Validation of quantitative loop-mediated isothermal amplification assay using a fluorescent distance-based paper device for detection of Escherichia coli in urine

被引:4
作者
Saengsawang, Natkrittaya [1 ]
Ruang-areerate, Panthita [2 ]
Kaeothaisong, Nuanlaong [3 ]
Leelayoova, Saovanee [3 ]
Mungthin, Mathirut [3 ]
Juntanawiwat, Piraporn [4 ]
Hanyanunt, Patomroek [4 ]
Potisuwan, Patsanun [4 ]
Kesakomol, Piyanate [5 ]
Butsararattanagomen, Pornphan [1 ]
Wichaiwong, Pattarawadee [1 ]
Dungchai, Wijitar [1 ]
Ruang-areerate, Toon [3 ]
机构
[1] King Mongkuts Univ Technol Thonburi, Dept Chem, Analyt Chem, Fac Sci, Bangkok 10140, Thailand
[2] Natl Sci & Technol Dev Agcy NSTDA, BIOTEC, Pathum Thani 12120, Thailand
[3] Phramongkutklao Coll Med, Dept Parasitol, Bangkok 10400, Thailand
[4] Phramongkutklao Hosp, Div Microbiol, Dept Clin Pathol, Bangkok 10400, Thailand
[5] Phramongkutklao Coll Med, Dept Microbiol, Bangkok 10400, Thailand
关键词
TRACT-INFECTION; RAPID DETECTION; DIAGNOSIS; LAMP; EPIDEMIOLOGY; CHILDREN; WOMEN; DNA;
D O I
10.1038/s41598-023-46001-6
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Uropathogenic Escherichia coli (UPEC) causes up to 90% of urinary tract infections (UTI) which is more prevalent among females than males. In urine, patients with symptomatic UTI usually have a high concentration of bacterial infection, >= 10(5) colony-forming units (CFU) per mL, in which the culture method is regularly the gold standard diagnosis. In this study, a simple and inexpensive distance-based paper device (dPAD) combined with the fluorescent closed tube LAMP assay was validated for simultaneously screening and semi-quantifying the infection level of E. coli in 440 urine samples of patients with UTI. The dPAD could measure the LAMP amplicons and semi-quantify the levels of E. coli infection in heavy (>= 10(4) CFU/mL), light (<= 10(3) CFU/mL) and no infection. The sensitivity and specificity had reliable performances, achieving as high as 100 and 92.7%, respectively. The one step LAMP assay could be performed within 3 h, which was 7.5 times faster than the culture method. To empower early UTI diagnosis and fast treatment, this inexpensive dPAD tool combined with the fluorescent closed tube LAMP assay is simple, reliably fast and practically portable for point-of-care settings, particularly in resource-limited areas, which can be set up in all levels of healthcare facilities.
引用
收藏
页数:10
相关论文
共 53 条
  • [51] Laboratory diagnosis of urinary tract infections in adult patients
    Wilson, ML
    Gaido, L
    [J]. CLINICAL INFECTIOUS DISEASES, 2004, 38 (08) : 1150 - 1158
  • [52] Genetic evidence supporting the fecal-perineal-urethral hypothesis in cystitis caused by Escherichia coli
    Yamamoto, S
    Tsukamoto, T
    Terai, A
    Kurazono, H
    Takeda, Y
    Yoshida, O
    [J]. JOURNAL OF UROLOGY, 1997, 157 (03) : 1127 - 1129
  • [53] A newly developed paper embedded microchip based on LAMP for rapid multiple detections of foodborne pathogens
    Zhang, Mimi
    Liu, Jinfeng
    Shen, Zhiqiang
    Liu, Yongxin
    Song, Yang
    Liang, Yu
    Li, Zhende
    Nie, Lingmei
    Fang, Yanjun
    Zhao, Youquan
    [J]. BMC MICROBIOLOGY, 2021, 21 (01)