Measurement of Occupational Eye and Thyroid Radiation Doses in Pediatric Interventional Cardiologists at a Tertiary Hospital

被引:3
|
作者
Khafaji, Mawya [1 ]
Ashour, Tariq M. [2 ]
Mozahim, Naif F. [2 ]
Tashkandi, Abdullah [2 ]
Alqarni, Basil [2 ]
Malaikah, Abdulrahman A. [2 ]
Bagabas, Abdulaziz K. [2 ]
Alwasabi, Abdullah A. [2 ]
Albahiti, Sarah K. [1 ]
机构
[1] King Abdulaziz Univ, Fac Med, Radiol Dept, Jeddah, Saudi Arabia
[2] King Abdulaziz Univ, Fac Med, Jeddah, Saudi Arabia
关键词
thyroid dose; eye lens dose; staff exposure; intervention pediatric cardiology; occupational radiation dose; cardiac catheterization; STAFF; EXTREMITIES; EXPOSURE; SCATTER;
D O I
10.7759/cureus.44894
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Advances in imaging techniques have led to increased utilization of imaging devices in catheterization laboratories. Invasive surgical procedures for cardiac disorders have been largely replaced by fluoroscopic cardiac catheterization. With this increase, concerns and risks associated with exposure to ionizing radiation among interventional cardiologists are growing. This study aims to measure and compare the occupational doses to the eye lens and thyroid of pediatric interventional cardiologists during different procedures in the catheterization laboratory and its significance. Methodology In this study, cardiologists wore bandanas with attached dosimeters to measure the absorbed doses to the eyes and thyroid gland. The dosimeters were collected for reading. The procedure types were also collected. In addition, the total fluoroscopy time and tube voltage of the biplane machine were measured. SPSS version 23 (IBM Corp., Armonk, NY, USA) was used to analyze the data. The characteristics of the study sample were described using simple counts and percentages, whereas means and standard deviations were used for continuous variables. Statistical significance was set at p-values <0.05. Results A total of 93 procedures were evaluated. The mean absorbed doses for all 93 procedures in both eyes and thyroid were 0.09 mGy and 0.08 mGy, respectively. A significant difference was found between the left and right eye measurements (p = 0.034), with higher doses administered to the left eye. However, no significant difference was observed between the right and left thyroid doses (p = 0.281). Significant correlations were found between the eye and thyroid doses and the procedure type (p = 0.02 and p = 0.009, respectively). Conclusions A significant amount of radiation was measured in the measurements of both organs. In addition, radiation dose measurements varied between different types of procedures. Our current results indicate the importance and necessity of applying the radiation protection concept of dose optimization.
引用
收藏
页数:8
相关论文
共 34 条
  • [1] Estimation of eye lens doses received by pediatric interventional cardiologists
    Alejo, L.
    Koren, C.
    Ferrer, C.
    Corredoira, E.
    Serrada, A.
    APPLIED RADIATION AND ISOTOPES, 2015, 103 : 43 - 47
  • [2] Eye lens radiation exposure to interventional cardiologists: a retrospective assessment of cumulative doses
    Jacob, Sophie
    Donadille, Laurent
    Maccia, Carlo
    Bar, Olivier
    Boveda, Serge
    Laurier, Dominique
    Bernier, Marie-Odile
    RADIATION PROTECTION DOSIMETRY, 2013, 153 (03) : 282 - 293
  • [3] Evaluation of patient doses and lens radiation doses to interventional cardiologists in a nationwide survey in Chile
    Ubeda, Carlos
    Vano, Eliseo
    Gonzalez, Luciano
    Miranda, Patricia
    Valenzuela, Edith
    Vergara, Francisco
    Guarda, Eduardo
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (01) : 36 - 43
  • [4] Assessment of radiation doses to cardiologists during interventional examinations
    Bor, D.
    Olgar, T.
    Onal, E.
    Caglan, A.
    Toklu, T.
    MEDICAL PHYSICS, 2009, 36 (08) : 3730 - 3736
  • [5] Measurement of occupational doses of ionising radiation to the lens of the eyes of interventional radiologists
    Higgins, Anthony
    JOURNAL OF RADIOLOGICAL PROTECTION, 2016, 36 (01) : 74 - 92
  • [6] Occupational radiation dose and protection to the eye lens for interventional cardiology operators
    Lian, Meng-En
    Wu, Guan-Yi
    Chao, Min-Fang
    Wang, Chong-Sheng
    Miao, Hsiao-Chien
    Tsai, Ming-Sung
    Hsieh, Mu-Yang
    Chen, Yeh-Peng
    Chu, Chun-Yuan
    Chang, Szu-Li
    Tsai, Hui-Yu
    RADIATION MEASUREMENTS, 2022, 154
  • [7] Eye lens dosimetry and the study on radiation cataract in interventional cardiologists
    Matsubara, Kosuke
    Lertsuwunseri, Vorarit
    Srimahachota, Suphot
    Krisanachinda, Anchali
    Tulvatana, Wasee
    Khambhiphant, Bharkbhum
    Sudchai, Waraporn
    Rehani, Madan
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 44 : 232 - 235
  • [8] Radiation-Induced Lens Opacities among Interventional Cardiologists: Retrospective Assessment of Cumulative Eye Lens Doses
    Struelens, L.
    Dabin, J.
    Carinou, E.
    Askounis, P.
    Ciraj-Bjelac, O.
    Domienik-Andrzejewska, J.
    Berus, D.
    Padovani, R.
    Farah, J.
    Covens, P.
    RADIATION RESEARCH, 2018, 189 (04) : 399 - 408
  • [9] Application of the ELDO approach to assess cumulative eye lens doses for interventional cardiologists
    Farah, J.
    Struelens, L.
    Auvinen, A.
    Jacob, S.
    Koukorava, C.
    Schnelzer, M.
    Vanhavere, F.
    Clairand, I.
    RADIATION PROTECTION DOSIMETRY, 2015, 164 (1-2) : 84 - 88
  • [10] Track, calculate and optimise eye lens doses of interventional cardiologists using mEyeDose and mEyeDose_X
    Covens, P.
    Dabin, J.
    De Troyer, O.
    Dragusin, O.
    Maushagen, J.
    Struelens, L.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 678 - 687