Reduced Iodinated Contrast Media Administration in Coronary CT Angiography on a Clinical Photon-Counting Detector CT System A Phantom Study Using a Dynamic Circulation Model

被引:39
作者
Emrich, Tilman [1 ,2 ,3 ]
O'Doherty, Jim [4 ]
Schoepf, U. Joseph [1 ,6 ]
Suranyi, Pal [1 ]
Aquino, Gilberto [1 ]
Kloeckner, Roman [2 ]
Halfmann, Moritz C. [2 ,3 ]
Allmendinger, Thomas [5 ]
Schmidt, Bernhard [5 ]
Flohr, Thomas [5 ]
Varga-Szemes, Akos [1 ]
机构
[1] Med Univ South Carolina, Dept Radiol & Radiol Sci, Div Cardiovasc Imaging, Charleston, SC USA
[2] Johannes Gutenberg Univ Mainz, Univ Med Ctr, Dept Diagnost & Intervent Radiol, Mainz, Germany
[3] German Ctr Cardiovasc Res, Partner site Rhine Main, Mainz, Germany
[4] Siemens Med Solut USA Inc, Malvern, PA USA
[5] Siemens Healthineers, Forchheim, Germany
[6] Med Univ SouthCarolina, Dept Radiol & Radiol Sci, Div Cardiovasc Imaging, Ashley River Tower, 25 Courtenay Dr,MSC 226, Charleston, SC 29425 USA
关键词
coronary CT angiography; spectral imaging; contrast agent; photon counting detector; COMPUTED-TOMOGRAPHY ANGIOGRAPHY; LOW TUBE VOLTAGE; DUAL-SOURCE; INJECTION PROTOCOL; IN-VITRO; RECONSTRUCTION; LOAD; OPTIMIZATION;
D O I
10.1097/RLI.0000000000000911
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe aim of this study was to evaluate strategies to reduce contrast media volumes for coronary computed tomography (CT) angiography on a clinical first-generation dual-source photon-counting detector (PCD)-CT system using a dynamic circulation phantom.Materials and MethodsCoronary CT angiograph is an established method for the assessment of coronary artery disease that relies on the administration of iodinated contrast media. Reduction of contrast media volumes while maintaining diagnostic image quality is desirable. In this study, a dynamic phantom containing a 3-dimensional-printed model of the thoracic aorta and coronary arteries was evaluated using a clinical contrast injection protocol with stepwise reduced contrast agent concentrations (100%, 75%, 50%, 40%, 30%, and 20% contrast media content of the same 50 mL bolus, resulting in iodine delivery rates of 1.5, 1.1, 0.7, 0.6, 0.4 and 0.3 gl/s) on a first-generation, dual-source PCD-CT. Polychromatic images (T3D) and virtual monoenergetic images were reconstructed in the range of 40 to 70 keV in 5-keV steps. Attenuation and noise were measured in the coronary arteries and background material and the contrast-to-noise ratio (CNR) were calculated. Attenuation of 350 HU and a CNR of the reference protocol at 70 keV were regarded as sufficient for simulation of diagnostic purposes. Vessel sharpness and noise power spectra were analyzed for the aforementioned reconstructions.ResultsThe standard clinical contrast protocol (bolus with 100% contrast) yielded diagnostic coronary artery attenuation for all tested reconstructions (>398 HU). A 50% reduction in contrast media concentration demonstrated sufficient attenuation of the coronary arteries at 40 to 55 keV (>366 HU). Virtual monoenergetic image reconstructions of 40 to 45 and 40 keV allowed satisfactory attenuation of the coronary arteries for contrast concentrations of 40% and 30% of the original protocol. A reduction of contrast agent concentration to 20% of the initial concentration provided insufficient attenuation in the target vessels for all reconstructions. The highest CNR was found for virtual monoenergetic reconstructions at 40 keV for all contrast media injection protocols, yielding a sufficient CNR at a 50% reduction of contrast agent concentration.ConclusionsUsing virtual monoenergetic image reconstructions at 40 keV on a dual-source PCD-CT system, contrast media concentration could be reduced by 50% to obtain diagnostic attenuation and objective image quality for coronary CT angiography in a dynamic vessel phantom. These initial feasibility study results have to be validated in clinical studies.
引用
收藏
页码:148 / 155
页数:8
相关论文
共 35 条
  • [1] Introduction of a dedicated circulation phantom for comprehensive in vitro analysis of intravascular contrast material application
    Behrendt, Florian F.
    Bruners, Philipp
    Kalafut, John
    Mahnken, Andreas H.
    Keil, Sebastian
    Plumhans, Cedric
    Das, Marco
    Stanzel, Sven
    Wildberger, Joachim E.
    Pfeffer, Jochen
    Guenther, Rolf W.
    Muehlenbruch, Georg
    [J]. INVESTIGATIVE RADIOLOGY, 2008, 43 (10) : 729 - 736
  • [2] Monochromatic image reconstruction by dual energy imaging allows half iodine load computed tomography coronary angiography
    Carrascosa, Patricia
    Leipsic, Jonathon A.
    Capunay, Carlos
    Deviggiano, Alejandro
    Vallejos, Javier
    Goldsmit, Alejandro
    Rodriguez-Granillo, Gaston A.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2015, 84 (10) : 1915 - 1920
  • [3] Optimizing Contrast Media Injection Protocols in Computed Tomography Angiography at Different Tube Voltages: Evaluation in a Circulation Phantom
    Caruso, Damiano
    Eid, Marwen
    Schoepf, U. Joseph
    De Santis, Domenico
    Varga-Szemes, Akos
    Mangold, Stefanie
    Canstein, Christian
    Lesslie, Virginia W.
    Fuller, Stephen R.
    Ball, B. Devon
    Laghi, Andrea
    De Cecco, Carlo N.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2017, 41 (05) : 804 - 810
  • [4] Contrast media injection protocol optimization for dual-energy coronary CT angiography: results from a circulation phantom
    De Santis, Domenico
    Caruso, Damiano
    Schoepf, U. Joseph
    Eid, Marwen
    Albrecht, Moritz H.
    Duguay, Taylor M.
    Varga-Szemes, Akos
    Laghi, Andrea
    De Cecco, Carlo N.
    [J]. EUROPEAN RADIOLOGY, 2018, 28 (08) : 3473 - 3481
  • [5] Spectral Optimization of Chest CT Angiography with Reduced Iodine Load: Experience in 80 Patients Evaluated with Dual-Source, Dual-Energy CT
    Delesalle, Marie-Aurelie
    Pontana, Francois
    Duhamel, Alain
    Faivre, Jean-Baptiste
    Flohr, Thomas
    Tacelli, Nunzia
    Remy, Jacques
    Remy-Jardin, Martine
    [J]. RADIOLOGY, 2013, 267 (01) : 256 - 266
  • [6] Coronary Computed Tomography Angiography-Based Calcium Scoring In Vitro and In Vivo Validation of a Novel Virtual Noniodine Reconstruction Algorithm on a Clinical, First-Generation Dual-Source Photon Counting-Detector System
    Emrich, Tilman
    Aquino, Gilberto
    Schoepf, U. Joseph
    Braun, Franziska M.
    Risch, Franka
    Bette, Stefanie J.
    Woznicki, Piotr
    Decker, Josua A.
    O'Doherty, Jim
    Brandt, Verena
    Allmendinger, Thomas
    Nowak, Tristan
    Schmidt, Bernhard
    Flohr, Thomas
    Kroencke, Thomas J.
    Scheurig-Muenkler, Christian
    Varga-Szemes, Akos
    Schwarz, Florian
    [J]. INVESTIGATIVE RADIOLOGY, 2022, 57 (08) : 536 - 543
  • [7] Soap-bubble visualization and quantitative analysis of 3D coronary magnetic resonance angiograms
    Etienne, A
    Botnar, RM
    van Muiswinkel, AMC
    Boesiger, P
    Manning, WJ
    Stuber, M
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (04) : 658 - 666
  • [8] High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose
    Euler, Andre
    Higashigaito, Kai
    Mergen, Victor
    Sartoretti, Thomas
    Zanini, Bettina
    Schmidt, Bernhard
    Flohr, Thomas G.
    Ulzheimer, Stefan
    Eberhard, Matthias
    Alkadhi, Hatem
    [J]. INVESTIGATIVE RADIOLOGY, 2022, 57 (02) : 115 - 121
  • [9] Hubbell JM, X-ray mass attenuation coefficients, NIST standard reference database 126
  • [10] Low-Contrast and Low-Radiation Dose Protocol in Cardiac Computed Tomography: Usefulness of Low Tube Voltage and Knowledge-Based Iterative Model Reconstruction Algorithm
    Iyama, Yuji
    Nakaura, Takeshi
    Yokoyama, Koichi
    Kidoh, Masafumi
    Harada, Kazunori
    Oda, Seitaro
    Tokuyasu, Shinichi
    Yamashita, Yasuyuki
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2016, 40 (06) : 941 - 947