Galvanic Deposition of Calcium Phosphate/Bioglass Composite Coating on AISI 316L

被引:2
作者
Zanca, Claudio [1 ]
Milazzo, Alessandro [1 ]
Campora, Simona [2 ]
Capuana, Elisa [1 ]
Pavia, Francesco Carfi [1 ,3 ]
Patella, Bernardo [1 ]
Lopresti, Francesco [1 ]
Brucato, Valerio [1 ]
La Carrubba, Vincenzo [1 ,4 ]
Inguanta, Rosalinda [1 ]
机构
[1] Univ Palermo, Dept Engn, Viale Sci, I-90128 Palermo, Italy
[2] Univ Palermo, Dept Biol Chem & Pharmaceut Sci & Technol, Viale Sci, Ed 16, I-90128 Palermo, Italy
[3] Consorzio Univ Caltanissetta, Corso Vittorio Emanuele 92, I-93100 Caltanissetta, Italy
[4] Univ Palermo, ATeN Ctr, Viale Sci, I-90133 Palermo, Italy
关键词
AISI; 316L; coating; corrosion; cytotoxicity; galvanic deposition; hydroxyapatite; orthopedic implant; BIOACTIVE GLASS COATINGS; METAL DISPLACEMENT DEPOSITION; BRUSHITE/HYDROXYAPATITE COATINGS; HYDROXYAPATITE; CORROSION; IMPLANTS; REDUCTION; BIOGLASS; ROUTE; EIS;
D O I
10.3390/coatings13061006
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Calcium phosphate/Bioglass composite coatings on AISI 316L were investigated with regard to their potential role as a beneficial coating for orthopedic implants. These coatings were realized by the galvanic co-deposition of calcium phosphate compounds and Bioglass particles. A different amount of Bioglass 45S5 was used to study its effect on the performance of the composite coatings. The morphology and chemical composition of the coatings were investigated before and after their aging in simulated body fluid. The coatings uniformly covered the AISI 316L substrate and consisted of a brushite and hydroxyapatite mixture. Both phases were detected using X-ray diffraction and Raman spectroscopy. Additionally, both analyses revealed that brushite is the primary phase. The presence of Bioglass was verified through energy-dispersive X-ray spectroscopy, which showed the presence of a silicon peak. During aging in simulated body fluid, the coating was subject to a dynamic equilibrium of dissolution/reprecipitation with total conversion in only the hydroxyapatite phase. Corrosion tests performed in simulated body fluid at different aging times revealed that the coatings made with 1 g/L of Bioglass performed best. These samples have a corrosion potential of -0.068V vs. Ag/AgCl and a corrosion current density of 8.87 x 10(-7) A/cm(2). These values are better than those measured for bare AISI 316L (-0.187 V vs. Ag/AgCl and 2.52 x 10(-6) A/cm(2), respectively) and remained superior to pure steel for all 21 days of aging. This behavior indicated the good protection of the coating against corrosion phenomena, which was further confirmed by the very low concentration of Ni ions (0.076 ppm) released in the aging solution after 21 days of immersion. Furthermore, the absence of cytotoxicity, verified through cell viability assays with MC3T3-E1 osteoblastic cells, proves the biocompatibility of the coatings.
引用
收藏
页数:19
相关论文
共 112 条
[1]   Toughening of Bioceramic Composites for Bone Regeneration [J].
Abbas, Zahid ;
Dapporto, Massimiliano ;
Tampieri, Anna ;
Sprio, Simone .
JOURNAL OF COMPOSITES SCIENCE, 2021, 5 (10)
[2]   Synthesis of silicon-substituted hydroxyapatite by a hydrothermal method with two different phosphorous sources [J].
Aminian, Alieh ;
Solati-Hashjin, Mehran ;
Samadikuchaksaraei, Ali ;
Bakhshi, Farhad ;
Gorjipour, Fazel ;
Farzadi, Arghavan ;
Moztarzadeh, Fattolah ;
Schmuecker, Martin .
CERAMICS INTERNATIONAL, 2011, 37 (04) :1219-1229
[3]   Minimizing Risk of Failure from Ceramic-on-Ceramic Total Hip Prosthesis by Selecting Ceramic Materials Based on Tresca Stress [J].
Ammarullah, Muhammad Imam ;
Santoso, Gatot ;
Sugiharto, S. ;
Supriyono, Toto ;
Wibowo, Dwi Basuki ;
Kurdi, Ojo ;
Tauviqirrahman, Mohammad ;
Jamari, J. .
SUSTAINABILITY, 2022, 14 (20)
[4]  
[Anonymous], About us
[5]  
[Anonymous], 2007, POW DIFFR FIL
[6]   A review on hydroxyapatite coatings for the biomedical applications: experimental and theoretical perspectives [J].
Awasthi, Shikha ;
Pandey, Sarvesh Kumar ;
Arunan, E. ;
Srivastava, Chandan .
JOURNAL OF MATERIALS CHEMISTRY B, 2021, 9 (02) :228-249
[7]   Electro-crystallization of hydroxyapatite coatings on Nitinol rotating disk electrode [J].
Azar, Z. ;
Khalil-Allafi, J. ;
Etminanfar, M. R. .
MATERIALS RESEARCH EXPRESS, 2019, 6 (05)
[8]   Electrophoretic Deposition of 45S5 Bioglass(R) Coatings on the Ti6Al4V Prosthetic Alloy with Improved Mechanical Properties [J].
Azzouz, Imen ;
Faure, Joel ;
Khlifi, Kaouther ;
Larbi, Ahmed Cheikh ;
Benhayoune, Hicham .
COATINGS, 2020, 10 (12) :1-15
[9]   Bioactive Glasses: Where Are We and Where Are We Going? [J].
Baino, Francesco ;
Hamzehlou, Sepideh ;
Kargozar, Saeid .
JOURNAL OF FUNCTIONAL BIOMATERIALS, 2018, 9 (01)
[10]  
Baino F, 2017, Biomedical glasses, V3, P1, DOI [10.1515/bglass-2017-0001, 10.1515/bglass-2017-0001, DOI 10.1515/BGLASS-2017-0001]