Analytical evaluation and quality assessment of the ARKRAY ADAMS A1c Lite HA-8380V for HbA1c

被引:0
作者
Urrechaga, Eloisa [1 ]
机构
[1] Hosp Galdakao Usansolo, Lab, Galdakao, Vizcaya, Spain
关键词
HbA1c; diabetes; performance specifications; verification; ARKRAY ADA MS A1c Lite HA-8380V analyzer; BIOLOGICAL VARIATION; GLYCATED HEMOGLOBIN; HBA(1C); HA-8190V; A(1C);
D O I
10.1080/00365513.2023.2225222
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: ARKRAY ADAMS A1c Lite HA-8380V is a fully automated high- performance liquid chromatography system for the measurement of HbA1c. We aimed to evaluate its analytical performance using the Variant Mode. Methods: Carry-over, linearity, imprecision, trueness, interferences and comparison against ARKRAY ADAMS A1c HA-8180V and HA-8180T analyzers were studied. Results: Total CVs 0.93% (IFCC units), 0.63% (NGSP units) at low concentration and 1.01% (IFCC) 0.74% (NGSP), at high concentration. Mean difference with the target values was -0.44 mmol/mol (IFCC) -0.04% (NGSP). Carry-over, linearity and method comparison were excellent. The results were not affected in the range of total Hb 39-199 g/L, labile fraction 5.7%, carbamylated Hb 9.1% nor acetylated Hb 7.8%, effect of common variants was negligible. Conclusions: the analyzer demonstrated very good analytical performances, according to the consensus criteria established for HbA1c; it is adequate for laboratories with medium-low workloads.
引用
收藏
页码:358 / 365
页数:8
相关论文
共 34 条
  • [1] Amer Diabet Assoc, 2014, DIABETES CARE, V37, pS81, DOI [10.2337/dc11-S062, 10.2337/dc14-S081, 10.2337/dc10-S062, 10.2337/dc12-s011, 10.2337/dc12-s064, 10.2337/dc11-S011, 10.2337/dc10-S011, 10.2337/dc13-S011, 10.2337/dc13-S067]
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Revaluation of biological variation of glycated hemoglobin (HbA1c) using an accurately designed protocol and an assay traceable to the IFCC reference system
    Braga, Federica
    Dolci, Alberto
    Montagnana, Martina
    Pagani, Franca
    Paleari, Renata
    Guidi, Gian Cesare
    Mosca, Andrea
    Panteghini, Mauro
    [J]. CLINICA CHIMICA ACTA, 2011, 412 (15-16) : 1412 - 1416
  • [4] Cercado AG, 2017, REV NEFROL DIAL TRAS, V37, P225
  • [5] CLSI, 2014, PREL EV QUANT CLIN L
  • [6] CLSI, 2003, Document EP6-A
  • [7] Diabetes in pregnancy: Effect on glycation and acetylation of the different chains of fetal and maternal hemoglobin
    Davison, A. S.
    Green, B. N.
    Roberts, N. B.
    [J]. CLINICAL BIOCHEMISTRY, 2011, 44 (2-3) : 198 - 202
  • [8] Labile glycated haemoglobin and carbamylated haemoglobin are still critical points for HbA1C measurement
    Desmons, Aurore
    Jaisson, Stephane
    Leroy, Nathalie
    Gillery, Philippe
    Guillard, Emmanuelle
    [J]. BIOCHEMIA MEDICA, 2017, 27 (02) : 378 - 386
  • [9] Finamore F., 2015, EuPA Open Proteomics, V8, P116, DOI DOI 10.1016/J.EUPROT.2015.04.003
  • [10] Review of hemoglobin A1c in the management of diabetes
    Gallagher, Emily Jane
    Le Roith, Derek
    Bloomgarden, Zachary
    [J]. JOURNAL OF DIABETES, 2009, 1 (01) : 9 - 17