Simulation of radon detection efficiency with small pulse ionization chamber based on Geant4

被引:0
作者
Gou, Rui [1 ]
Shi, Rui [1 ]
Zhang, Qian [1 ]
Dong, Xiang-Long [2 ]
机构
[1] Sichuan Univ Sci & Engn, Yibin 644005, Sichuan, Peoples R China
[2] Hunan Commun Engn Polytech, Changsha 410000, Hunan, Peoples R China
基金
中国国家自然科学基金;
关键词
Pulsed ionization chamber; Radon; Detection efficiency; Monte Carlo simulation; Geant4;
D O I
10.1016/j.apradiso.2023.111152
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Radon measurement is crucial in assessing the damage to the human body caused by natural radiation. Pulsed ionization chambers are effective for real-time radon measurement and have widespread applications in other radiation techniques. However, due to practical constraints such as limited space and portability concerns, it becomes imperative to consider not only the detection efficiency but also their ease of transportation. This work utilizes the Geant4 Monte Carlo simulation toolkit to characterize the detection models of small cylindrical and flat plate-type pulsed ionization chambers, and carry out a simulation study to analyze the three crucial factors that influence detection efficiency, including the geometry of the chamber, electrode size, and operating temperature. The results indicate that the cylindrical pulse ionization chamber, with a length of 8 cm and radius of 2 cm, has the best detection efficiency and portability in terms of geometric dimensions, achieving a detection efficiency of (58 +/- 4)%. Meanwhile, the flat plate pulse ionization chamber, with dimensions of 7 cm in length and 3 cm in width, achieves the best detection efficiency and portability, with a detection efficiency of (54 +/- 3)%. In terms of electrode wire size, the cylindrical ionization chamber electrode wire with a length of 7 cm and a radius of 2.5 mm was optimal with a detection efficiency of (59 +/- 4)%. In terms of operating temperature, the detection efficiency of the flat-plate pulsed ionization chamber was the highest at 30 degrees C, which was (58 +/- 4)%, and that of the cylindrical pulsed ionization chamber was the highest at 20 degrees C, which was (63 +/- 4)%. By analyzing the influencing factors of the detection efficiency of the pulsed ionization chamber, it has a certain reference value and guiding significance for the research and design of small pulsed ionization chamber detectors for radon measuring instruments.
引用
收藏
页数:6
相关论文
共 50 条
[41]   Monte Carlo simulation of a PhosWatch detector using Geant4 for xenon isotope beta-gamma coincidence spectrum profile and detection efficiency calculations [J].
Mekarski, P. ;
Zhang, W. ;
Ungar, K. ;
Bean, M. ;
Korpach, E. .
APPLIED RADIATION AND ISOTOPES, 2009, 67 (10) :1957-1963
[42]   Scatter in an uncollimated x-ray CT machine based on a Geant4 Monte Carlo simulation [J].
Wadeson, Nicola ;
Morton, Edward ;
Lionheart, William .
MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
[43]   Geant4 track structure simulation of electron beam interaction with a gold nanoparticle [J].
Moradi, F. ;
Jalili, M. ;
Saraee, Kh. Rezaee Ebrahim ;
Khandaker, M. U. ;
Bradley, D. A. .
RADIATION PHYSICS AND CHEMISTRY, 2022, 200
[44]   GePEToS: A Geant4 Monte Carlo simulation package for positron emission tomography [J].
Jan, S ;
Collot, J ;
Gallin-Martel, ML ;
Martin, P ;
Mayet, F ;
Tournefier, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (01) :102-106
[45]   Validation of a GEANT4 simulation model for pinhole SPECT including calibration parameters [J].
Aguiar, Pablo ;
Pino, Francisco ;
Ros, Domenec ;
El Bitar, Ziad .
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, :2827-+
[46]   Evaluation of double photon coincidence Compton imaging method with GEANT4 simulation [J].
Yoshihara, Yuri ;
Shimazoe, Kenji ;
Mizumachi, Yuki ;
Takahashi, Hiroyuki .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 873 :51-55
[47]   An implementation of discrete electron transport models for gold in the Geant4 simulation toolkit [J].
Sakata, D. ;
Incerti, S. ;
Bordage, M. C. ;
Lampe, N. ;
Okada, S. ;
Emfietzoglou, D. ;
Kyriakou, I. ;
Murakami, K. ;
Sasaki, T. ;
Tran, H. ;
Guatelli, S. ;
Ivantchenko, V. N. .
JOURNAL OF APPLIED PHYSICS, 2016, 120 (24)
[48]   Interfacing Geant4, Garfield+ plus and Degrad for the simulation of gaseous detectors [J].
Pfeiffer, Dorothea ;
De Keukeleere, Lennert ;
Azevedo, Carlos ;
Belloni, Francesca ;
Biagi, Stephen ;
Grichine, Vladimir ;
Hayen, Leendert ;
Hanu, Andrei R. ;
Hrivnacova, Ivana ;
Ivanchenko, Vladimir ;
Krylov, Vladyslav ;
Schindler, Heinrich ;
Veenhof, Rob .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 935 :121-134
[49]   Geant4 simulation of the response of phosphor screens for X-ray imaging [J].
Pistrui-Maximean, S. A. ;
Freud, N. ;
Letang, J. M. ;
Koch, A. ;
Munier, B. ;
Walenta, A. H. ;
Montarou, G. ;
Babot, D. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 563 (01) :196-199
[50]   Evaluation of Neutron Field Parameters at the VITA Facility Using Geant4 Simulation [J].
Azarkin, M. Yu. ;
Zavestovskaya, I. N. ;
Kirakosyan, M. R. ;
Petrunya, D. S. ;
Kasatov, D. A. ;
Konovalova, V. D. ;
Taskaev, S. Yu. .
BULLETIN OF THE LEBEDEV PHYSICS INSTITUTE, 2025, 52 (02) :72-77