Feasibility and efficacy of a novel size adjustable cryoballoon for ablation of atrial fibrillation

被引:10
作者
Frommeyer, Gerrit [1 ]
Ellermann, Christian [1 ]
Wolfes, Julian [1 ]
Lange, Philipp S. [1 ]
Guener, Fatih [1 ]
Eckardt, Lars [1 ]
机构
[1] Univ Hosp Munster, Dept Cardiol Electrophysiol 2, Albert Schweitzer Campus 1,Gebaude A1, D-48149 Munster, Germany
关键词
Pulmonary vein isolation; Cryoballoon; Atrial fibrillation;
D O I
10.1007/s10840-023-01645-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of the present case series was to characterize the feasibility of a novel size adjustable cryoballoon system (PolarX Fit, Boston Scientific, Marlborough, MA, USA). This cryoballoon catheter can be inflated to two different diameters (28 mm and 31 mm) within the same procedure allowing vein adapted PVI. In summary, the novel size adjustable cryoballoon shows similar characteristics as the established versions. The intraprocedural flexibility of balloon size led to employment of the larger variant in the majority of freeze applications. Of note, in all but one procedure, both sizes were employed to ensure optimal occlusion for all veins. This initial series suggests that the size adjustable balloon offers more flexibility of obtain optimal occlusions in particular, in challenging anatomies, including common pulmonary vein ostia.
引用
收藏
页码:253 / 255
页数:3
相关论文
共 6 条
[1]   Novel cryoballoon ablation system for pulmonary vein isolation: multicenter assessment of efficacy and safety-ANTARCTICA study [J].
Heeger, Christian H. ;
Pott, Alexander ;
Sohns, Christian ;
Riesinger, Lisa ;
Sommer, Philipp ;
Gasperetti, Alessio ;
Tondo, Claudio ;
Fassini, Gaetano ;
Moser, Fabian ;
Lucas, Philipp ;
Weinmann, Karolina ;
Bohnen, Jan Eric ;
Dahme, Tillman ;
Rillig, Andreas ;
Kuck, Karl Heinz ;
Wakili, Reza ;
Metzner, Andreas ;
Tilz, Roland R. .
EUROPACE, 2022, 24 (12) :1917-1925
[2]  
Hindricks G, 2021, EUR HEART J, V42, P546, DOI [10.1093/eurheartj/ehaa945, 10.1093/eurheartj/ehaa612]
[3]   Cryoballoon or radiofrequency ablation for symptomatic paroxysmal atrial fibrillation: reintervention, rehospitalization, and quality-of-life outcomes in the FIRE AND ICE trial [J].
Kuck, Karl-Heinz ;
Fuernkranz, Alexander ;
Chun, K. R. Julian ;
Metzner, Andreas ;
Ouyang, Feifan ;
Schlueter, Michael ;
Elvan, Arif ;
Lim, Hae W. ;
Kueffer, Fred J. ;
Arentz, Thomas ;
Albenque, Jean-Paul ;
Tondo, Claudio ;
Kuehne, Michael ;
Sticherling, Christian ;
Brugada, Josep .
EUROPEAN HEART JOURNAL, 2016, 37 (38) :2858-2865
[4]   Comparing learning curves of two established "single-shot" devices for ablation of atrial fibrillation [J].
Leitz, Patrick ;
Moennig, G. ;
Guener, F. ;
Dechering, D. G. ;
Wasmer, K. ;
Reinke, F. ;
Lange, Philipp S. ;
Eckardt, L. ;
Frommeyer, G. .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2018, 53 (03) :317-322
[5]   Acute procedural efficacy and safety of a novel cryoballoon for the treatment of paroxysmal atrial fibrillation: Results from the POLAR ICE study [J].
Martin, Claire A. A. ;
Tilz, Roland R. R. ;
Anic, Ante ;
Defaye, Pascal ;
Luik, Armin ;
de Asmundis, Carlo ;
Champ-Rigot, Laure ;
Iacopino, Saverio ;
Sommer, Philipp ;
Albrecht, Elizabeth M. M. ;
Raybuck, Jonathan D. D. ;
Richards, Elizabeth ;
Cielen, Nele ;
Yap, Sing-Chien .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (04) :833-840
[6]   Novel Cryoballoon Ablation System for Single Shot Pulmonary Vein Isolation - The Prospective ICE-AGE-X Study - [J].
Tilz, Roland Richard ;
Meyer-Saraei, Roza ;
Eitel, Charlotte ;
Fink, Thomas ;
Sciacca, Vanessa ;
Lopez, Lisbeth Delgado ;
Kirstein, Bettina ;
Schluter, Michael ;
Vogler, Julia ;
Kuck, Karl-Heinz ;
Heeger, Christian-H .
CIRCULATION JOURNAL, 2021, 85 (08) :1296-1304