Fully Integrated Laser Guidance for CT-Based Punctures: A Study in Phantoms and Patients

被引:3
作者
Becker, Lena S. [1 ,3 ]
Ringe, Kristina I. [1 ]
Dewald, Cornelia L. A. [1 ]
Canstein, Christian [2 ]
Hinrichs, Jan B. [1 ]
Wacker, Frank K. [1 ]
Meyer, Bernhard C. [1 ]
机构
[1] Hannover Med Sch, Dept Diagnost & Intervent Radiol, Hannover, Germany
[2] Siemens Healthineers, Forchheim, Germany
[3] Hannover Med Sch, Dept Diagnost & Intervent Radiol, Carl Neuberg Str 1, D-30625 Hannover, Germany
关键词
PERCUTANEOUS PUNCTURES; IMAGE; NAVIGATION; ACCURACY; BIOPSIES; SYSTEMS;
D O I
10.1016/j.jvir.2022.09.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To test the hypothesis of equal or even superior applicability and accuracy of a fully integrated, laser-based computed tomography (CT) navigation system compared with conventional CT guidance for percutaneous interventions.Materials and Methods: CT-guided punctures were first performed in phantoms. Four radiologists with different experi-ence levels (2 residents (L.B., C.D.) and 2 board-certified radiologists (B.M., K.R.) performed 48 punctures using both conventional image-guided and laser-guided approaches. Subsequently, 12 punctures were performed in patients during a clinical pilot trial. Phantom targets required an in-plane or a single-/double-angulated, out-of-plane approach. Planning and intervention time, control scan number, radiation exposure, and accuracy of needle placement (measured by deviation of the needle tip to the designated target) were assessed for each guidance technique and compared (Mann-Whitney U test and t test). Patient interventions were additionally analyzed for applicability in a clinical setting.Results: The application of laser guidance software in the phantom study and in 12 human patients in a clinical setting was both technically and clinically feasible in all cases. The mean planning time (P = .009), intervention time (P = .005), control scan number (P < .001), and radiation exposure (P = .013) significantly decreased for laser-navigated punctures compared with those for conventional CT guidance and especially in punctures with out-of-plane-trajectories. The accuracy signifi-cantly increased for laser-guided interventions compared with that for conventional CT (P < .001).Conclusions: Interventional radiologists with differing levels of experience performed faster and more accurate punctures for out-of-plane trajectories in the phantom models, using a new, fully integrated, laser-guided CT software and demon-strated excellent clinical and technical success in initial clinical experiments.
引用
收藏
页码:130 / 138
页数:9
相关论文
共 25 条
[1]   System for CT-guided needle placement in the thorax and abdomen: A design for clinical acceptability, applicability and usability [J].
Arnolli, Maarten M. ;
Buijze, Martijn ;
Franken, Michel ;
de Jong, Koert P. ;
Brouwer, Dannis M. ;
Broeders, Ivo A. M. J. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2018, 14 (01)
[2]   An overview of systems for CT- and MRI-guided percutaneous needle placement in the thorax and abdomen [J].
Arnolli, Maarten M. ;
Hanumara, Nevan C. ;
Franken, Michel ;
Brouwer, Dannis M. ;
Broeders, Ivo A. M. J. .
INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2015, 11 (04) :458-475
[3]   Evaluation of a CT-Guided Robotic System for Precise Percutaneous Needle Insertion [J].
Ben-David, Eliel ;
Shochat, Moran ;
Roth, Ido ;
Nissenbaum, Isaac ;
Sosna, Jacob ;
Goldberg, S. Nahum .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 29 (10) :1440-1446
[4]   Cone beam CT guidance provides superior accuracy for complex needle paths compared with CT guidance [J].
Busser, W. M. H. ;
Braak, S. J. ;
Futterer, J. J. ;
van Strijen, M. J. L. ;
Hoogeveen, Y. L. ;
de Lange, F. ;
Kool, L. J. Schultze .
BRITISH JOURNAL OF RADIOLOGY, 2013, 86 (1030)
[5]   Computer assisted electromagnetic navigation improves accuracy in computed tomography guided interventions: A prospective randomized clinical trial [J].
Durand, Pierre ;
Moreau-Gaudry, Alexandre ;
Silvent, Anne-Sophie ;
Frandon, Julien ;
Chipon, Emilie ;
Medici, Maud ;
Bricault, Ivan .
PLOS ONE, 2017, 12 (03)
[6]   Phantom and Animal Study of a Robot-Assisted, CT-Guided Targeting System using Image-Only Navigation for Stereotactic Needle Insertion without Positional Sensors [J].
Fong, Khi Yung ;
Tan, Alexander Sheng Ming ;
Sulaiman, Muhammad Syadad Bin ;
Leong, Siang Huei ;
Ng, Ka Wei ;
Too, Chow Wei .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 33 (11) :1416-+
[7]   A LIGHT-GUIDANCE SYSTEM TO BE USED FOR CT-GUIDED BIOPSY [J].
FREDERICK, PR ;
BROWN, TH ;
MILLER, MH ;
BAHR, AL ;
TAYLOR, KH .
RADIOLOGY, 1985, 154 (02) :535-536
[8]   Guidance of percutaneous pulmonary biopsies with real-time CT fluoroscopy [J].
Froelich, JJ ;
Ishaque, N ;
Regn, J ;
Saar, B ;
Walthers, EM ;
Klose, KJ .
EUROPEAN JOURNAL OF RADIOLOGY, 2002, 42 (01) :74-79
[9]   Accuracy and complications in computed tomography fluoroscopy-guided needle biopsies of lung masses [J].
Heck, SL ;
Blom, P ;
Berstad, A .
EUROPEAN RADIOLOGY, 2006, 16 (06) :1387-1392
[10]   Robotic systems for percutaneous needle-guided interventions [J].
Kettenbach, Joachim ;
Kronreif, Gernot .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2015, 24 (01) :45-53