Friction and wear behaviors of HXLPE, PEEK and CoCrMo against articular cartilage under cross-shear motion

被引:5
作者
Lu, Weiping [1 ]
Cui, Wen [1 ]
Zhang, Xiaogang [1 ]
Zhang, Yali [1 ]
Yang, Shu [1 ]
Pu, Jian [1 ]
Li, Junyan [1 ]
Jin, Zhongmin [1 ,2 ,3 ]
机构
[1] Southwest Jiaotong Univ, Sch Mech Engn, Tribol Res Inst, Chengdu 610031, Peoples R China
[2] Xi An Jiao Tong Univ, Sch Mech Engn, Xian, Peoples R China
[3] Univ Leeds, Sch Mech Engn, Leeds, W Yorkshire, England
基金
中国国家自然科学基金;
关键词
Cross-shear motion; friction; wear; cartilage; RESURFACING IMPLANT; SLIDING MOTIONS; HYDROGELS; DAMAGE;
D O I
10.1177/13506501221119017
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The friction and wear behaviors of cartilage replacement materials against articular cartilage are considered essential characteristics that determine the function and performance of the replacement. Previous studies have often been conducted under a reciprocal linear motion, while a cross-shear motion has been shown to widely exist in joints. The difference resulting between cross-shear motion and reciprocal linear motion, and the effect of bearing surface roughness on the frictional and wear behaviors of cartilage should be further investigated. In this study, experimental tribology investigations were conducted under the cross-shear motion for articular cartilage against commonly used three replacement materials, including high cross-linked polyethylene (HXLPE), polyether ether ketone (PEEK), and cobalt-chromium molybdenum alloy (CoCrMo). The experimental results showed that there was no significant difference in friction and wear behavior between cross-shear and reciprocal linear motion. For the implants with low surface roughness, the water contact angle of the bearing surface affected the friction and wear behaviors of the cartilage, while for the implants with high rough surfaces, the surface roughness of the material affected the friction and wear behaviors of the cartilage. The damage of cartilage under cross-shear motion mainly consisted of deformations and accumulations. This phenomenon was especially evident at high roughness. Of the three bearing materials considered, PEEK was more suitable for osteochondral implants.
引用
收藏
页码:552 / 561
页数:10
相关论文
共 38 条
[1]   Hemiarthroplasties: the choice of prosthetic material causes different levels of damage in the articular cartilage [J].
Ajdari, Niloofar ;
Tempelaere, Christine ;
Masouleh, Maryam Imani ;
Abel, Richard ;
Delfosse, Daniel ;
Emery, Roger ;
Dini, Daniele ;
Hansen, Ulrich .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2020, 29 (05) :1019-1029
[2]   Simple geometry tribological study of osteochondral graft implantation in the knee [J].
Bowland, Philippa ;
Ingham, Eileen ;
Fisher, John ;
Jennings, Louise M. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2018, 232 (03) :249-256
[3]   Measurement of protein denaturation in human synovial fluid and its analogs using differential scanning calorimetry [J].
Briere, Lee-Ann K. ;
Brandt, Jan-M. ;
Medley, John B. .
JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2010, 102 (01) :99-106
[4]   Torsional fretting wear behaviour of nature articular cartilage in vitro [J].
Cai, Zhen-bing ;
Gao, Shan-shan ;
Gan, Xue-qi ;
Yu, Hai-yang ;
Zhu, Min-hao .
INTERNATIONAL JOURNAL OF SURFACE SCIENCE AND ENGINEERING, 2011, 5 (5-6) :348-368
[5]   Friction and Wear of Hemiarthroplasty Biomaterials in Reciprocating Sliding Contact With Articular Cartilage [J].
Chan, S. M. T. ;
Neu, C. P. ;
Komvopoulos, K. ;
Reddi, A. H. ;
Di Cesare, P. E. .
JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2011, 133 (04)
[6]   Effect of Rotational and Sliding Motions on Friction and Degeneration of Articular Cartilage under Dry and Wet Friction [J].
Cilingir, Ahmet C. .
JOURNAL OF BIONIC ENGINEERING, 2015, 12 (03) :464-472
[7]   Polyurethane scaffold implants for partial meniscus lesions: delayed intervention leads to an inferior outcome [J].
Condello, Vincenzo ;
Dei Giudici, Luca ;
Perdisa, Francesco ;
Screpis, Daniele Umberto ;
Guerriero, Massimo ;
Filardo, Giuseppe ;
Zorzi, Claudio .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2021, 29 (01) :109-116
[8]   Replacement of the Medial Tibial Plateau by a Metallic Implant in a Goat Model [J].
Custers, Roel J. H. ;
Saris, Daniel B. F. ;
Creemers, Laura B. ;
Verbout, Ab J. ;
van Rijen, Mattie H. P. ;
Mastbergen, Simon C. ;
Lafeber, Floris P. ;
Dhert, Wouter J. A. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2010, 28 (04) :429-435
[9]   Bone response to zirconia ceramic implants: an experimental study in rabbit [J].
Di Carlo, F ;
Prosper, L ;
Ripari, F ;
Scarano, A .
BIOCERAMICS, 2000, 192-1 :403-407
[10]   Effects of focal metallic implants on opposing cartilage - an in-vitro study with an abrasion test machine [J].
Diermeier, Theresa ;
Venjakob, Arne ;
Byrne, Kevin ;
Burgkart, Rainer ;
Foehr, Peter ;
Milz, Stefan ;
Imhoff, Andreas B. ;
Vogt, Stephan .