Influence of the device used for obstructive sleep apnea diagnosis on body position: a comparison between polysomnography and portable monitor

被引:4
|
作者
Mello, Andre A. F. [1 ]
D'Angelo, Giovanna [1 ]
Santos, Ronaldo B. [2 ]
Bensenor, Isabela [2 ]
Lotufo, Paulo A. [2 ]
Lorenzi-Filho, Geraldo [1 ]
Drager, Luciano F. [3 ,4 ]
Genta, Pedro R. [1 ]
机构
[1] Univ Sao Paulo, Hosp Clin HCFMUSP, Inst Coracao InCor, Lab Sono,Div Pneumol, LIM 63,Ave Doutor Eneas de Carvalho Aguiar 44, BR-05403900 Sao Paulo, SP, Brazil
[2] Univ Sao Paulo, Ctr Clin & Epidemiol Res CPCE, Cidade Univ,Av Prof Lineu Prestes 2565,4 Andar, BR-05508000 Sao Paulo, SP, Brazil
[3] Univ Sao Paulo, Renal Div, Hypertens Unit, Med Sch, Sao Paulo, SP, Brazil
[4] Univ Sao Paulo, Heart Inst InCor, Hypertens Unit, Med Sch, Sao Paulo, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
Obstructive sleep apnea; Polysomnography; Portable monitor; Posture; ADULT;
D O I
10.1007/s11325-022-02674-3
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose Different devices have been used for the diagnosis of obstructive sleep apnea (OSA), which differ in the number of sensors used. The numerous sensors used in more complex sleep studies such as in-lab polysomnography may influence body position during sleep. We hypothesized that patients submitted to in-lab polysomnography (PSG) would spend more time in the supine position than patients submitted to an ambulatory Portable Monitor (PM) sleep study. Methods Body position during PSG and PM studies was compared among two distinct groups of patients matched for age, body-mass index (BMI), apnea-hypopnea index (AHI), and gender. Predictors of time spent in the supine position were determined using a multiple linear regression model. Results Of 478 participants who underwent either PSG or PM studies, mean age: 61[43-66] years; males: 43.9%; BMI: 28.4[26.1-31.1]kg/m(2); AHI 14[7-27] events/hour). Participants who underwent PSG studies spent more time in the supine position (41[16-68]% than participants who underwent PM studies (34[16-51]%), P = 0.014. Participants with OSA spent more time in the supine position than participants without OSA, both among the PSG and PM groups P < 0.05). Gender, BMI, OSA severity, and sleep study type were independent predictors of time spent in the supine position. Conclusion In-lab PSG may increase time spent in the supine position and overestimate OSA severity compared to a PM sleep study. OSA diagnosis is also associated with increased time spent in the supine position. The potential influence on the sleeping position should be taken into account when choosing among the different sleep study types for OSA diagnosis.
引用
收藏
页码:887 / 891
页数:5
相关论文
共 50 条
  • [1] Influence of the device used for obstructive sleep apnea diagnosis on body position: a comparison between polysomnography and portable monitor
    André A. F. Mello
    Giovanna D´Angelo
    Ronaldo B. Santos
    Isabela Bensenor
    Paulo A. Lotufo
    Geraldo Lorenzi-Filho
    Luciano F. Drager
    Pedro R. Genta
    Sleep and Breathing, 2023, 27 : 887 - 891
  • [2] A COMPARISON OF A WRIST WORN PORTABLE DEVICE (WATCHPAT)™ WITH IN-LAB POLYSOMNOGRAPHY FOR THE DIAGNOSIS OF OBSTRUCTIVE SLEEP APNEA
    Mittal, S.
    Suri, T. M.
    Hadda, V.
    Madan, K.
    Mohan, A.
    Pb, S.
    Tiwari, P.
    Guleria, R.
    SLEEP MEDICINE, 2019, 64 : S259 - S260
  • [3] A comparison of polysomnography and a portable home sleep study in the diagnosis of obstructive sleep apnea syndrome
    Su, S
    Baroody, FM
    Kohrman, M
    Suskind, D
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2004, 131 (06) : 844 - 850
  • [4] Comparison of portable monitoring devices and polysomnography for the diagnosis of obstructive sleep apnea syndrome
    Yano, T.
    Aoshima, Y.
    Tanaka, K.
    Ogasawara, T.
    Kasamatsu, N.
    JOURNAL OF SLEEP RESEARCH, 2014, 23 : 118 - 118
  • [5] A comparison of polysomnography and the WatchPAT in the diagnosis of obstructive sleep apnea
    Pang, Kenny P.
    Gourin, Christine G.
    Terris, David J.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2007, 137 (04) : 665 - 668
  • [6] Pilot evaluation of a portable monitor for diagnosis of obstructive sleep apnea in a consecutive series of patients compared with polysomnography
    Helen, D. S.
    Bjerring, K. A.
    Stewart, S. C.
    Munt, P.
    Fitzpatrick, M. P.
    SLEEP, 2008, 31 : A341 - A341
  • [7] A Comparison of a Wrist-Worn Portable Device (WatchPAT 200)™ with In-Lab Polysomnography for the Diagnosis of Obstructive Sleep Apnea
    Boppana T.K.
    Suri T.M.
    Hadda V.
    Madan K.
    Mohan A.
    Tiwari P.
    Mittal S.
    Sleep and Vigilance, 2024, 8 (2) : 209 - 214
  • [8] Comparative study of a wearable intelligent sleep monitor and polysomnography monitor for the diagnosis of obstructive sleep apnea
    Xu, Yanxia
    Ou, Qiong
    Cheng, Yilu
    Lao, Miaochan
    Pei, Guo
    SLEEP AND BREATHING, 2023, 27 (01) : 205 - 212
  • [9] Comparative study of a wearable intelligent sleep monitor and polysomnography monitor for the diagnosis of obstructive sleep apnea
    Yanxia Xu
    Qiong Ou
    Yilu Cheng
    Miaochan Lao
    Guo Pei
    Sleep and Breathing, 2023, 27 : 205 - 212
  • [10] VALIDATION OF PORTABLE MONITOR COMPARED WITH POLYSOMNOGRAPHY FOR SCREENING OF OBSTRUCTIVE SLEEP APNEA IN POLIO SURVIVORS
    Ding, Q.
    Liu, J.
    Wu, J.
    Du, J.
    Li, X.
    Wang, M.
    Sun, Y.
    Yu, Y.
    Wang, J.
    Sun, T.
    Zhang, C.
    Lv, C.
    Strohl, K. P.
    Han, F.
    Dong, X.
    SLEEP MEDICINE, 2024, 115 : 401 - 401