelies
    Echahidi, Fedoua
    Vancutsem, Ellen
    Vandoorslaer, Kristof
    Pierard, Denis
    JOURNAL OF CLINICAL MICROBIOLOGY, 2012, 50 (04) : 1415 - 1418
  • [44] Identification of Challenging Dermatophyte Species Using Matrix-Assisted Laser Desorption/Ionization Time-of-Flight Mass Spectrometry
    Tsai, Tsung-Fu
    Fan, Yun-Chen
    Lu, Jang-Jih
    Chien, Chun-Chih
    Wang, Hsin-Yao
    Sun, Pei-Lun
    JOURNAL OF FUNGI, 2025, 11 (02)
  • [45] Rapid identification and typing of Listeria species by matrix-assisted laser desorption ionization-time of flight mass spectrometry
    Barbuddhe, Sukhadeo B.
    Maier, Thomas
    Schwarz, Gerold
    Kostrzewa, Markus
    Hof, Herbert
    Domann, Eugen
    Chakraborty, Trinad
    Hain, Torsten
    APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2008, 74 (17) : 5402 - 5407
  • [46] Pseudallescheria/Scedosporium complex species identification by matrix-assisted laser desorption ionization time-of-flight mass spectrometry
    Coulibaly, Oumar
    Marinach-Patrice, Carine
    Cassagne, Carole
    Piarroux, Renaud
    Mazier, Dominique
    Ranque, Stephane
    MEDICAL MYCOLOGY, 2011, 49 (06) : 621 - 626
  • [47] A Canine Case of Nocardia africana Infection Detected by Matrix-Assisted Laser Desorption Ionization-Time-of-Flight Mass Spectrometry
    Yoon, Ji-Seon
    So, Hyungjae
    Joo, Beomsung
    Park, Jihong
    Jeong, In-Seong
    Lee, Gi-Jong
    Park, Jinho
    VETERINARY SCIENCES, 2022, 9 (06)
  • [48] Guanidination of Tryptic Peptides without Desalting for Matrix-Assisted Laser Desorption/Ionization-Time-of-Flight Mass Spectrometry Analysis
    Baker, Margaret R.
    Li, Qing X.
    ANALYTICAL CHEMISTRY, 2013, 85 (18) : 8873 - 8880
  • [49] Phyloproteomics:: Species identification of Enterobacteriaceae using matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
    Conway, GC
    Smole, SC
    Sarracino, DA
    Arbeit, RD
    Leopold, PE
    JOURNAL OF MOLECULAR MICROBIOLOGY AND BIOTECHNOLOGY, 2001, 3 (01) : 103 - 112
  • [50] Species Identification and Delineation of Pathogenic Mucorales by Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry
    Shao, Jin
    Wan, Zhe
    Li, Ruoyu
    Yu, Jin
    JOURNAL OF CLINICAL MICROBIOLOGY, 2018, 56 (04)