Mid-term outcomes of a kinematically designed cruciate retaining total knee arthroplasty

被引:0
作者
Katzman, Jonathan L. [1 ]
Habibi, Akram A. [1 ]
Haider, Muhammad A. [1 ]
Cardillo, Casey [1 ]
Fernandez-Madrid, Ivan [1 ]
Meftah, Morteza [1 ]
Schwarzkopf, Ran [1 ]
机构
[1] NYU Langone Orthoped Hosp, Dept Orthoped Surg, Div Adult Reconstruct Surg, NYU Langone Hlth, 301 East 17th St,Suite 1402, New York, NY 10010 USA
来源
WORLD JOURNAL OF ORTHOPEDICS | 2024年 / 15卷 / 02期
关键词
Total knee arthroplasty; Cruciate retaining; Kinematic design; Survivorship; Bearing material; Prosthetic design; Clinical outcomes; Patient-reported outcome measures; PATIENT SATISFACTION; OXIDIZED ZIRCONIUM; POLYETHYLENE;
D O I
10.5312/wjo.v15.i2.118
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BACKGROUND Advances in implant material and design have allowed for improvements in total knee arthroplasty (TKA) outcomes. A cruciate retaining (CR) TKA provides the least constraint of TKA designs by preserving the native posterior cruciate ligament. Limited research exists that has examined clinical outcomes or patient reported outcome measures (PROMs) of a large cohort of patients undergoing a CR TKA utilizing a kinematically designed implant. It was hypothesized that the studied CR Knee System would demonstrate favorable outcomes and a clinically significant improvement in pain and functional scores. AIM To assess both short-term and mid-term clinical outcomes and PROMs of a novel CR TKA design. METHODS A retrospective, multi-surgeon study identified 255 knees undergoing a TKA utilizing a kinematically designed CR Knee System (JOURNEY (TM) II CR; Smith and Nephew, Inc., Memphis, TN) at an urban, academic medical institution between March 2015 and July 2021 with a minimum of two-years of clinical follow-up with an orthopedic surgeon. Patient demographics, surgical information, clinical outcomes, and PROMs data were collected via query of electronic medical records. The PROMs collected in the present study included the Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) and Patient-Reported Outcomes Measurement Information System (PROMIS (R)) scores. The significance of improvements in mean PROM scores from preoperative scores to scores collected at six months and two-years postoperatively was analyzed using Independent Samples t-tests. RESULTS Of the 255 patients, 65.5% were female, 43.8% were White, and patients had an average age of 60.6 years. Primary osteoarthritis (96.9%) was the most common primary diagnosis. The mean surgical time was 105.3 minutes and mean length of stay was 2.1 d with most patients discharged home (92.5%). There were 18 emergency department (ED) visits within 90 d of surgery resulting in a 90 d ED visit rate of 7.1%, including a 2.4% orthopedic-related ED visit rate and a 4.7% non-orthopedic-related ED visit rate. There were three (1.2%) hospital readmissions within 90 d postoperatively. With a mean time to latest follow-up of 3.3 years, four patients (1.6%) required revision, two for arthrofibrosis, one for aseptic femoral loosening, and one for peri-prosthetic joint infection. There were significant improvements in KOOS JR, PROMIS Pain Intensity, PROMIS Pain Interference, PROMIS Mobility, and PROMIS Physical Health from preoperative scores to six month and two-year postoperative scores. CONCLUSION The evaluated implant is an effective, novel design offering excellent outcomes and low complication rates. At a mean follow up of 3.3 years, four patients required revisions, three aseptic and one septic, resulting in an overall implant survival rate of 98.4% and an aseptic survival rate of 98.8%. The results of our study demonstrate the utility of this kinematically designed implant in the setting of primary TKA.
引用
收藏
页码:118 / 128
页数:12
相关论文
共 23 条
[1]  
Breugem SJM, 2017, WORLD J ORTHOP, V8, P922, DOI 10.5312/wjo.v8.i12.922
[2]   The use of highly crosslinked polyethylene in total knee arthroplasty [J].
Brown, T. S. ;
Van Citters, D. W. ;
Berry, D. J. ;
Abdel, M. P. .
BONE & JOINT JOURNAL, 2017, 99B (08) :996-1002
[3]   Factors Influencing Discharge Readiness After Total Knee Replacement [J].
Causey-Upton, Renee ;
Howell, Dana M. ;
Kitzman, Patrick H. ;
Custer, Melba G. ;
Dressler, Emily V. .
ORTHOPAEDIC NURSING, 2019, 38 (01) :6-14
[4]  
D'Apuzzo M, 2017, J BONE JOINT SURG AM, V99, P1093, DOI 10.2106/JBJS.16.00874
[5]   More Patients Are Being Discharged Home After Total Knee Arthroplasty, However Rates Vary Between Large Databases [J].
DeMik, David E. ;
Carender, Christopher N. ;
Glass, Natalie A. ;
Callaghan, John J. ;
Bedard, Nicholas A. .
JOURNAL OF ARTHROPLASTY, 2021, 36 (01) :173-179
[6]   A multicenter analysis of axial femorotibial rotation after total knee arthroplasty [J].
Dennis, DA ;
Komistek, RD ;
Mahfouz, MR ;
Walker, SA ;
Tucker, A .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :180-189
[7]  
Di Benedetto Paolo, 2019, Acta Biomed, V90, P91, DOI 10.23750/abm.v90i12-S.8984
[8]   New polys for old: Contribution or caveat? [J].
Greenwald, AS ;
Bauer, TW ;
Ries, MD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A :27-31
[9]   A comparative surface topographical analysis of explanted total knee replacement prostheses: Oxidised zirconium vs cobalt chromium femoral components [J].
Kennard, Emma ;
Scholes, Susan C. ;
Sidaginamale, Raghavendra ;
Gangadharan, Rajkumar ;
Weir, David J. ;
Holland, James ;
Deehan, David ;
Joyce, Thomas J. .
MEDICAL ENGINEERING & PHYSICS, 2017, 50 :59-64
[10]   Periprosthetic Joint Infection Is the Main Cause of Failure for Modern Knee Arthroplasty: An Analysis of 11,134 Knees [J].
Koh, Chuan Kong ;
Zeng, Irene ;
Ravi, Saiprassad ;
Zhu, Mark ;
Vince, Kelly G. ;
Young, Simon W. .