Diagnostic accuracy of apple watch ECG outputs in identifying dysrhythmias: A comparison with 12-Lead ECG in emergency department

被引:2
|
作者
Pasli, Sinan [1 ,4 ]
Topcuoglu, Hazal [2 ]
Yilmaz, Mutlu [1 ]
Yadigaroglu, Metin [3 ]
Imamoglu, Melih [1 ]
Karaca, Yunus [1 ]
机构
[1] Karadeniz Tech Univ, Fac Med, Dept Emergency Med, Trabzon, Turkiye
[2] Siirt Educ & Res Hosp, Dept Emergency Med, Siirt, Turkiye
[3] Samsun Univ, Fac Med, Dept Emergency Med, Samsun, Turkiye
[4] Karadeniz Tech Univ, Dept Emergency Med, TR-61080 Trabzon, Turkiye
来源
AMERICAN JOURNAL OF EMERGENCY MEDICINE | 2024年 / 79卷
关键词
Apple watch; ECG; Smartwatch; Photoplethysmography; Wearable device; Dysrhythmia; WEARABLE DEVICES; HEART-RATE; VALIDATION;
D O I
10.1016/j.ajem.2024.01.046
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Wearable devices, particularly smartwatches like the Apple Watch (AW), can record important cardiac information, such as single-lead electrocardiograms (ECGs). Although they are increasingly used to detect conditions such as atrial fibrillation (AF), research on their effectiveness in detecting a wider range of dysrhythmias and abnormal ECG findings remains limited. The primary objective of this study is to evaluate the accuracy of the AW in detecting various cardiac rhythms by comparing it with standard ECG's lead-I. Methods: This single -center prospective observational study was conducted in a tertiary care emergency department (ED) between 1.10.2023 and 31.10.2023. The study population consisted of all patients assessed in the critical care areas of the ED, all of whom underwent standard 12-lead ECGs for various clinical reasons. Participants in the study were included consecutively. An AW was attached to patients' wrists and an ECG lead-I printout was obtained. Heart rate, rhythm and abnormal findings were evaluated and compared with the lead-I of standard ECG. Two emergency medicine specialists performed the ECG evaluations. Rhythms were categorized as normal sinus rhythm and abnormal rhythms, while ECG findings were categorized as the presence or absence of abnormal findings. AW and 12-lead ECG outputs were compared using the McNemar test. Predictive performance analyses were also performed for subgroups. Bland -Altman analysis using absolute mean differences and concordance correlation coefficients was used to assess the level of heart rate agreement between devices. Results: The study was carried out on 721 patients. When analyzing ECG rhythms and abnormal findings in lead-I, the effectiveness of AW in distinguishing between normal and abnormal rhythms was similar to standard ECGs (p = 0.52). However, there was a significant difference between AW and standard ECGs in identifying abnormal findings in lead-I (p < 0.05). Using Bland -Altman analysis for heart rate assessment, the absolute mean difference for heart rate was 0.81 +/- 6.12 bpm (r = 0.94). There was strong agreement in 658 out of 700 (94%) heart rate measurements. Conclusion: Our study indicates that the AW has the potential to detect cardiac rhythms beyond AF. ECG tracings obtained from the AW may help evaluate cardiac rhythms prior to the patient's arrival in the ED. However, further research with a larger patient cohort is essential, especially for specific diagnoses. (c) 2024 Elsevier Inc. All rights reserved.
引用
收藏
页码:25 / 32
页数:8
相关论文
共 50 条
  • [1] A comparison of manual electrocardiographic interval and waveform analysis in lead 1 of 12-lead ECG and Apple Watch ECG: A validation study
    Saghir, Nabeel
    Aggarwal, Arjun
    Soneji, Nisha
    Valencia, Victoria
    Rodgers, George
    Kurian, Thomas
    CARDIOVASCULAR DIGITAL HEALTH JOURNAL, 2020, 1 (01): : 30 - 36
  • [2] ACCURACY OF 22-LEAD ECG ANALYSIS FOR DIAGNOSIS OF ACUTE MYOCARDIAL-INFARCTION AND CORONARY-ARTERY DISEASE IN THE EMERGENCY DEPARTMENT - A COMPARISON WITH 12-LEAD ECG
    JUSTIS, DL
    HESSION, WT
    ANNALS OF EMERGENCY MEDICINE, 1992, 21 (01) : 1 - 9
  • [3] COMPARISON OF INTRACORONARY ECG WITH STANDARD 12-LEAD ECG IN PTCA
    TATAMI, R
    YASUHARA, S
    YAMAMURA, M
    SHIBATA, K
    ISHISE, S
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1987, 51 (07): : 812 - 813
  • [4] Prehospital 12-lead ECG diagnostic programs
    Urban, MJ
    Edmondson, DA
    Aufderheide, TP
    EMERGENCY MEDICINE CLINICS OF NORTH AMERICA, 2002, 20 (04) : 825 - +
  • [5] APPLE WATCH SERIES 4 VS. 12-LEAD ECG: A COMPARISON OF MANUAL ELECTROCARDIOGRAPHIC WAVEFORM ANALYSIS
    Saghir, Nabeel
    Aggarwal, Arjun
    Soneji, Nishita
    Valencia, Victoria
    Rodgers, George
    Kurian, Thomas
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2020, 75 (11) : 378 - 378
  • [6] Evaluation of the Accuracy of ECG Captured by CardioChip Through Comparison of ECG Recording to a Standard 12-Lead ECG Recording Device
    Zou, Rui
    Hung, Chung-Lieh
    Lo, Chi Ln
    Yang, Catherine
    Zhang, Wenfeng
    PROCEEDINGS 2016 IEEE 25TH INTERNATIONAL SYMPOSIUM ON INDUSTRIAL ELECTRONICS (ISIE), 2016, : 1094 - 1098
  • [7] 12-Lead ECG interpretation by database comparison
    Gregg, Richard E.
    Smith, Stephen W.
    Babaeizadeh, Saeed
    JOURNAL OF ELECTROCARDIOLOGY, 2019, 57 : S79 - S85
  • [8] COMPARISON OF DIGITAL 12-LEAD ECG AND DIGITAL 12-LEAD HOLTER ECG RECORDINGS IN HEALTHY MALE SUBJECTS
    Taubel, J.
    Naseem, A.
    Shakeri-Nejad, K.
    Dingemanse, J.
    Ferber, G.
    Camm, A. J.
    CLINICAL PHARMACOLOGY & THERAPEUTICS, 2011, 89 : S11 - S11
  • [9] Comparison between a 6-lead smartphone ECG and 12-lead ECG in athletes
    Orchard, Jessica J.
    Orchard, John W.
    Raju, Hariharan
    La Gerche, Andre
    Puranik, Rajesh
    Semsarian, Chris
    JOURNAL OF ELECTROCARDIOLOGY, 2021, 66 : 95 - 97
  • [10] Automated serial ECG comparison improves computerized interpretation of 12-lead ECG
    Gregg, Richard E.
    Deluca, Daniel C.
    Chien, Cheng-hao Simon
    Helfenbein, Eric D.
    Ariet, Mario
    JOURNAL OF ELECTROCARDIOLOGY, 2012, 45 (06) : 561 - 565