Comparison of Short and Long-Tunnel Needle Track for Ahmed Glaucoma Valve Implantation in a Private Eye Center in the Philippines: A Retrospective Study

被引:0
作者
San Pedro, Marie Jeremy B. [1 ]
Gomez, James Paul S. [1 ]
Leuenberger, Edgar U. [1 ,2 ,3 ]
机构
[1] Asian Eye Inst, Makati, Philippines
[2] Univ East Ramon Magsaysay Mem Med Ctr Inc, Dept Ophthalmol, Quezon City, Philippines
[3] Ospital Ng Makati, Dept Ophthalmol, Makati, Philippines
来源
CLINICAL OPHTHALMOLOGY | 2023年 / 17卷
关键词
glaucoma drainage device; Ahmed glaucoma valve; long needle track technique; short needle track technique; POSTOPERATIVE HYPOTONY; SURGICAL TECHNIQUE; RISK-FACTORS; GRAFT;
D O I
10.2147/OPTH.S405924
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: This study aimed to compare the success and complication rates among patients implanted with Ahmed Glaucoma Valve (AGV) using the short and long tunnel technique through retrospective chart review.Patients and Methods: We reviewed 54 charts of adult patients who underwent AGV implantation using a Short-Needle Track (SNT) or Long-Needle Track (LNT) technique. Intraocular pressures (IOP), Best Corrected Visual Acuity (BCVA) and number of medications were recorded pre-operatively, and at Day 1, 3, 7, Month 1, 3, 6 post-operatively. Treatment success, occurrence of Hypertensive Phase (HP), complication and procedures done after AGV implantation were compared between the two groups usingResults: A total of 20 (LNT) and 21 (SNT) charts were included in the study. There was no significant difference between the median postoperative IOP, BCVA, and number of anti-glaucoma medications between the two groups at each time interval. The comparison between the occurrence of HP (P = 0.435) and success rates (P = 0.476) between the two groups yielded no significant difference. Flat/ shallow anterior chamber (AC) was seen exclusively in three eyes (14%) in the SNT group (P = 0.039). There was one occurrence of plate exposure in the LNT group (P = 0.149).Conclusion: The LNT technique of AGV Implantation may be used as an alternative to the traditional SNT (with autologous graft). The long needle track offers the advantage reducing the risk of complications arising from shallow anterior chamber post-operatively.
引用
收藏
页码:1691 / 1699
页数:9
相关论文
共 29 条
[1]   Evaluation of Ahmed glaucoma valve implantation through a needle-generated scleral tunnel in Mexican children with glaucoma [J].
Albis-Donado, Oscar ;
Gil-Carrasco, Felix ;
Romero-Quijada, Rafael ;
Thomas, Ravi .
INDIAN JOURNAL OF OPHTHALMOLOGY, 2010, 58 (05) :365-373
[2]   Ahmed glaucoma valve in vitrectomized eyes: An implantation technique to minimize early postoperative fluid-loss hypotony [J].
Arpa, Paolo ;
Arpa, Cristina .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (05) :2759-2764
[3]   A clinical study of the Ahmed glaucoma valve implant in advanced glaucoma [J].
Ayyala, RS ;
Zurakowski, D ;
Smith, JA ;
Monshizadeh, R ;
Netland, PA ;
Richards, DW ;
Layden, WE .
OPHTHALMOLOGY, 1998, 105 (10) :1968-1976
[4]   Risk factors of implant exposure outside the conjunctiva after Ahmed glaucoma valve implantation [J].
Byun, Yong Soo ;
Lee, Na Young ;
Park, Chan Kee .
JAPANESE JOURNAL OF OPHTHALMOLOGY, 2009, 53 (02) :114-119
[5]   Outcome of fixed volume intracameral sodium hyaluronate 1.4% injection for early postoperative hypotony after Baerveldt glaucoma implant [J].
Chiam, Patrick J. ;
Chen, Xiaoli ;
Haque, M. Sayeed ;
Sung, Velota C. T. .
CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2018, 46 (09) :1035-1040
[6]   Outcomes and Complications of Ahmed Tube Implantation in Asian Eyes [J].
Choo, Jessica Q. H. ;
Chen, Z. D. ;
Koh, Victor ;
Liang, Shen ;
Aquino, Cecilia M. ;
Sng, Chelvin ;
Chew, Paul .
JOURNAL OF GLAUCOMA, 2018, 27 (08) :733-738
[7]   Comparison of surgical outcomes of trabeculectomy, Ahmed shunt, and Baerveldt shunt in uveitic glaucoma [J].
Chow A. ;
Burkemper B. ;
Varma R. ;
Rodger D.C. ;
Rao N. ;
Richter G.M. .
Journal of Ophthalmic Inflammation and Infection, 8 (1)
[8]   Single long scleral tunnel technique for prevention of Ahmed valve tube exposure [J].
Eslami, Yadollah ;
Azaripour, Ebrahim ;
Mohammadi, Masoud ;
Kiarudi, Mohammad Yaser ;
Fakhraie, Ghasem ;
Zarei, Reza ;
Alizadeh, Yousef ;
Moghimi, Sasan .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2019, 29 (01) :52-56
[9]   Ab Interno Tube Occlusion for Postoperative Hypotony in a Patient With an Ahmed Glaucoma Drainage Device [J].
Feinstein, Max ;
Moussa, Kareem ;
Han, Ying .
JOURNAL OF GLAUCOMA, 2018, 27 (03) :E61-E63
[10]   Peritubular filtration as cause of severe hypotony after ahmed valve implantation for glaucoma [J].
García-Feijoó, J ;
Cuiña-Sardiña, R ;
Méndez-Fernández, C ;
Castillo-Gómez, A ;
García-Sánchez, J .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2001, 132 (04) :571-572