Comparison of radium-226 separation methods based on chromatographic and extraction resins for its determination by ICP-MS in drinking waters

被引:4
作者
Roulier, Marine [1 ]
Baya, Pascale Anabelle [2 ]
Roberge, Steeve [2 ]
Lariviere, Dominic [1 ]
机构
[1] Univ Laval, Dept Chim, Lab Radioecol, Quebec City, PQ G1V 0A6, Canada
[2] Quebec Minist Environm Lutte Changements Climat, Div Chim Inorgan Milieu, Direct Anal Chim, Direct Gen Coordinat Sci, Faune & Parcs, Quebec City, PQ, Canada
来源
JOURNAL OF MASS SPECTROMETRY | 2024年 / 59卷 / 02期
基金
加拿大自然科学与工程研究理事会;
关键词
chromatographic resin; drinking waters; environmental; ICP-MS; radium; MASS-SPECTROMETRY; ENVIRONMENTAL-SAMPLES; RAPID-DETERMINATION; NATURAL-WATERS; RADIUM; PRECONCENTRATION; QUANTIFICATION;
D O I
10.1002/jms.5005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Over the past century, human activities have contributed to the release of 226Ra (t1/2 = 1,600 y) in the environment, increasing the potential risks for human exposure and thus prompting scientists to monitor it. Inductively coupled plasma mass spectrometry (ICP-MS) is an alternative to alpha-spectrometry for the quantification of 226Ra. However, the performances of radioanalytical procedures are rarely compared in a rigorous framework, which means that researchers may choose one on subjective factors or guesses.This article compares five published methods for the separation and preconcentration of 226Ra in drinking waters based on chromatographic and extraction resins prior to its analysis by ICP-MS. We evaluated the turnaround time, generated wastes, and final cost of each protocol as the economic aspect can be an important criterion when selecting a method, particularly for sustainable environmental monitoring. Our results showed that 226Ra was successfully separated and preconcentrated, yielding recoveries ranging between 84% and 105%. Method detection and quantification limits of respectively 2-7 fg L-1 (0.1-0.3 mBq L-1) and 6-24 fg L-1 (0.2-0.9 mBq L-1) were achieved when the separation method was coupled with ICP-MS. The turnaround times ranged between 6 and 21 hours, whereas the cost of the methods varied between 40 and 132 USD.This study highlights for the first time that methodologies recently published on the evaluation of 226Ra levels in drinking water by ICP-MS have comparable figures of merit. Our results offer essential insights into the selection of the most suitable separation method.
引用
收藏
页数:10
相关论文
共 30 条
  • [1] A review of the analytical methodology to determine Radium-226 and Radium-228 in drinking waters
    Abbasi, Akbar
    [J]. RADIOCHIMICA ACTA, 2018, 106 (10) : 819 - 829
  • [2] [Anonymous], CEAEQ, 20230801, Methode d'Analyse 603 Radium226, Determination du radium226 dans l'eau potable, les eaux naturelles et les eaux usees: methode par coprecipitation, purification sur resine cationique et dosage par ICPMS/MS
  • [3] [Anonymous], 2005, ATLAS EH PH DIAGRAMS
  • [4] Becker J. S., 2003, Spectroscopy Europe, V14, P8
  • [5] Determination of long-lived radionuclides by inductively coupled plasma quadrupole mass spectrometry using different nebulizers
    Becker, JS
    Soman, RS
    Sutton, KL
    Caruso, JA
    Dietze, HJ
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (06) : 933 - 937
  • [6] Detection of radium at the attogram per gram level in copper by inductively coupled plasma mass spectrometry after cation-exchange chromatography
    Bonin, Melodie
    Lariviere, Dominic
    Povinec, Pavel P.
    [J]. ANALYTICAL METHODS, 2020, 12 (17) : 2272 - 2278
  • [7] 226Ra and 137Cs determination by inductively coupled plasma mass spectrometry: state of the art and perspectives including sample pretreatment and separation steps
    Boudias, Marine
    Gourgiotis, Alkiviadis
    Montavon, Gilles
    Cazala, Charlotte
    Pichon, Valerie
    Delaunay, Nathalie
    [J]. JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2022, 244
  • [8] Copia L., 2015, J ANAL SCI TECHNOL, V6, P22, DOI DOI 10.1186/S40543-015-0062-5
  • [9] Rapid, versatile and sensitive method for the quantification of radium in environmental samples through cationic extraction and inductively coupled plasma mass spectrometry
    Dalencourt, Claire
    Michaud, Annie
    Habibi, Azza
    Leblanc, Alexa
    Lariviere, Dominic
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (06) : 1031 - 1040
  • [10] A rapid determination of 226Ra in water using composite ion exchanger MnO2-PAN
    Dulanska, Silvia
    Gardonova, Veronika
    Sebesta, Ferdinand
    Matel, Lubomir
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2015, 303 (01) : 47 - 51