Evaluation of a new predictive equation for automated calculation of size-specific dose estimate (SSDE) in CT imaging

被引:2
作者
Osman, N. D. [1 ]
Abdulkadir, M. K. [1 ,2 ,3 ]
Shuaib, I. L. [1 ]
Nasirudin, R. A. [2 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Kepala Batas 13200, Penang, Malaysia
[2] Univ Ilorin, Fac Basic Clin Sci, Dept Radiog, Ilorin 240213, Nigeria
[3] Univ Ilorin, Fac Basic Clin Sci, Dept Radiog, Ilorin 1515, Kwara State, Nigeria
关键词
Computed tomography; Size-specific dose; Software; Effective diameter; Dose optimisation; DIAMETER;
D O I
10.1016/j.radi.2023.11.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: The adoption of size-specific dose estimate (SSDE) in clinical practice is still limited owing to the tedious and complex manual measurement of individual patient size for the clinical calculation of SSDE. Thus, the automation of SSDE is imperative. This study aims to evaluate a predictive equation for the automated calculation of SSDE. Methods: A user-friendly software was developed to accurately predict the individual size-specific dose estimation of paediatric patients undergoing computed tomography (CT) scans of the head, thorax, and abdomen. The software includes a calculation equation developed based on a novel SSDE prediction equation that used a population's pre-determined percentage difference between volume-weighted computed tomography dose index (CTDIvol) and SSDE with age. American Association of Physicists in Medicine (AAPM RPT 204) method (manual) and segmentation-based SSDE calculators (indoseCT and XXautocalc) were used to assess the proposed software predictions comparatively. Results: The results of this study show that the automated equation-based calculation of SSDE and the manual and segmentation-based calculation of SSDE are in good agreement for patients. The differences between the automated equation-based calculation of SSDE and the manual and segmentation-based calculation are less than 3%. Conclusion: This study validated an accurate SSDE calculator that allows users to enter key input values and calculate SSDE. Implication for practice: The automated equation-based SSDE software (PESSD) seems a promising tool for estimating individualised CT doses during CT scans. (c) 2023 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:237 / 244
页数:8
相关论文
共 48 条
[31]   Variations in size-specific effective dose with patient stature and beam width for kV cone beam CT imaging in radiotherapy [J].
Martin, C. J. ;
Abuhaimed, A. .
JOURNAL OF RADIOLOGICAL PROTECTION, 2022, 42 (03)
[32]   A SIMPLE METHOD FOR CALIBRATING PIXEL VALUES OF THE CT LOCALIZER RADIOGRAPH FOR CALCULATING WATER-EQUIVALENT DIAMETER AND SIZE-SPECIFIC DOSE ESTIMATE [J].
Anam, Choirul ;
Fujibuchi, Toshioh ;
Toyoda, Takatoshi ;
Sato, Naoki ;
Haryanto, Freddy ;
Widita, Rena ;
Arif, Idam ;
Dougherty, Geoff .
RADIATION PROTECTION DOSIMETRY, 2018, 179 (02) :158-168
[33]   LOCAL DRLS FOR PAEDIATRIC CT EXAMINATIONS BASED ON SIZE-SPECIFIC DOSE ESTIMATES IN KERMANSHAH, IRAN [J].
Mohammadbeigi, Ahmad ;
Khoshgard, Karim ;
Haghparast, Abbas ;
Eivazi, Mohammad Taghi .
RADIATION PROTECTION DOSIMETRY, 2019, 186 (04) :496-506
[34]   RADIATION DOSE DETERMINATION IN ABDOMINAL CT EXAMINATIONS OF CHILDREN AT SUDANESE HOSPITALS USING SIZE-SPECIFIC DOSE ESTIMATES [J].
Bashier, Einas H. ;
Suliman, I. I. .
RADIATION PROTECTION DOSIMETRY, 2019, 183 (04) :444-449
[35]   Radiation dose in chest CT: Assessment of size-specific dose estimates based on water-equivalent correction [J].
Gabusi, M. ;
Riccardi, L. ;
Aliberti, C. ;
Vio, S. ;
Paiusco, M. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (02) :393-397
[36]   Optimizing Radiation Dose in High-Resolution Chest CT: The Impact of Patient-Specific Factors and Size-Specific Dose Estimates [J].
Abuzaid, Mohamed .
DIAGNOSTICS, 2025, 15 (06)
[37]   Size-specific dose estimations for pediatric chest, abdomen/pelvis and head CT scans with the use of GATE [J].
Kostou, Theodora ;
Papadimitroulas, Panagiotis ;
Papaconstadopoulos, Pavlos ;
Devic, Slobodan ;
Seuntjens, Jan ;
Kagadis, George C. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 :181-190
[38]   Influence of 320-detector-row volume scanning and AAPM report 111 CT dosimetry metrics on size-specific dose estimate: a Monte Carlo study [J].
Haba, Tomonobu ;
Koyama, Shuji ;
Kinomura, Yutaka ;
Ida, Yoshihiro ;
Kobayashi, Masanao .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (03) :697-703
[39]   Estimating Patient Dose from X-ray Tube Output Metrics: Automated Measurement of Patient Size from CT Images Enables Large-scale Size-specific Dose Estimates [J].
Ikuta, Ichiro ;
Warden, Graham I. ;
Andriole, Katherine P. ;
Khorasani, Ramin ;
Sodickson, Aaron .
RADIOLOGY, 2014, 270 (02) :472-480
[40]   Size-specific dose estimates for pediatric head CT protocols based on the AAPM report TG-293 [J].
Sapignoli, Sonia ;
Roggio, Antonella ;
Boschini, Alessio ;
Guida, Federica ;
Merlo, Carlo ;
Paiusco, Marta ;
Zorz, Alessandra ;
De Monte, Francesca .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 100 :26-30