Impact of photon counting detector CT derived virtual monoenergetic images and iodine maps on the diagnosis of pleural empyema

被引:19
作者
Jungblut, Lisa [1 ]
Abel, Frederik [1 ]
Nakhostin, Dominik [1 ]
Mergen, Viktor [1 ]
Sartoretti, Thomas [1 ]
Euler, Andre [1 ]
Frauenfelder, Thomas [1 ]
Martini, Katharina [1 ,2 ]
机构
[1] Univ Zurich, Univ Hosp Zurich, Inst Diagnost & Intervent Radiol, Raemistr 100, CH-8091 Zurich, Switzerland
[2] Univ Hosp Zurich, Inst Diagnost & Intervent Radiol, Raemistr 100, F-8091 Zurich, France
关键词
Computed tomography; Empyema; Pleura; Image processing; Computer -assisted image interpretation; Chest; DUAL-ENERGY CT; PARAPNEUMONIC EFFUSIONS; FEASIBILITY; CHILDREN; SYSTEM; FAT;
D O I
10.1016/j.diii.2022.09.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate the impact of virtual monoenergetic image (VMI) energies and iodine maps on the diagnosis of pleural empyema with photon counting detector computed tomography (PCD-CT). Materials and methods: In this IRB-approved retrospective study, consecutive patients with non-infectious pleural effusion or histopathology-proven empyema were included. PCD-CT examinations were performed on a dual-source PCD-CT in the multi-energy (QuantumPlus) mode at 120 kV with weight-adjusted intravenous contrast-agent. VMIs from 40-70 keV obtained in 10 keV intervals and an iodine map was reconstructed for each scan. CT attenuation was measured in the aorta, the pleura and the peripleural fat (between autochthonous dorsal muscles and dorsal ribs). Contrast-to-noise (CNR) and signal-to-noise (SNR) ratios were calculated. Two blinded radiologists evaluated if empyema was present (yes/no), and rated diagnostic confidence (1 to 4; not confident to fully confident, respectively) with and without using the iodine map. Sensitivity, specificity and diagnostic confidence were estimated. Interobserver agreement was estimated using an unweighted Cohen kappa test. A one-way ANOVA was used to compare variables. Differences in sensitivity and specificity between the different levels of energy were searched using McNemar test. Results: Sixty patients (median age, 60 years; 26 women) were included. A strong negative correlation was found between image noise and VMI energies (r = -0.98; P = 0.001) and CNR increased with lower VMI energies (r = -0.98; P = 0.002). Diagnostic accuracy (96%; 95% CI: 82-100) as well as diagnostic confidence (3.4 + 0.75 [SD]) were highest at 40 keV. Diagnostic accuracy and confidence at higher VMI energies improved with the addition of iodine maps (P <= 0.001). Overall, no difference in CT attenuation of peripleural fat between patients with empyema and those with pleural effusion was found (P = 0.07). Conclusion: Low VMI energies lead to a higher diagnostic accuracy and diagnostic confidence in the diagnosis of pleural empyema. Iodine maps help in diagnosing empyema only at high VMI energies. (c) 2022 The Authors. Published by Elsevier Masson SAS on behalf of Societe francaise de radiologie. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/)
引用
收藏
页码:84 / 90
页数:7
相关论文
共 50 条
  • [31] Dual-energy CT with virtual monoenergetic images and iodine maps improves tumor conspicuity in patients with pancreatic ductal adenocarcinoma
    Liang, Hongwei
    Zhou, Yang
    Zheng, Qiao
    Yan, Gaowu
    Liao, Hongfan
    Du, Silin
    Zhang, Xiaohui
    Lv, Fajin
    Zhang, Zhiwei
    Li, Yong-mei
    INSIGHTS INTO IMAGING, 2022, 13 (01)
  • [32] Noise characteristics of virtual monoenergetic images from a novel detector-based spectral CT scanner
    Kalisz, Kevin
    Rassouli, Negin
    Dhanantwari, Amar
    Jordan, David
    Rajiah, Prabhakar
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 98 : 118 - 125
  • [33] Optimizing Arterial Vessel Contrast in Portal Venous Phase with Virtual Monoenergetic Images from Photon-Counting Detector CT Scans of the Abdomen-First Clinical Experiences
    Dillinger, Daniel
    Overhoff, Daniel
    Ayx, Isabelle
    Kaatsch, Hanns L.
    Hagen, Achim
    Schoenberg, Stefan O.
    Waldeck, Stephan
    DIAGNOSTICS, 2024, 14 (06)
  • [34] Photon-counting detector CT-based virtual monoenergetic reconstructions: repeatability and reproducibility of radiomics features of an organic phantom and human myocardium
    Wolf, Elias V.
    Mueller, Lukas
    Schoepf, U. Joseph
    Fink, Nicola
    Griffith, Joseph P.
    Zsarnoczay, Emese
    Baruah, Dhiraj
    Suranyi, Pal
    Kabakus, Ismael M.
    Halfmann, Moritz C.
    Emrich, Tilman
    Varga-Szemes, Akos
    O'Doherty, Jim
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2023, 7 (01)
  • [35] Quantitative performance of photon-counting CT at low dose: Virtual monochromatic imaging and iodine quantification
    Vrbaski, Stevan
    Bache, Steve
    Rajagopal, Jayasai
    Samei, Ehsan
    MEDICAL PHYSICS, 2023, 50 (09) : 5421 - 5433
  • [36] Photon Counting CT Angiography of the Head and Neck: Image Quality Assessment of Polyenergetic and Virtual Monoenergetic Reconstructions
    Michael, Arwed Elias
    Boriesosdick, Jan
    Schoenbeck, Denise
    Lopez-Schmidt, Ingo
    Kroeger, Jan Robert
    Moenninghoff, Christoph
    Horstmeier, Sebastian
    Pennig, Lenhard
    Borggrefe, Jan
    Niehoff, Julius Henning
    DIAGNOSTICS, 2022, 12 (06)
  • [37] Photon-Counting Detector CT Virtual Monoengergetic Images for Cochlear Implant Visualization-A Head to Head Comparison to Energy-Integrating Detector CT
    Waldeck, Stephan
    Overhoff, Daniel
    Alizadeh, Leona
    Becker, Benjamin, V
    Port, Matthias
    Froelich, Matthias F.
    Brockmann, Marc A.
    Schumann, Sven
    Vogl, Thomas J.
    Schoenberg, Stefan O.
    Schmidt, Sandra
    TOMOGRAPHY, 2022, 8 (04) : 1642 - 1648
  • [38] Spectral detector CT-derived virtual non-contrast images: comparison of attenuation values with unenhanced CT
    Ananthakrishnan, Lakshmi
    Rajiah, Prabhakar
    Ahn, Richard
    Rassouli, Negin
    Xi, Yin
    Soesbe, Todd C.
    Lewis, Matthew A.
    Lenkinski, Robert E.
    Leyendecker, John R.
    Abbara, Suhny
    ABDOMINAL RADIOLOGY, 2017, 42 (03) : 702 - 709
  • [39] Virtual Monoenergetic Images from Spectral Detector CT Enable Radiation Dose Reduction in Unenhanced Cranial CT
    Reimer, R. P.
    Flatten, D.
    Lichtenstein, T.
    Zopfs, D.
    Neuhaus, V.
    Kabbasch, C.
    Maintz, D.
    Borggrefe, J.
    Hokamp, N. Grosse
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (10) : 1617 - 1623
  • [40] Photon-Counting Detector CT for Liver Lesion Detection-Optimal Virtual Monoenergetic Energy for Different Simulated Patient Sizes and Radiation Doses
    Racine, Damien
    Mergen, Victor
    Viry, Anais
    Frauenfelder, Thomas
    Alkadhi, Hatem
    Vitzthum, Veronika
    Euler, Andre
    INVESTIGATIVE RADIOLOGY, 2024, 59 (08) : 554 - 560