Developing an evidence-based approach to quality control

被引:13
作者
Badrick, Tony [1 ,3 ]
Loh, Tze Ping [2 ]
机构
[1] RCPA Qual Assurance Programs, Sydney, NSW, Australia
[2] Natl Univ Singapore Hosp, Singapore, Singapore
[3] RCPAQAP, 8 Herbert St, St Leonards, NSW 2065, Australia
关键词
Quality control strategy; Sigma metric; Error detection; Patient-based real-time quality control; MOVING AVERAGE; SIGMA-METRICS; TIME; PERFORMANCE; OPTIMIZATION; IMPLEMENTATION; VALIDATION; ALGORITHM; LENGTH; TOOLS;
D O I
10.1016/j.clinbiochem.2023.01.011
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Effective Quality Control remains one of the pillars of Clinical Biochemistry. An understanding of the possible analytical errors that may occur, how to detect them efficiently and how to prevent them from causing patient harm are critical components of a Quality System. For some time, there have been questions about the theoretical basis of the models used to describe and detect analytical error. The current theory recognises two types of error, systematic and random and a system based on sampling the analytical process using a synthetic material to detect these errors. However, there are at least two other errors that are present. One is related to the QC material and the other, irregular errors. In this Opinion Piece, some of the underlying assumptions of Quality Control systems are described and analysed.
引用
收藏
页码:39 / 42
页数:4
相关论文
共 57 条
  • [21] Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices
    Koerbin, Gus
    Liu, Jiakai
    Eigenstetter, Alex
    Tan, Chin Hon
    Badrick, Tony
    Loh, Tze Ping
    [J]. BIOCHEMIA MEDICA, 2018, 28 (01)
  • [22] Lim C., 2022, PATHOLOGY
  • [23] Lim C.Y., 2020, J Lab Precis Med., V5, P29
  • [24] Moving standard deviation and moving sum of outliers as quality tools for monitoring analytical precision
    Liu, Jiakai
    Tan, Chin Hon
    Badrick, Tony
    Loh, Tze Ping
    [J]. CLINICAL BIOCHEMISTRY, 2018, 52 : 112 - 116
  • [25] Recommendation for performance verification of patient-based real-time quality control
    Loh, Tze Ping
    Bietenbeck, Andreas
    Cervinski, Mark A.
    van Rossum, Huub H.
    Katayev, Alex
    Badrick, Tony
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (08) : 1205 - 1213
  • [26] Recommendations for laboratory informatics specifications needed for the application of patient-based real time quality control
    Loh, Tze Ping
    Cervinski, Mark A.
    Katayev, Alex
    Bietenbeck, Andreas
    van Rossum, Huub
    Badrick, Tony
    [J]. CLINICA CHIMICA ACTA, 2019, 495 : 625 - 629
  • [27] Optimizing moving average control procedures for small-volume laboratories: can it be done?
    Lukic, Vera
    Ignjatovic, Svetlana
    [J]. BIOCHEMIA MEDICA, 2019, 29 (03)
  • [28] Assay Stability, the missing component of the Error Budget
    Mackay, Mark
    Hegedus, Gabe
    Badrick, Tony
    [J]. CLINICAL BIOCHEMISTRY, 2017, 50 (18) : 1136 - 1144
  • [29] Steady state errors and risk of a QC strategy
    Mackay, Mark A.
    Badrick, Tony C.
    [J]. CLINICAL BIOCHEMISTRY, 2019, 64 : 37 - 43
  • [30] Routine internal- and external-quality control data in clinical laboratories for estimating measurement and diagnostic uncertainty using GUM principles
    Magnusson, Bertil
    Ossowicki, Haakan
    Rienitz, Olaf
    Theodorsson, Elvar
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2012, 72 (03) : 212 - 220