MRI compatibility study of a prototype radiofrequency penetrable oval PET insert at 3 T

被引:0
作者
Akram, Md Shahadat Hossain [1 ]
Nishikido, Fumihiko [1 ]
Levin, Craig S. [2 ]
Takyu, Sodai [1 ]
Obata, Takayuki [3 ]
Yamaya, Taiga [1 ]
机构
[1] Natl Inst Quantum Sci & Technol QST, Inst Quantum Med Sci, Dept Adv Nucl Med Sci, 4-9-1 Anagawa, Inage, Chiba 2638555, Japan
[2] Stanford Univ, Sch Med, Dept Radiol, Stanford, CA 94305 USA
[3] Natl Inst Quantum Sci & Technol QST, Inst Quantum Med Sci, Dept Appl MRI Res, 4-9-1 Anagawa, Inage, Chiba 2638555, Japan
关键词
Oval PET; PET/MRI; PET insert; RF floating; RF-penetrable PET; NOISE RATIOS; PET/MRI; DETECTORS; FEASIBILITY; PERFORMANCE; SCANNER; SYSTEM; COIL;
D O I
10.1007/s11604-023-01514-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo perform an MRI compatibility study of an RF field-penetrable oval-shaped PET insert that implements an MRI built-in body RF coil both as a transmitter and a receiver.MethodsTwelve electrically floating RF shielded PET detector modules were used to construct the prototype oval PET insert with a major axis of 440 mm, a minor axis of 350 mm, and an axial length of 225 mm. The electric floating of the PET detector modules was accomplished by isolating the cable shield from the detector shield using plastic tape. Studies were conducted on the transmit (B1) RF field, the image signal-to-noise ratio (SNR), and the RF pulse amplitude for a homogeneous cylindrical (diameter: 160 mm and length: 260 mm) phantom (NaCl + NiSO4 solution) in a 3 T clinical MRI system (Verio, Siemens, Erlangen, Germany).ResultsThe B1 maps for the oval insert were similar to the MRI-only field responses. Compared to the MRI-only values, SNR reductions of 51%, 45%, and 59% were seen, respectively, for the spin echo (SE), gradient echo (GE), and echo planar (EPI) images for the case of oval PET insert. Moreover, the required RF pulse amplitudes for the SE, GE, and EPI sequences were, respectively, 1.93, 1.85, and 1.36 times larger. However, a 30% reduction in the average RF reception sensitivity was observed for the oval insert.ConclusionsThe prototype floating PET insert was a safety concern for the clinical MRI system, and this compatibility study provided clearance for developing a large body size floating PET insert for the existing MRI system. Because of the RF shield of the insert, relatively large RF powers compared to the MRI-only case were required. Because of this and also due to low RF sensitivity of the body coil, the SNRs reduced largely.
引用
收藏
页码:382 / 390
页数:9
相关论文
共 43 条
[1]   Hybrid PET/MR Imaging and Brain Connectivity [J].
Aiello, Marco ;
Cavaliere, Carlo ;
Salvatore, Marco .
FRONTIERS IN NEUROSCIENCE, 2016, 10
[2]   Toward simultaneous PET/MR breast imaging: Systematic evaluation and integration of a radiofrequency breast coil [J].
Aklan, Bassim ;
Paulus, Daniel H. ;
Wenkel, Evelyn ;
Braun, Harald ;
Navalpakkam, Bharath K. ;
Ziegler, Susanne ;
Geppert, Christian ;
Sigmund, Eric E. ;
Melsaether, Amy ;
Quick, Harald H. .
MEDICAL PHYSICS, 2013, 40 (02)
[3]   Study on the radiofrequency transparency of partial-ring oval-shaped prototype PET inserts in a 3 T clinical MRI system [J].
Akram, Md Shahadat Hossain ;
Levin, Craig S. ;
Nishikido, Fumihiko ;
Takyu, Sodai ;
Obata, Takayuki ;
Yamaya, Taiga .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2024, 17 (01) :60-70
[4]   Study on the radiofrequency transparency of electrically floating and ground PET inserts in a 3 T clinical MRI system [J].
Akram, Md Shahadat Hossain ;
Obata, Takayuki ;
Nishikido, Fumihiko ;
Yamaya, Taiga .
MEDICAL PHYSICS, 2022, 49 (05) :2965-2978
[5]   Geometry optimization of electrically floating PET inserts for improved RF penetration for a 3T MRI system [J].
Akram, Md Shahadat Hossain ;
Levin, Craig S. ;
Obata, Takayuki ;
Hirumi, Genki ;
Yamaya, Taiga .
MEDICAL PHYSICS, 2018, 45 (10) :4627-4641
[6]   MRI compatibility study of an integrated PET/RF-coil prototype system at 3T [J].
Akram, Md Shahadat Hossain ;
Obata, Takayuki ;
Suga, Mikio ;
Nishikido, Fumihiko ;
Yoshida, Eiji ;
Saito, Kazuyuki ;
Yamaya, Taiga .
JOURNAL OF MAGNETIC RESONANCE, 2017, 283 :62-70
[7]  
Akram MSH, PROTOTYPE OVAL PET I
[8]   Performance Study of a Radio-Frequency Field-Penetrable PET Insert for Simultaneous PET/MRI [J].
Chang, Chen-Ming ;
Lee, Brian J. ;
Grant, Alexander M. ;
Groll, Andrew N. ;
Levin, Craig S. .
IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2018, 2 (05) :422-431
[9]   Calculation of Radiofrequency Electromagnetic Fields and Their Effects in MRI of Human Subjects [J].
Collins, Christopher M. ;
Wang, Zhangwei .
MAGNETIC RESONANCE IN MEDICINE, 2011, 65 (05) :1470-1482
[10]   A method for accurate calculation of B-1 fields in three dimensions. Effects of shield geometry on field strength and homogeneity in the birdcage coil [J].
Collins, CM ;
Li, SZ ;
Yang, QX ;
Smith, MB .
JOURNAL OF MAGNETIC RESONANCE, 1997, 125 (02) :233-241