Insertion Torque (IT) and Implant Stability Quotient (ISQ) Assessment in Dental Implants with and without Healing Chambers: A Polyurethane In Vitro Study

被引:4
作者
Mello, Bruno Freitas [1 ]
Formiga, Marcio De Carvalho [1 ]
Bianchini, Marco Aurelio [2 ]
Borges Jr, Ivan [1 ]
Coura, Gustavo [1 ]
Tumedei, Margherita [3 ]
Fuller, Renato [1 ]
Petrini, Morena [4 ]
Romasco, Tea [4 ,5 ]
Vaz, Paula [6 ]
Piattelli, Adriano [7 ,8 ]
Di Pietro, Natalia [4 ,5 ]
机构
[1] Unisul Univ, Dept Periodontol & Oral Implantol, BR-88101001 Florianopolis, Brazil
[2] Fed Univ Santa Catarina UFSC, Postgrad Program Implant Dent PPGO, BR-88040900 Florianopolis, Brazil
[3] Univ Milan, Dept Med Surg & Dent Sci, I-20122 Milan, Italy
[4] G dAnnunzio Univ Chieti Pescara, Dept Med Oral & Biotechnol Sci, I-66013 Chieti, Italy
[5] G dAnnunzio Univ Chieti Pescara, Ctr Adv Studies & Technol CAST, I-66013 Chieti, Italy
[6] Univ Porto, Genet Fac Dent Med, Fixed Prosthodont, P-4200393 Porto, Portugal
[7] St Camillus Int Univ Hlth & Med Sci, Sch Dent, I-00131 Rome, Italy
[8] UCAM Univ Catolica San Antonio Murcia, Fac Med, Murcia 30107, Spain
来源
APPLIED SCIENCES-BASEL | 2023年 / 13卷 / 18期
关键词
dental implants; healing chambers; implant primary stability; insertion torque; polyurethane; removal torque; OSSEOINTEGRATION; DESIGN; MACROGEOMETRY; VIVO;
D O I
10.3390/app131810215
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Primary Stability (PS) depends on different factors, such as bone structure (quality and density), implant characteristics (macro and micro), and the relationship between thread shape and implant insertion hole size. PS is crucial for the prognosis and success of dental implants. The use of healing chambers, which create empty spaces between the implant and peri-implant bone, helps reduce the risk of compressive bone necrosis. High Insertion Torque (IT) values could potentially lead to the deterioration of the implant-abutment connection. However, the optimal implant macro-structure and IT values remain unknown. In this study, implants with healing chambers (Test) and without (Control) were inserted into polyurethane blocks with densities of 10, 20, 30, and 40 pounds per cubic foot (PCF). In blocks with densities of 30 and 40 PCF, Control implants had significantly higher IT values than Test implants. Additionally, Control implants exhibited significantly higher Removal Torque (RT) values than Test implants, but only in the 30 PCF density block (p < 0.0001). However, no differences were found between the implants in the 10, 20, and 40 PCF density blocks. Similarly, no significant differences were observed in the Implant Stability Quotient (ISQ) values between the Test and Control implants. The results of the present study confirm that adding healing chambers to the macro-structure of dental implants can significantly reduce IT values without affecting ISQ values. This suggests the potential for avoiding compression and damage to peri-implant bone while maintaining consistent levels of PS.
引用
收藏
页数:14
相关论文
共 35 条
  • [1] Macrogeometric Differentiation of Dental Implant Primary Stability: An In Vitro Study
    Al-Tarawneh, Sandra K.
    Thalji, Ghadeer
    Cooper, Lyndon F.
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (06) : 1110 - 1118
  • [2] Effects of insertion torque values on the marginal bone loss of dental implants installed in sheep mandibles
    Alexandre Gehrke, Sergio
    Aramburu, Jaime, Jr.
    Eirles Treichel, Tiago Luis
    do Prado, Tales Dias
    Dedavid, Berenice Anina
    de Aza, Piedad N.
    [J]. SCIENTIFIC REPORTS, 2022, 12 (01)
  • [3] Can changes in implant macrogeometry accelerate the osseointegration process?: An in vivo experimental biomechanical and histological evaluations
    Alexandre Gehrke, Sergio
    Aramburu, Jaime Junior
    Perez-Diaz, Leticia
    do Prado, Tales Dias
    Dedavid, Berenice Anina
    Mazon, Patricia
    De Aza, Piedad N.
    [J]. PLOS ONE, 2020, 15 (05):
  • [4] New Implant Macrogeometry to Improve and Accelerate the Osseointegration: An In Vivo Experimental Study
    Alexandre Gehrke, Sergio
    Aramburu Junior, Jaime
    Perez-Diaz, Leticia
    Eirles Treichel, Tiago Luis
    Dedavid, Berenice Anina
    De Aza, Piedad N.
    Carlos Prados-Frutos, Juan
    [J]. APPLIED SCIENCES-BASEL, 2019, 9 (15):
  • [5] Impact of Different Titanium Implant Thread Designs on Bone Healing: A Biomechanical and Histometric Study with an Animal Model
    Alexandre Gehrke, Sergio
    Eliers Treichel, Tiago Luis
    Perez-Diaz, Leticia
    Calvo-Guirado, Jose Luis
    Aramburu, Jaime, Jr.
    Mazon, Patricia
    de Aza, Piedad N.
    [J]. JOURNAL OF CLINICAL MEDICINE, 2019, 8 (06)
  • [6] [Anonymous], 2008, ASTM F1839, DOI [10.1520/F1839-08, DOI 10.1520/F1839-08]
  • [7] Drilling dimension effects in early stages of osseointegration and implant stability in a canine model
    Baires-Campos, Felipe-Eduardo
    Jimbo, Ryo
    Bonfante, Estevam-Augusto
    Fonseca-Oliveira, Maiolino-Thomaz
    Moura, Camila
    Zanetta-Barbosa, Darceny
    Coelho, Paulo-Guilherme
    [J]. MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2015, 20 (04): : E471 - E479
  • [8] Implant design and its effects on osseointegration over time within cortical and trabecular bone
    Beutel, Bryan G.
    Danna, Natalie R.
    Granato, Rodrigo
    Bonfante, Estevam A.
    Marin, Charles
    Tovar, Nick
    Suzuki, Marcelo
    Coelho, Paulo G.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2016, 104 (06) : 1091 - 1097
  • [9] Bonfante EA, 2011, INT J ORAL MAX IMPL, V26, P75
  • [10] Enhanced bone apposition to a chemically modified SLA titanium surface
    Buser, D
    Broggini, N
    Wieland, M
    Schenk, RK
    Denzer, AJ
    Cochran, DL
    Hoffmann, B
    Lussi, A
    Steinemann, SG
    [J]. JOURNAL OF DENTAL RESEARCH, 2004, 83 (07) : 529 - 533