Evaluation of Image Quality and Detectability of Deep Learning Image Reconstruction (DLIR) Algorithm in Single- and Dual-energy CT

被引:12
作者
Zhong, Jingyu [1 ]
Shen, Hailin [2 ]
Chen, Yong [3 ]
Xia, Yihan [3 ]
Shi, Xiaomeng [4 ]
Lu, Wei [5 ]
Li, Jianying [6 ]
Xing, Yue [1 ]
Hu, Yangfan [1 ]
Ge, Xiang [1 ]
Ding, Defang [1 ]
Jiang, Zhenming [1 ]
Yao, Weiwu [1 ]
机构
[1] Shanghai Jiao Tong Univ, Tongren Hosp, Dept Imaging, Sch Med, 1111 Xianxia Rd, Shanghai 200336, Peoples R China
[2] Shanghai Jiao Tong Univ, Suzhou Kowloon Hosp, Dept Radiol, Sch Med, Suzhou 215028, Peoples R China
[3] Shanghai Jiao Tong Univ, Ruijin Hosp, Sch Med, Dept Radiol, Shanghai 200025, Peoples R China
[4] Imperial Coll London, Dept Mat, South Kensington Campus, London SW7 2AZ, England
[5] GE Healthcare, Computed Tomog Res Ctr, Shanghai 201203, Peoples R China
[6] GE Healthcare, Computed Tomog Res Ctr, Beijing 100176, Peoples R China
基金
中国国家自然科学基金;
关键词
Multidetector computed tomography; Deep learning; Image reconstruction; Image enhancement; Radiation dosage; TASK-BASED PERFORMANCE; ITERATIVE RECONSTRUCTION; COMPUTED-TOMOGRAPHY; DOSE REDUCTION; ABDOMINAL CT; NOISE; CHEST; RESOLUTION; SYSTEM;
D O I
10.1007/s10278-023-00806-z
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study is aimed to evaluate effects of deep learning image reconstruction (DLIR) on image quality in single-energy CT (SECT) and dual-energy CT (DECT), in reference to adaptive statistical iterative reconstruction-V (ASIR-V). The Gammex 464 phantom was scanned in SECT and DECT modes at three dose levels (5, 10, and 20 mGy). Raw data were reconstructed using six algorithms: filtered back-projection (FBP), ASIR-V at 40% (AV-40) and 100% (AV-100) strength, and DLIR at low (DLIR-L), medium (DLIR-M), and high strength (DLIR-H), to generate SECT 120kVp images and DECT 120kVp-like images. Objective image quality metrics were computed, including noise power spectrum (NPS), task transfer function (TTF), and detectability index (d '). Subjective image quality evaluation, including image noise, texture, sharpness, overall quality, and low- and high-contrast detectability, was performed by six readers. DLIR-H reduced overall noise magnitudes from FBP by 55.2% in a more balanced way of low and high frequency ranges comparing to AV-40, and improved the TTF values at 50% for acrylic inserts by average percentages of 18.32%. Comparing to SECT 20 mGy AV-40 images, the DECT 10 mGy DLIR-H images showed 20.90% and 7.75% improvement in d ' for the small-object high-contrast and large-object low-contrast tasks, respectively. Subjective evaluation showed higher image quality and better detectability. At 50% of the radiation dose level, DECT with DLIR-H yields a gain in objective detectability index compared to full-dose AV-40 SECT images used in daily practice.
引用
收藏
页码:1390 / 1407
页数:18
相关论文
共 45 条
  • [1] Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT
    Akagi, Motonori
    Nakamura, Yuko
    Higaki, Toru
    Narita, Keigo
    Honda, Yukiko
    Zhou, Jian
    Yu, Zhou
    Akino, Naruomi
    Awai, Kazuo
    [J]. EUROPEAN RADIOLOGY, 2019, 29 (11) : 6163 - 6171
  • [2] Bornet PA, 2022, EUR RADIOL, V32, P3161, DOI 10.1007/s00330-021-08410-x
  • [3] A study of using a deep learning image reconstruction to improve the image quality of extremely low-dose contrast-enhanced abdominal CT for patients with hepatic lesions
    Cao, Le
    Liu, Xiang
    Li, Jianying
    Qu, Tingting
    Chen, Lihong
    Cheng, Yannan
    Hu, Jieliang
    Sun, Jingtao
    Guo, Jianxin
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1118)
  • [4] Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods
    Chen, Baiyu
    Christianson, Olav
    Wilson, Joshua M.
    Samei, Ehsan
    [J]. MEDICAL PHYSICS, 2014, 41 (07)
  • [5] Multivendor Comparison of Quantification Accuracy of Iodine Concentration and Attenuation Measurements by Dual-Energy CT: A Phantom Study
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Chang, Rui
    Fan, Jing
    Jiang, Jiang
    Xia, Yihan
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 827 - 839
  • [6] Validation of algorithmic CT image quality metrics with preferences of radiologists
    Cheng, Yuan
    Abadi, Ehsan
    Smith, Taylor Brunton
    Ria, Francesco
    Meyer, Mathias
    Marin, Daniele
    Samei, Ehsan
    [J]. MEDICAL PHYSICS, 2019, 46 (11) : 4837 - 4846
  • [7] An Improved Index of Image Quality for Task-based Performance of CT Iterative Reconstruction across Three Commercial Implementations
    Christianson, Olav
    Chen, Joseph J. S.
    Yang, Zhitong
    Saiprasad, Ganesh
    Dima, Alden
    Filliben, James J.
    Peskin, Adele
    Trimble, Christopher
    Siegel, Eliot L.
    Samei, Ehsan
    [J]. RADIOLOGY, 2015, 275 (03) : 725 - 734
  • [8] A Third-Generation Adaptive Statistical Iterative Reconstruction Technique: Phantom Study of Image Noise, Spatial Resolution, Lesion Detectability, and Dose Reduction Potential
    Euler, Andre
    Solomon, Justin
    Marin, Daniele
    Nelson, Rendon C.
    Samei, Ehsan
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 210 (06) : 1301 - 1308
  • [9] Image Quality Evaluation in Dual-Energy CT of the Chest, Abdomen, and Pelvis in Obese Patients With Deep Learning Image Reconstruction
    Fair, Eric
    Profio, Mark
    Kulkarni, Naveen
    Laviolette, Peter S.
    Barnes, Bret
    Bobholz, Samuel
    Levenhagen, Maureen
    Ausman, Robin
    Griffin, Michael O.
    Duvnjak, Petar
    Zorn, Adam P.
    Foley, W. Dennis
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2022, 46 (04) : 604 - 611
  • [10] Simulated Dose Reduction for Abdominal CT With Filtered Back Projection Technique: Effect on Liver Lesion Detection and Characterization
    Falt, Tobias
    Soderberg, Marcus
    Horberg, Lisa
    Christoffersen, Christina
    Lang, Kristina
    Abul-Kasim, Kasim
    Leander, Peter
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2019, 212 (01) : 84 - 93