A prediction method and evaluation for assessing guide-supported dental implant accuracy

被引:0
作者
Xu, Liangwei [1 ]
Wu, Jianlei [1 ]
Chen, Lin [2 ]
Gao, Hongyan [2 ]
Zhu, Xian [2 ]
Yang, Guichan [2 ]
Chen, Yinghao [2 ]
机构
[1] Ningbo Polytech, Mech & Elect Engn Inst, Ningbo 315800, Peoples R China
[2] Zhejiang Jiamin Plast Co Ltd, Jiaxing 314027, Peoples R China
关键词
Surgical guide; implant accuracy; computer-assisted design and manufacture; implant deviation; SURGICAL TEMPLATE; IN-VITRO; SURGERY; IMPACT;
D O I
10.1080/10255842.2023.2284092
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Functional and esthetic results require accurate implant placement. We aimed to develop a predictive method for assessing dental implant accuracy, and to evaluate the cumulative system influence of surgical guides. A mathematical model was constructed to determine the influence of surface changes on a specific point, using Jacobian matrix expressions. Tooth-supported guides with a specific locating surface were designed for model verification. Ten guides were fabricated for in vitro virtual implant accuracy tests. Locating surface errors of 0.15 mm caused significant deviations in implant positioning. Tooth-supported guide location errors significantly impacted implant accuracy, emphasizing the need for precise guide placement.
引用
收藏
页码:188 / 196
页数:9
相关论文
共 36 条
[1]  
Bai S Z, 2021, Zhonghua Kou Qiang Yi Xue Za Zhi, V56, P170, DOI 10.3760/cma.j.cn112144-20210107-00008
[2]   Factors influencing transfer accuracy of cone beam CT-derived template-based implant placement [J].
Behneke, Alexandra ;
Burwinkel, Matthias ;
Behneke, Nikolaus .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (04) :416-423
[3]   Influence of surgical guide support and implant site location on accuracy of static Computer-Assisted Implant Surgery [J].
El Kholy, Karim ;
Lazarin, Rafael ;
Janner, Simone F. M. ;
Faerber, Karin ;
Buser, Ramona ;
Buser, Daniel .
CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (11) :1067-1075
[4]   The influence of guided sleeve height, drilling distance, and drilling key length on the accuracy of static Computer-Assisted Implant Surgery [J].
El Kholy, Karim ;
Janner, Simone F. M. ;
Schimmel, Martin ;
Buser, Daniel .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (01) :101-107
[5]   Precision of Dental Implant Digitization Using Intraoral Scanners [J].
Fluegge, Tabea V. ;
Att, Wael ;
Metzger, Marc C. ;
Nelson, Katja .
INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2016, 29 (03) :277-283
[6]  
Geng W, 2015, INT J CLIN EXP MED, V8, P8442
[7]   The accuracy of static vs. dynamic computer-assisted implant surgery in single tooth space: A randomized controlled trial [J].
Kaewsiri, Dechawat ;
Panmekiate, Soontra ;
Subbalekha, Keskanya ;
Mattheos, Nikos ;
Pimkhaokham, Atiphan .
CLINICAL ORAL IMPLANTS RESEARCH, 2019, 30 (06) :505-514
[8]   Influence of 3D-printing method, resin material, and sterilization on the accuracy of virtually designed surgical implant guides [J].
Kessler, Andreas ;
Dosch, Maximilian ;
Reymus, Marcel ;
Folwaczny, Matthias .
JOURNAL OF PROSTHETIC DENTISTRY, 2022, 128 (02) :196-204
[9]   Accuracy and dimensional reproducibility by model scanning, intraoral scanning, and CBCT imaging for digital implant dentistry [J].
Komuro, Akira ;
Yamada, Yoichi ;
Uesugi, Satoshi ;
Terashima, Hiroaki ;
Kimura, Masashi ;
Kishimoto, Hiroto ;
Iida, Tsutomu ;
Sakamoto, Katsuya ;
Okuda, Kenichi ;
Kusano, Kaoru ;
Baba, Shunsuke ;
Sakamoto, Takashi .
INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2021, 7 (01)
[10]  
Lee DH, 2016, J ADV PROSTHODONT, V8, P207