Biomechanical design optimization of proximal humerus locked plates: A review

被引:3
|
作者
Zdero, Radovan [1 ]
Brzozowski, Pawel [1 ,3 ]
Schemitsch, Emil H. [1 ,2 ]
机构
[1] Victoria Hosp, Orthopaed Biomech Lab, London, ON, Canada
[2] Western Univ, Div Orthopaed Surg, London, ON, Canada
[3] Victoria Hosp, Room A6-144,800 Commissioners Rd East, London, ON N6A 5W9, Canada
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2024年 / 55卷 / 02期
关键词
Biomechanics; Optimization; Proximal humerus; Locked plates; Review; SURGICAL NECK FRACTURES; LOCKING PLATE; MEDIAL SUPPORT; INTERNAL-FIXATION; STAINLESS-STEEL; CALCAR SCREWS; IN-VITRO; BONE; STABILITY; IMPLANTS;
D O I
10.1016/j.injury.2023.111247
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: Proximal humerus locked plates (PHLPs) are widely used for fracture surgery. Yet, non-union, malunion, infection, avascular necrosis, screw cut-out (i.e., perforation), fixation failure, and re-operation occur. Most biomechanical investigators compare a specific PHLP configuration to other implants like nonlocked plates, nails, wires, and arthroplasties. However, it is unknown whether the PHLP configuration is biomechanically optimal according to some well-known biomechanical criteria. Therefore, this is the first review of the systematic optimization of plate and/or screw design variables for improved PHLP biomechanical performance.Methods: The PubMed website was searched for papers using the terms "proximal humerus" or "shoulder" plus "biomechanics/biomechanical" plus "locked/locking plates". PHLP papers were included if they were (a) optimization studies that systematically varied plate and screw variables to determine their influence on PHLP's biomechanical performance; (b) focused on plate and screw variables rather than augmentation techniques (i.e., extra implants, bone struts, or cement); (c) published after the year 2000 signaling the commercial availability of locked plate technology; and (d) written in English.Results: The 41 eligible papers involved experimental testing and/or finite element modeling. Plate variables investigated by these papers were geometry, material, and/or position, while screw variables studied were number, distribution, angle, size, and/or threads. Numerical outcomes given by these papers included stiffness, strength, fracture motion, bone and implant stress, and/or the number of loading cycles to failure. But, no paper fully optimized any plate or screw variable for a PHLP by simultaneously applying four well-established biomechanical criteria: (a) allow controlled fracture motion for early callus generation; (b) reduce bone and implant stress below the material's ultimate stress to prevent failure; (c) maintain sufficient bone-plate interface stress to reduce bone resorption (i.e., stress shielding); and (d) increase the number of loading cycles before failure for a clinically beneficial lifespan (i.e., fatigue life). Finally, this review made suggestions for future work, identified clinical implications, and assessed the quality of the papers reviewed.Conclusions: Applying biomechanical optimization criteria can assist biomedical engineers in designing or evaluating PHLPs, so orthopaedic surgeons can have superior PHLP constructs for clinical use.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Biomechanical analysis of plate osteosynthesis systems for proximal humerus fractures
    Lever, James P.
    Aksenov, Sergei A.
    Zdero, Rad
    Ahn, Henry
    Mckee, Michael D.
    Schemitsch, Emil H.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (01) : 23 - 29
  • [32] Biomechanical Evaluation of Two-Part Surgical Neck Fractures of the Humerus Fixed by an Angular Stable Locked Intramedullary Nail
    Horn, Joachim
    Gueorguiev, Boyko
    Brianza, Stefano
    Steen, Harald
    Schwieger, Karsten
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (07) : 406 - 413
  • [33] Locked Plate Fixation for Fractures of the Proximal Humerus
    Ricci, William M.
    TECHNIQUES IN ORTHOPAEDICS, 2007, 22 (04) : 192 - 196
  • [34] Role of inferomedial supporting screws for secondary varus deformity in non-osteoporotic proximal humerus fracture: A biomechanical study
    Shin, Myung Jin
    Kim, Hyojune
    Kim, Dong Min
    Park, Dongjun
    Jeon, In-Ho
    Koh, Kyoung Hwan
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2021, 141 (09) : 1517 - 1523
  • [35] Why locking plates for the proximal humerus do not fit well
    Kim, Hyungsuk
    Chung, Yang-Guk
    Jang, Ji Seok
    Kim, Yongdeok
    Park, Soo Bin
    Song, Hyun Seok
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (02) : 219 - 226
  • [36] Biomechanical effects of calcar screws and bone block augmentation on medial support in locked plating of proximal humeral fractures
    Katthagen, Jan Christoph
    Schwarze, Michael
    Meyer-Kobbe, Josefin
    Voigt, Christine
    Hurschler, Christof
    Lill, Helmut
    CLINICAL BIOMECHANICS, 2014, 29 (07) : 735 - 741
  • [37] Internal fixation of proximal humerus fractures. Influence of implant design on biomechanical properties
    Hessmann, MH
    Sternstein, W
    Krummenauer, F
    Hofmann, A
    Rommens, PM
    CHIRURG, 2005, 76 (02): : 167 - 174
  • [38] Retrospective analysis of proximal humeral fracture-dislocations managed with locked plates
    Trikha, Vivek
    Singh, Vivek
    Choudhury, Buddhadeb
    Das, Saubhik
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2017, 26 (10) : E293 - E299
  • [39] Analysis of efficacy and failure in proximal humerus fractures treated with locking plates
    Agudelo, Juan
    Schuermann, Matthias
    Stahel, Philip
    Helwig, Peter
    Morgan, Steven J.
    Zechel, Wolfgang
    Bahrs, Christian
    Parekh, Anand
    Ziran, Bruce
    Williams, Allison
    Smith, Wade
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2007, 21 (10) : 676 - 681
  • [40] A comprehensive update on current fixation options for two-part proximal humerus fractures A biomechanical investigation
    Yoon, Richard S.
    Dziadosz, Daniel
    Porter, David A.
    Frank, Matthew A.
    Smith, Wade R.
    Liporace, Frank A.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 (03): : 510 - 514