Sensitivity measurements for a 250 MHz quartz shear-horizontal surface acoustic wave biosensor under liquid viscous loading

被引:2
作者
Adetula, Ololade [1 ]
Aimofumhe, Eshiobomhe [1 ]
Badewole, Favour [1 ]
Ijale, Collins [1 ]
Thomas, Marlon [1 ]
机构
[1] Fisk Univ, Coll Nat Sci & Math, Dept Life & Phys Sci, Bioengn Program, Nashville, TN 37027 USA
基金
美国能源部; 美国国家科学基金会;
关键词
PHASE-NOISE; VISCOSITY; SENSORS;
D O I
10.1063/5.0166035
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Surface acoustic wave (SAW) devices have been used in biochemical assays due to their high sensitivity. The device sensitivity is a function of changes in the density and viscosity of the liquid. Here, we studied the effect of fluid viscosity using a 250 MHz quartz shear-horizontal (SH)-SAW biosensor by monitoring different concentrations of binary aqueous/glycerol solutions. In this study, the sensitivity of the biosensor was determined by fitting the data to models derived from perturbation theory. Measurements in water were used as the reference. For a 0% to 50% glycerol solution, an 87(degrees)-204(degrees) separation in the phase shift was observed. The slope of the plot of the phase shift vs (eta rho)(0.5) was used to indicate the sensor's sensitivity. The sensitivity for our 250 MHz quartz SH-SAW sensors was calculated to be 3.7x10(-3 )m(2)s/Kg. The corresponding mass sensitivity was determined to be 9.25 x 10(5) m(2)/Kg. The limit of detection was calculated to be 36 picograms (pg), while the limit of quantification or LOQ was calculated to be 109 pg. Traditionally, liquid phase measurements have been challenging for SAW devices because liquids dampen the vibrating sensors severely. This problem has been largely solved using a transverse (shear) wave instead of the more popular longitudinal or Rayleigh waves. Liquid measurements are now possible using transverse waves, also known as shear waves, because transverse waves are only minimally attenuated by liquids. Shear-horizontal SAW sensors (SH-SAW) show great promise as label-free biosensors because of their ability to handle liquid samples. However, the viscosity of the liquid still induces loading effects and can be measured when the liquid is loaded onto the SH-SAW propagating surface (delay line). When the liquid above the delay line is perturbed by physical or chemical changes, such as binding to a receptor, it alters the propagating acoustic wave. The SH-SAW device can measure these changes in liquid properties as a change in the wave's phase compared to the original wave. The device's phase shift was recorded as a function of the changes in the density and viscosity of the binary glycerol solution and used to determine the sensitivity in the linear dynamic range of responses.
引用
收藏
页数:7
相关论文
共 29 条
[1]   Electrothermal Modeling of Surface Acoustic Wave Resonators and Filters [J].
Akstaller, Wolfgang ;
Weigel, Robert ;
Hagelauer, Amelie .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2020, 67 (11) :2423-2432
[2]  
Auld B.A., 1990, Acoustic Fields and Waves in Solids, V2
[3]   PHASE NOISE IN SURFACE-ACOUSTIC-WAVE FILTERS AND RESONATORS [J].
BAER, RL .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (03) :421-425
[4]  
Ballantine D. S., 1996, Acoustic Wave Sensors: Theory, Design and Physico-Chemical Applications, P83
[5]   Analysis of Binary Mixtures of Aqueous Aromatic Hydrocarbons with Low-Phase-Noise Shear-Horizontal Surface Acoustic Wave Sensors Using Multielectrode Transducer Designs [J].
Bender, Florian ;
Mohler, Rachel E. ;
Ricco, Antonio J. ;
Josse, Fabien .
ANALYTICAL CHEMISTRY, 2014, 86 (22) :11464-11471
[6]  
Branch T. L., 2008, Shear Horizontal Surface Acoustic Wave Microsensor for Class A Viral and Bacterial Detection
[7]   Shear-horizontal surface acoustic wave sensor for non-invasive monitoring of dynamic cell spreading and attachment in wound healing assays [J].
Brugger, Manuel S. ;
Schnitzler, Lukas G. ;
Nieberle, Timo ;
Wixforth, Achim ;
Westerhausen, Christoph .
BIOSENSORS & BIOELECTRONICS, 2021, 173 (173)
[8]  
Campbell C., 1998, Surface Acoustic Wave Devices for Mobile and Wireless Communications, pxxvi
[9]   MULTIPARAMETER TUNING OF SURFACE-ACOUSTIC-WAVE FILTERS [J].
COHN, RW ;
WAGERS, RS .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (03) :255-264
[10]   Potential of surface acoustic wave biosensors for early sepsis diagnosis [J].
Csete, Marie ;
Hunt, William D. .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2013, 27 (04) :427-431