Optimization of power feedback control system for HCN interferometer on EAST Tokamak

被引:1
|
作者
Zhang, J. B. [1 ]
Liu, H. Q. [1 ]
Zhang, Y. [1 ]
Jie, Y. . X. [1 ]
Wei, X. C. [1 ]
Xie, J. X. [1 ,2 ]
Wang, S. X. [1 ]
Lian, H. [1 ]
机构
[1] Chinese Acad Sci, Inst Plasma Phys, 350 ShuShanHu Rd, Hefei 230031, Peoples R China
[2] Univ Sci & Technol China, 96 JinZhai Rd, Hefei 230026, Peoples R China
来源
JOURNAL OF INSTRUMENTATION | 2023年 / 18卷 / 11期
关键词
Nuclear instruments and methods for hot plasma diagnostics; Plasma diagnostics-interferometry; spectroscopy and imaging; INFRARED LASER INTERFEROMETER; DIAGNOSTICS;
D O I
10.1088/1748-0221/18/11/C11016
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Hydrogen Cyanide (HCN) interferometer stands as an indispensable diagnostic tool, designated to measure the line-integrated electron density on the Experimental Advanced Superconducting Tokamak (EAST), thereby offering essential density feedback signals for EAST operation. The HCN laser used in the interferometer is a continuous glow discharge gas laser. However, owing to variations in external ambient temperature, the HCN laser exhibits output power instability. The current power automatic feedback control system for the HCN laser interferometer exhibits low actuator adjustment precision, thereby limiting its ability to achieve maximum output power. Moreover, the system may cause a temporary power drop to zero during adjustment, while EAST operation requires the HCN laser interferometer to maintain high power continuously. In order to solve the issues, this paper developed an innovative actuator that ingeniously amalgamates the virtues of piezoelectric ceramic and stepper motor. This novel actuator, demonstrating a broad adjustment range and high precision, is intended to replace the present one in the HCN interferometer power control system. While the stepper motor executes coarse adjustment, the piezoelectric ceramic enables precise adjustment, thus enhancing the adjustment accuracy of the automatic power control system to a sub-micrometer level and an adjustment range greater than 5 mm. Accurate determination of the regulator's adjustment direction can significantly improve the efficiency of adjustment. The slope of the laser output power peak decreases as it approaches the maximum value, and increases as it moves away from the maximum value. Therefore, this system replaces the original threshold algorithm with a slope algorithm. By correctly judging the adjustment direction, the efficiency of adjustment is improved. Furthermore, this algorithm can keep the output power stable on a power peak. There will be no instances of zero power output during the adjustment, allowing the power to remain stable above the set threshold for a long time. Currently, this system has been successfully applied to the HCN interferometer, and testing has found that the system can maintain stability for at least 16 hours. Furthermore, even when the laser output power was intentionally reduced to below the set threshold, the system was able to respond quickly and adjust the output, rapidly restoring it to above the predefined threshold. Multiple tests have shown the strong robustness of the system, indicating it meets the stringent demands of the EAST's complex environment.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Optimization of the POlarimeter-INTerferometer diagnostics system on EAST
    Wang, Y. F.
    Liu, H. Q.
    Jie, Y. X.
    Lian, H.
    Chu, Y. Q.
    Li, W. M.
    Wang, S. X.
    Zhu, X.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (06)
  • [22] Optimization of Frequency Feedback Control Strategy for Magnet Power Supply of EAST Fishtail Divertor
    Zhou, Yu
    Wang, Yuchen
    Guan, Rui
    Sun, Haozhang
    Huang, Yiyun
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2024, 52 (02) : 514 - 520
  • [23] Conceptual design of Remote Control System for EAST tokamak
    Sun, X. Y.
    Wang, F.
    Wang, Y.
    Li, S.
    FUSION ENGINEERING AND DESIGN, 2014, 89 (05) : 741 - 744
  • [24] MULTICHANNEL HCN INTERFEROMETER FOR ELECTRON-DENSITY PROFILE MEASUREMENTS OF TOKAMAK PLASMAS
    VERON, D
    CERTAIN, J
    CRENN, JP
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1977, 67 (07) : 964 - 967
  • [25] The 7-channel FIR HCN interferometer on J-TEXT tokamak
    Chen, W.
    Gao, L.
    Chen, J.
    Li, Q.
    Wang, Z. J.
    Zhuang, G.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [26] ELECTRON-DENSITY MEASUREMENT OF A TOKAMAK PLASMA BY AN HCN-LASER INTERFEROMETER
    NISHIZAWA, A
    MASUZAKI, M
    MOHRI, A
    JAPANESE JOURNAL OF APPLIED PHYSICS, 1976, 15 (09) : 1753 - 1760
  • [27] Control System for Dispersion Interferometer on Globus-M2 Tokamak
    Kovalenko, Yu. V.
    Zubarev, P. V.
    Ivanenko, S. V.
    Solomakhin, A. L.
    Puriga, E. A.
    Bagryansky, P. A.
    PHYSICS OF ATOMIC NUCLEI, 2024, 87 (SUPPL1) : S159 - S165
  • [28] Study of a DCN laser interferometer/polarimeter for EAST tokamak
    Liu, H. Q.
    Gao, X.
    Jie, Y. X.
    Xu, Q.
    Ming, T. F.
    Shi, N.
    Cheng, Y. F.
    Tong, X. D.
    2007 JOINT 32ND INTERNATIONAL CONFERENCE ON INFRARED AND MILLIMETER WAVES AND 15TH INTERNATIONAL CONFERENCE ON TERAHERTZ ELECTRONICS, VOLS 1 AND 2, 2007, : 374 - 375
  • [29] Bench Test of the Vibration Compensation Interferometer for EAST Tokamak
    李恭顺
    杨曜
    刘海庆
    揭银先
    邹志勇
    王正兴
    曾龙
    魏学朝
    李维明
    兰婷
    朱翔
    刘煜锴
    高翔
    Plasma Science and Technology, 2016, 18 (02) : 206 - 210
  • [30] Bench Test of the Vibration Compensation Interferometer for EAST Tokamak
    Li Gongshun
    Yang Yao
    Liu Haiqing
    Jie Yinxian
    Zou Zhiyong
    Wang Zhengxing
    Zeng Long
    Wei Xuechao
    Li Weiming
    Lan Ting
    Zhu Xiang
    Liu Yukai
    Gao Xiang
    PLASMA SCIENCE & TECHNOLOGY, 2016, 18 (02) : 206 - 210