Vanguard Study: Initial experience with the new fourth generation Pipeline Vantage Flow Diverter (PVFD): 6-month results, technical and clinical considerations

被引:17
作者
de Villiers, Laetitia [1 ]
Carraro do Nascimento, Vinicius [1 ]
Domitrovic, Luis [1 ]
Dhillon, Permesh Singh [2 ]
Rice, Hal [3 ]
机构
[1] Gold Coast Univ Hosp, Intervent Neuroradiol, Gold Coast, Qld, Australia
[2] Univ Nottingham, Radiol Sci, Mental Hlth & Clin Neurosci, Nottingham, England
[3] Gold Coast Univ Hosp, Dept Intervent Neuroradiol, Southport, Qld, Australia
关键词
Device; Aneurysm; Flow Diverter; EMBOLIZATION DEVICE; ANEURYSMS; TECHNOLOGY;
D O I
10.1136/jnis-2023-021182
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Background The Pipeline Embolization Device has proven to be a safe and effective device for the treatment of intracranial aneurysms. The Pipeline Vantage Flow Diverter (PVFD) with Shield Technology is the new fourth generation of this implant, with modifications made compared to previous iterations. We aimed to evaluate the mechanical properties and clinical safety and efficacy of this device.Methods Vanguard is a single arm, single center, prospective study. Between April 2021 and April 2023, all consecutive patients with an unruptured aneurysm treated with Pipeline Vantage flow-diverting stents were included. There were no aneurysm size or location exclusion criteria. Safety (neurological serious adverse events) and efficacy (device deployment and aneurysm occlusion) were independently reviewed. Imaging follow-up data, and immediate, early (<30 days), and delayed (>30 days) neurological serious adverse events were independently assessed.Results 101 consecutive patients with a total of 115 aneurysms were included. The aneurysms were situated in the anterior (90.4%) or posterior (9.6%) circulations. A total of 124 devices were implanted. The deployment success rate was 100%. In four (4.0%) cases post-deployment angioplasty was required to optimize device wall apposition. Occlusion rates at 1 month were 54.7%, at 3 months 72.1%, and at 6 months 81.7%. Morbidity and mortality were 4.9% and 0%, respectively, at 6 months. Eight cases (6.9%) demonstrated in-stent stenosis, four of which had 'fish mouth' deformity.Conclusion Initial results of the new generation PVFD for unruptured intracranial aneurysm treatment demonstrate overall satisfactory device performance, safety profile, and effectiveness.
引用
收藏
页码:e166 / e171
页数:6
相关论文
共 23 条
[1]   Evaluation of Pipeline Flex delivery system for the treatment of unruptured aneurysms [J].
Aguilar-Salinas, Pedro ;
Brasiliense, Leonardo B. C. ;
Gonsales, Douglas ;
Mitchell, Bartley ;
Lima, Andrey ;
Sauvageau, Eric ;
Hanel, Ricardo .
EXPERT REVIEW OF MEDICAL DEVICES, 2016, 13 (10) :885-897
[2]   Pipeline for Uncoilable or Failed Aneurysms: Results from a Multicenter Clinical Trial [J].
Becske, Tibor ;
Kallmes, David F. ;
Saatci, Isil ;
McDougall, Cameron G. ;
Szikora, Istvn ;
Lanzino, Giuseppe ;
Moran, Christopher J. ;
Woo, Henry H. ;
Lopes, Demetrius K. ;
Berez, Aaron L. ;
Cher, Daniel J. ;
Siddiqui, Adnan H. ;
Levy, Elad I. ;
Albuquerque, Felipe C. ;
Fiorella, David J. ;
Berentei, Zsolt ;
Marosfoi, Miklos ;
Cekirge, Saruhan H. ;
Nelson, Peter K. .
RADIOLOGY, 2013, 267 (03) :858-868
[3]   Multicenter Study of Pipeline Flex for Intracranial Aneurysms [J].
Brasiliense, Leonardo B. C. ;
Aguilar-Salinas, Pedro ;
Lopes, Demetrius K. ;
Nogueira, Danilo ;
DeSousa, Keith ;
Nelson, Peter K. ;
Moran, Christopher J. ;
Mazur, Marcus D. ;
Taussky, Philipp ;
Park, Min S. ;
Dabus, Guilherme ;
Linfante, Italo ;
Chaudry, Imran ;
Turner, Ray D. ;
Spiotta, Alex M. ;
Turk, Aquilla S. ;
Siddiqui, Adnan H. ;
Levy, Elad I. ;
Hopkins, L. Nelson ;
Arthur, Adam S. ;
Nickele, Christopher ;
Gonsales, Douglas ;
Sauvageau, Eric ;
Hanel, Ricardo A. .
NEUROSURGERY, 2019, 84 (06) :E402-E409
[4]   Effects of stent sizing on endothelial and vessel wall stress: potential mechanisms for in-stent restenosis [J].
Chen, Henry Y. ;
Hermiller, James ;
Sinha, Anjan K. ;
Sturek, Michael ;
Zhu, Luoding ;
Kassab, Ghassan S. .
JOURNAL OF APPLIED PHYSIOLOGY, 2009, 106 (05) :1686-1691
[5]  
Chia GS., 2022, SVIN, V2, DOI [10.1161/SVIN.121.000292, DOI 10.1161/SVIN.121.000292]
[6]   Review of current intracranial aneurysm flow diversion technology and clinical use [J].
Dandapat, Sudeepta ;
Mendez-Ruiz, Alan ;
Martinez-Galdamez, Mario ;
Macho, Juan ;
Derakhshani, Shahram ;
Foa Torres, Gustavo ;
Pereira, Vitor M. ;
Arat, Anil ;
Wakhloo, Ajay K. ;
Ortega-Gutierrez, Santiago .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2021, 13 (01) :54-+
[7]   Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysms [J].
Fiorella, D. ;
Lylyk, P. ;
Szikora, I. ;
Kelly, M. E. ;
Albuquerque, F. C. ;
McDougall, C. G. ;
Nelson, P. K. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2009, 1 (01) :56-65
[8]  
Girdhar Gaurav, 2019, eNeurologicalSci, V14, P77, DOI 10.1016/j.ensci.2019.01.004
[9]   Thrombogenicity assessment of Pipeline Flex, Pipeline Shield, and FRED flow diverters in an in vitro human blood physiological flow loop model [J].
Girdhar, Gaurav ;
Andersen, Arielle ;
Pangerl, Elizabeth ;
Jahanbekam, Reza ;
Ubl, Samantha ;
Nguyen, Kevin ;
Wainwright, John ;
Wolf, Michael F. .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2018, 106 (12) :3195-3202
[10]   Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results [J].
Hanel, Ricardo A. ;
Kallmes, David F. ;
Lopes, Demetrius Klee ;
Nelson, Peter Kim ;
Siddiqui, Adnan ;
Jabbour, Pascal ;
Pereira, Vitor M. ;
Istvan, Istvan Szikora ;
Zaidat, Osama O. ;
Bettegowda, Chetan ;
Colby, Geoffrey P. ;
Mokin, Maxim ;
Schirmer, Clemens ;
Hellinger, Frank R. ;
Given, Curtis, II ;
Krings, Timo ;
Taussky, Philipp ;
Toth, Gabor ;
Fraser, Justin F. ;
Chen, Michael ;
Priest, Ryan ;
Kan, Peter ;
Fiorella, David ;
Frei, Don ;
Aagaard-Kienitz, Beverly ;
Diaz, Orlando ;
Malek, Adel M. ;
Cawley, C. Michael ;
Puri, Ajit S. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2020, 12 (01) :62-66