Clinical image quality assessment and mean glandular dose for full field digital mammography

被引:0
作者
Nassar, Joyce [1 ]
Rizk, Chadia [1 ,2 ]
Fares, Georges [1 ]
Tohme, Carla [3 ]
Braidy, Chady [3 ]
Farah, Jad [4 ]
机构
[1] St Joseph Univ, Fac Sci, POB 11-514, Beirut 11072050, Lebanon
[2] Natl Council Sci Res, Lebanese Atom Energy Commiss, Beirut 118281, Lebanon
[3] Hotel Dieu France Hosp, Radiol Dept, POB 166830, Beirut, Lebanon
[4] Vis RT Ltd, Dove House,Arcadia Ave, London N3 2JU, England
关键词
mammography; image quality; system performance; positioning performance; average glandular dose; DIAGNOSTIC REFERENCE LEVELS; EQUIPMENT PERFORMANCE; CRITERIA; RELIABILITY; VALIDITY; FAILURE;
D O I
10.1088/1361-6498/ad1cd4
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study aims to assess the image quality (IQ) of 12 mammographic units and to identify units with potential optimisation needs. Data for 350 mammography examinations meeting inclusion criteria were collected retrospectively from April 2021 to April 2022. They were categorised based on the medical reports into 10 normal cases, 10 cases displaying calcifications and 10 cases presenting lesions. Two radiologists assessed the IQ of 1400 mammograms, evaluating system performance per Boita et al's study and positioning performance following European guidelines. To measure agreement between the two radiologists, the Cohen's Kappa coefficient (kappa) was computed, quantifying the excess of agreement beyond chance. The visual grading analysis score (VGAS) was computed to compare system and positioning performance assessments across different categories and facilities. Median average glandular dose (AGD) values for cranio caudal and medio lateral oblique views were calculated for each category and facility and compared to the national diagnostic reference levels. The health facilities were categorised by considering both IQ VGAS and AGD levels. Inter-rater agreement between radiologists ranged from poor (kappa < 0.20) to moderate (0.41 < kappa < 0.60), likely influenced by inherent biases and distinct IQ expectations. 50% of the facilities were classified as needing corrective actions for their system performance as they had IQ or high AGD that could increase recall rate and radiation risk and 50% of the health facilities exhibited insufficient positioning performance that could mask tumour masses and microcalcifications. The study's findings emphasise the importance of implementing quality assurance programs to ensure optimal IQ for accurate diagnoses while adhering to radiation exposure guidelines. Additionally, comprehensive training for technologists is essential to address positioning challenges. These initiatives collectively aim to enhance the overall quality of breast imaging services, contributing to improved patient care.
引用
收藏
页数:15
相关论文
共 50 条
  • [41] Digital mammography screening: Average glandular dose and first performance parameters
    Weigel, S.
    Girnus, R.
    Czwoydzinski, J.
    Decker, T.
    Spital, S.
    Heindel, W.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2007, 179 (09): : 892 - 895
  • [42] Dosimetric evaluation of the mean glandular dose for mammography in Korean women: A preliminary report
    Oh, KK
    Hur, J
    Kim, EK
    Choo, SS
    YONSEI MEDICAL JOURNAL, 2003, 44 (05) : 863 - 868
  • [43] Comprehensive assessment of image quality in synthetic and digital mammography: a quantitative comparison
    Barca, Patrizio
    Lamastra, Rocco
    Aringhieri, Giacomo
    Tucciariello, Raffaele Maria
    Traino, Antonio
    Fantacci, Maria Evelina
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2019, 42 (04) : 1141 - 1152
  • [44] Assessment of Mean Glandular Dose in Mammography System with Different Anode-Filter Combinations Using MCNP Code
    Gholamkar, Lida
    Mowlavi, Ali Asghar
    Sadeghi, Mahdi
    Athari, Mitra
    IRANIAN JOURNAL OF RADIOLOGY, 2016, 13 (04)
  • [45] Picture Quality According to the Type of Detector in Full-field Digital Mammography
    Kim, Hyun-Soo
    Jeong, Jae-Ho
    Lee, Jong-Woong
    Kang, Hee-Doo
    Dong, Kyung-Rae
    Chung, Woon-Kwan
    Kim, Eun-Soo
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 58 (02) : 364 - 371
  • [46] Comprehensive assessment of image quality in synthetic and digital mammography: a quantitative comparison
    Patrizio Barca
    Rocco Lamastra
    Giacomo Aringhieri
    Raffaele Maria Tucciariello
    Antonio Traino
    Maria Evelina Fantacci
    Australasian Physical & Engineering Sciences in Medicine, 2019, 42 : 1141 - 1152
  • [47] A Monte Carlo model for mean glandular dose evaluation in spot compression mammography
    Sarno, Antonio
    Dance, David R.
    van Engen, Ruben E.
    Young, Kenneth C.
    Russo, Paolo
    Di Lillo, Francesca
    Mettivier, Giovanni
    Bliznakova, Kristina
    Fei, Baowei
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2017, 44 (07) : 3848 - 3860
  • [48] A Study of Mean Glandular Dose during Diagnostic Mammography in Hospitals in Hanoi, Vietnam
    Nguyen Thai Ha
    Nguyen Duc Thuan
    Nguyen Thu Van
    THIRD INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM, 2010, 27 : 263 - 265
  • [49] Investigation into the computer assessment of image quality in mammography
    Kaplish, A
    Pretlove, JRG
    Young, KC
    Horton, PW
    MEDICAL IMAGING 1996: IMAGE PROCESSING, 1996, 2710 : 979 - 987
  • [50] Study on patient dosimetry and image quality in digital mammography
    Xavier A.C.S.
    Andrade M.E.A.
    Pinto B.V.-C.C.
    de Barros V.S.M.
    Kramer R.
    Khoury H.J.
    Andrade, Marcos Ely Almeida (marcos.ely@gmail.com), 1600, Brazilian Society of Biomedical Engineering (33): : 138 - 143