Evaluation of a new predictive equation for automated calculation of size-specific dose estimate (SSDE) in CT imaging

被引:2
|
作者
Osman, N. D. [1 ]
Abdulkadir, M. K. [1 ,2 ,3 ]
Shuaib, I. L. [1 ]
Nasirudin, R. A. [2 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Kepala Batas 13200, Penang, Malaysia
[2] Univ Ilorin, Fac Basic Clin Sci, Dept Radiog, Ilorin 240213, Nigeria
[3] Univ Ilorin, Fac Basic Clin Sci, Dept Radiog, Ilorin 1515, Kwara State, Nigeria
关键词
Computed tomography; Size-specific dose; Software; Effective diameter; Dose optimisation; DIAMETER;
D O I
10.1016/j.radi.2023.11.012
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: The adoption of size-specific dose estimate (SSDE) in clinical practice is still limited owing to the tedious and complex manual measurement of individual patient size for the clinical calculation of SSDE. Thus, the automation of SSDE is imperative. This study aims to evaluate a predictive equation for the automated calculation of SSDE. Methods: A user-friendly software was developed to accurately predict the individual size-specific dose estimation of paediatric patients undergoing computed tomography (CT) scans of the head, thorax, and abdomen. The software includes a calculation equation developed based on a novel SSDE prediction equation that used a population's pre-determined percentage difference between volume-weighted computed tomography dose index (CTDIvol) and SSDE with age. American Association of Physicists in Medicine (AAPM RPT 204) method (manual) and segmentation-based SSDE calculators (indoseCT and XXautocalc) were used to assess the proposed software predictions comparatively. Results: The results of this study show that the automated equation-based calculation of SSDE and the manual and segmentation-based calculation of SSDE are in good agreement for patients. The differences between the automated equation-based calculation of SSDE and the manual and segmentation-based calculation are less than 3%. Conclusion: This study validated an accurate SSDE calculator that allows users to enter key input values and calculate SSDE. Implication for practice: The automated equation-based SSDE software (PESSD) seems a promising tool for estimating individualised CT doses during CT scans. (c) 2023 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:237 / 244
页数:8
相关论文
共 46 条
  • [1] The impact of patient table on size-specific dose estimate (SSDE)
    Anam, Choirul
    Haryanto, Freddy
    Widita, Rena
    Arif, Idam
    Dougherty, Geoff
    McLean, Donald
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 153 - 158
  • [2] Size-Specific Dose Estimate (SSDE) Provides a Simple Method to Calculate Organ Dose for Pediatric CT Examinations
    Moore, B.
    Brady, S.
    Mirro, A.
    Kaufman, R.
    MEDICAL PHYSICS, 2014, 41 (06) : 424 - +
  • [3] A comparison study of size-specific dose estimate calculation methods
    Parikh, Roshni A.
    Wien, Michael A.
    Novak, Ronald D.
    Jordan, David W.
    Klahr, Paul
    Soriano, Stephanie
    Ciancibello, Leslie
    Berlin, Sheila C.
    PEDIATRIC RADIOLOGY, 2018, 48 (01) : 56 - 65
  • [4] Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE)
    Kidoh, Masafumi
    Utsunomiya, Daisuke
    Oda, Seitaro
    Nakaura, Takeshi
    Funama, Yoshinori
    Yuki, Hideaki
    Hirata, Kenichiro
    Namimoto, Tomohiro
    Sakabe, Daisuke
    Hatemura, Masahiro
    Yamashita, Yasuyuki
    EUROPEAN RADIOLOGY, 2017, 27 (06) : 2267 - 2274
  • [5] A comparison study of size-specific dose estimate calculation methods
    Roshni A. Parikh
    Michael A. Wien
    Ronald D. Novak
    David W. Jordan
    Paul Klahr
    Stephanie Soriano
    Leslie Ciancibello
    Sheila C. Berlin
    Pediatric Radiology, 2018, 48 : 56 - 65
  • [6] The role of Size-Specific Dose Estimate (SSDE) in patient-specific organ dose and cancer risk estimation in paediatric chest and abdominopelvic CT examinations
    Franck, Caro
    Vandevoorde, Charlot
    Goethals, Ingeborg
    Smeets, Peter
    Achten, Eric
    Verstraete, Koenraad
    Thierens, Hubert
    Bacher, Klaus
    EUROPEAN RADIOLOGY, 2016, 26 (08) : 2646 - 2655
  • [7] Automated size-specific CT dose monitoring program: Assessing variability in CT dose
    Christianson, Olav
    Li, Xiang
    Frush, Donald
    Samei, Ehsan
    MEDICAL PHYSICS, 2012, 39 (11) : 7131 - 7139
  • [8] Consideration of the usefulness of a size-specific dose estimate in pediatric CT examination
    Tsujiguchi, Takakiyo
    Obara, Hideki
    Ono, Shuichi
    Saito, Yoko
    Kashiwakura, Ikuo
    JOURNAL OF RADIATION RESEARCH, 2018, 59 (04) : 430 - 435
  • [9] THE SIZE-SPECIFIC DOSE ESTIMATE OF PAEDIATRIC HEAD CT EXAMINATIONS FOR VARIOUS PROTOCOLS
    Fahmi, Arif
    Anam, Choirul
    Ali, Mohd Hanafi
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (04) : 522 - 528
  • [10] ACCURACY OF SIZE-SPECIFIC DOSE ESTIMATE CALCULATION FROM CENTER SLICE IN COMPUTED TOMOGRAPHY
    Boos, Johannes
    Kroepil, Patric
    Bethge, Oliver Th.
    Aissa, Joel
    Schleich, Christoph
    Sawicki, Lino Morris
    Heinzler, Niklas
    Antoch, Gerald
    Thomas, Christoph
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (01) : 8 - 19