Characterization of 250 MeV Protons from the Varian ProBeam PBS System for FLASH Radiation Therapy

被引:8
作者
Charyyev, Serdar [1 ]
Chang, Chih-Wei [2 ,3 ]
Zhu, Mingyao [2 ,3 ]
Lin, Liyong [2 ,3 ]
Langen, Katja [2 ,3 ]
Dhabaan, Anees [2 ,3 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Palo Alto, CA 94304 USA
[2] Emory Univ, Dept Radiat Oncol, Atlanta, GA 30322 USA
[3] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
关键词
FLASH; proton therapy; commissioning; IONIZATION-CHAMBER; IRRADIATION; DELIVERY;
D O I
10.14338/IJPT-22-00027.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Shoot-through proton FLASH radiation therapy has been proposed where the highest energy is extracted from a cyclotron to maximize the dose rate (DR). Although our proton pencil beam scanning system can deliver 250 MeV (the highest energy), this energy is not used clinically, and as such, 250 MeV has yet to be characterized during clinical commissioning. We aim to characterize the 250-MeV proton beam from the Varian ProBeam system for FLASH and assess the usability of the clinical monitoring ionization chamber (MIC) for FLASH use. We measured the following data for beam commissioning: integral depth dose curve, spot sigma, and absolute dose. To evaluate the MIC, we measured output as a function of beam current. To characterize a 250 MeV FLASH beam, we measured (1) the central axis DR as a function of current and spot spacing and arrangement, (2) for a fixed spot spacing, the maximum field size that achieves FLASH DR (ie, > 40 Gy/s), and (3) DR reproducibility. All FLASH DR measurements were performed using an ion chamber for the absolute dose, and irradiation times were obtained from log files. We verified dose measurements using EBT-XD films and irradiation times using a fast, pixelated spectral detector. R90 and R80 from integral depth dose were 37.58 and 37.69 cm, and spot sigma at the isocenter were sigma(x) = 3.336 and sigma(y) = 3.332 mm, respectively. The absolute dose output was measured as 0.343 Gy*mm(2)/MU for the commissioning conditions. Output was stable for beam currents up to 15 nA and gradually increased to 12-fold for 115 nA. Dose and DR depended on beam current, spot spacing, and arrangement and could be reproduced with 6.4% and 4.2% variations, respectively. Although FLASH was achieved and the largest field size that delivers FLASH DR was determined as 35 x 35 mm(2), the current MIC has DR dependence, and users should measure dose and DR independently each time for their FLASH applications.
引用
收藏
页码:279 / 289
页数:11
相关论文
共 44 条
  • [21] Comparison of multi-institutional Varian ProBeam pencil beam scanning proton beam commissioning data
    Langner, Ulrich W.
    Eley, John G.
    Dong, Lei
    Langen, Katja
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (03): : 96 - 107
  • [22] A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles
    Lin, Liyong
    Ainsley, Christopher G.
    Mertens, Thierry
    De Wilde, Olivier
    Talla, Patrick T.
    McDonough, James E.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (12) : N171 - N180
  • [23] Ion recombination and polarity correction factors for a plane-parallel ionization chamber in a proton scanning beam
    Liszka, Malgorzata
    Stolarczyk, Liliana
    Klodowska, Magdalena
    Kozera, Anna
    Krzempek, Dawid
    Mojzeszek, Natalia
    Pezdracka, Anna
    Waligorski, Michael Patrick Russell
    Olko, Pawel
    [J]. MEDICAL PHYSICS, 2018, 45 (01) : 391 - 401
  • [24] Delivery of Ultra-Rapid Flash Radiation Therapy and Demonstration of Normal Tissue Sparing After Abdominal Irradiation of Mice
    Loo, Billy W.
    Schuler, Emil
    Lartey, Frederick M.
    Rafat, Marjan
    King, Gregory J.
    Trovati, Stefania
    Koong, Albert C.
    Maxim, Peter G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 98 (02): : E16 - E16
  • [25] Martin J., 1999, ESTAR, PSTAR, and ASTAR: computer programs for calculating stopping-power and range tables for electrons, protons, and helium ions (version 1.21)
  • [26] Proton FLASH Radiotherapy for the Treatment of Symptomatic Bone Metastases The FAST-01 Nonrandomized Trial
    Mascia, Anthony E.
    Daugherty, Emily C.
    Zhang, Yongbin
    Lee, Eunsin
    Xiao, Zhiyan
    Sertorio, Mathieu
    Woo, Jennifer
    Backus, Lori R.
    McDonald, Julie M.
    McCann, Claire
    Russell, Kenneth
    Levine, Lisa
    Sharma, Ricky A.
    Khuntia, Dee
    Bradley, Jeffrey D.
    Simone, Charles B.
    Perentesis, John P.
    Breneman, John C.
    [J]. JAMA ONCOLOGY, 2023, 9 (01) : 62 - 69
  • [27] Long-term neurocognitive benefits of FLASH radiotherapy driven by reduced reactive oxygen species
    Montay-Gruel, Pierre
    Acharya, Munjal M.
    Petersson, Kristoffer
    Alikhani, Leila
    Yakkala, Chakradhar
    Allen, Barrett D.
    Ollivier, Jonathan
    Petit, Benoit
    Jorge, Patrik Goncalves
    Syage, Amber R.
    Nguyen, Thuan A.
    Baddour, Al Anoud D.
    Lu, Celine
    Singh, Paramvir
    Moeckli, Raphael
    Bochud, Francois
    Germond, Jean-Francois
    Froidevaux, Pascal
    Bailat, Claude
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Limoli, Charles L.
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2019, 116 (22) : 10943 - 10951
  • [28] X-rays can trigger the FLASH effect: Ultra-high dose-rate synchrotron light source prevents normal brain injury after whole brain irradiation in mice
    Montay-Gruel, Pierre
    Bouchet, Audrey
    Jaccard, Maud
    Patin, David
    Serduc, Raphael
    Aim, Warren
    Petersson, Kristoffer
    Petit, Benoit
    Bailat, Claude
    Bourhis, Jean
    Brauer-Krisch, Elke
    Vozenin, Marie-Catherine
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 129 (03) : 582 - 588
  • [29] Irradiation in a flash: Unique sparing of memory in mice after whole brain irradiation with dose rates above 100 Gy/s
    Montay-Gruel, Pierre
    Petersson, Kristoffer
    Jaccard, Maud
    Boivin, Gael
    Germond, Jean-Francois
    Petit, Benoit
    Doenlen, Raphael
    Favaudon, Vincent
    Bochud, Francois
    Bailat, Claude
    Bourhis, Jean
    Vozenin, Marie-Catherine
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 124 (03) : 365 - 369
  • [30] Paganetti H., 2019, Series in medical physics and biomedical engineering, V2nd ed.