A Comparison of Percutaneous Pedicle Screw Accuracy Between Robotic Navigation and Novel Fluoroscopy-Based Instrument Tracking for Patients Undergoing Instrumented Thoracolumbar Surgery
被引:3
|作者:
Wang, Timothy Y.
[1
]
Tabarestani, Troy Q.
[2
]
Mehta, Vikram A.
[1
]
Sankey, Eric W.
[1
]
Karikari, Isaac O.
[1
]
Goodwin, C. Rory
[1
]
Than, Khoi D.
[1
]
Abd-El-Barr, Muhammad M.
[1
]
机构:
[1] Duke Univ, Dept Neurol Surg, Durham, NC USA
[2] Duke Univ, Sch Med, Durham, NC 27708 USA
基金:
美国国家卫生研究院;
关键词:
Fluoroscopy;
Instrument navigation;
Percutaneous;
-Accuracy;
-Robot;
PLACEMENT;
FIXATION;
GUIDANCE;
D O I:
10.1016/j.wneu.2023.01.037
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
BACKGROUND: The accuracy of pedicle screws placed with instrument tracking and robotic navigation are indi-vidually comparable or superior to placement using stan-dard fluoroscopy, however head-to-head comparisons between these adjuncts in a similar surgical population have yet to be performed.METHODS: Consecutive patients undergoing percuta-neous thoracic and lumbosacral spinal instrumentation were retrospectively enrolled. Instrumentation was per-formed using either fluoroscopy-based instrument tracking system (TrackX, TrackX Technologies) or robotic-navigation (ExcelsiusGPS, Globus Medical). Postinstrumentation computed tomography scans were graded for breach ac-cording to the Gertzbein-Robbins scale, with "acceptable" screws deemed as Grade A or B and "unacceptable" screws deemed as Grades C through E. Accuracy data was compared between both instrumentation modalities.RESULTS: Fifty-three patients, comprising a total of 250 screws (167 robot, 83 instrument tracking) were included. The overall accuracy between both modalities was similar, with 96.4% and 97.6% of screws with acceptable accuracy between instrument tracking and robotic navigation, respectively (I-squared 0.30, df = 1, P = 0.58). Between instrument tracking and robotic navigation, 92.8% and 95.8% of screws received Grade A, 3.6% and 1.8% a Grade B, 1.2% and 1.2% a Grade C, 1.2% and 0.6% a Grade D, and 1.2% and 0.6% a Grade E, respectively. The robot was abandoned intraoperatively in 2 cases due to unrecoverable registration inaccuracy or software failure, leading to abandonment of 8 potential screws (4.8%). -CONCLUSIONS: In a similar patient population, there is a similarly high degree of instrumentation accuracy be-tween fluoroscopy-based instrument tracking and robotic navigation. There is a rare chance for screw breach with either surgical adjunct.
引用
收藏
页码:E389 / E395
页数:7
相关论文