Characterization of 250 MeV Protons from the Varian ProBeam PBS System for FLASH Radiation Therapy

被引:8
作者
Charyyev, Serdar [1 ]
Chang, Chih-Wei [2 ,3 ]
Zhu, Mingyao [2 ,3 ]
Lin, Liyong [2 ,3 ]
Langen, Katja [2 ,3 ]
Dhabaan, Anees [2 ,3 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Palo Alto, CA 94304 USA
[2] Emory Univ, Dept Radiat Oncol, Atlanta, GA 30322 USA
[3] Emory Univ, Winship Canc Inst, Atlanta, GA 30322 USA
关键词
FLASH; proton therapy; commissioning; IONIZATION-CHAMBER; IRRADIATION; DELIVERY;
D O I
10.14338/IJPT-22-00027.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Shoot-through proton FLASH radiation therapy has been proposed where the highest energy is extracted from a cyclotron to maximize the dose rate (DR). Although our proton pencil beam scanning system can deliver 250 MeV (the highest energy), this energy is not used clinically, and as such, 250 MeV has yet to be characterized during clinical commissioning. We aim to characterize the 250-MeV proton beam from the Varian ProBeam system for FLASH and assess the usability of the clinical monitoring ionization chamber (MIC) for FLASH use. We measured the following data for beam commissioning: integral depth dose curve, spot sigma, and absolute dose. To evaluate the MIC, we measured output as a function of beam current. To characterize a 250 MeV FLASH beam, we measured (1) the central axis DR as a function of current and spot spacing and arrangement, (2) for a fixed spot spacing, the maximum field size that achieves FLASH DR (ie, > 40 Gy/s), and (3) DR reproducibility. All FLASH DR measurements were performed using an ion chamber for the absolute dose, and irradiation times were obtained from log files. We verified dose measurements using EBT-XD films and irradiation times using a fast, pixelated spectral detector. R90 and R80 from integral depth dose were 37.58 and 37.69 cm, and spot sigma at the isocenter were sigma(x) = 3.336 and sigma(y) = 3.332 mm, respectively. The absolute dose output was measured as 0.343 Gy*mm(2)/MU for the commissioning conditions. Output was stable for beam currents up to 15 nA and gradually increased to 12-fold for 115 nA. Dose and DR depended on beam current, spot spacing, and arrangement and could be reproduced with 6.4% and 4.2% variations, respectively. Although FLASH was achieved and the largest field size that delivers FLASH DR was determined as 35 x 35 mm(2), the current MIC has DR dependence, and users should measure dose and DR independently each time for their FLASH applications.
引用
收藏
页码:279 / 289
页数:11
相关论文
共 44 条
  • [1] [Anonymous], 2021, RayStation 10B Reference Manual
  • [2] Treatment of a first patient with FLASH-radiotherapy
    Bourhis, Jean
    Sozzi, Wendy Jeanneret
    Jorge, Patrik Goncalves
    Gaide, Olivier
    Bailat, Claude
    Duclos, Frederic
    Patin, David
    Ozsahin, Mahmut
    Bochud, Francois
    Germond, Jean-Francois
    Moeckli, Raphael
    Vozenin, Marie-Catherine
    [J]. RADIOTHERAPY AND ONCOLOGY, 2019, 139 : 18 - 22
  • [3] Validation of a deep learning-based material estimation model for Monte Carlo dose calculation in proton therapy
    Chang, Chih-Wei
    Zhou, Shuang
    Gao, Yuan
    Lin, Liyong
    Liu, Tian
    Bradley, Jeffrey D.
    Zhang, Tiezhi
    Zhou, Jun
    Yang, Xiaofeng
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (21)
  • [4] A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems
    Chang, Chih-Wei
    Huang, Sheng
    Harms, Joseph
    Zhou, Jun
    Zhang, Rongxiao
    Dhabaan, Anees
    Slopsema, Roelf
    Kang, Minglei
    Liu, Tian
    McDonald, Mark
    Langen, Katja
    Lin, Liyong
    [J]. MEDICAL PHYSICS, 2020, 47 (04) : 1545 - 1557
  • [5] Charyyev S, 2022, Arxiv, DOI arXiv:2209.01178
  • [6] A novel proton counting detector and method for the validation of tissue and implant material maps for Monte Carlo dose calculation
    Charyyev, Serdar
    Chang, Chih-Wei
    Harms, Joseph Mark
    Oancea, Cristina
    Yoon, S. Tim
    Yang, Xiaofeng
    Zhang, Tiezhi
    Zhou, Jun
    Lin, Liyong
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (04)
  • [7] Optimization of hexagonal-pattern minibeams for spatially fractionated radiotherapy using proton beam scanning
    Charyyev, Serdar
    Artz, Mark
    Szalkowski, Gregory
    Chang Chih-Wei
    Stanforth, Alexander
    Lin Liyong
    Zhang Rongxiao
    Wang, C. -K. Chris
    [J]. MEDICAL PHYSICS, 2020, 47 (08) : 3485 - 3495
  • [8] FLASH Proton Pencil Beam Scanning Irradiation Minimizes Radiation-Induced Leg Contracture and Skin Toxicity in Mice
    Cunningham, Shannon
    McCauley, Shelby
    Vairamani, Kanimozhi
    Speth, Joseph
    Girdhani, Swati
    Abel, Eric
    Sharma, Ricky A.
    Perentesis, John P.
    Wells, Susanne I.
    Mascia, Anthony
    Sertorio, Mathieu
    [J]. CANCERS, 2021, 13 (05) : 1 - 15
  • [9] Spread-out Bragg peak proton FLASH irradiation using a clinical synchrocyclotron: Proof of concept and ion chamber characterization
    Darafsheh, Arash
    Hao, Yao
    Zhao, Xiandong
    Zwart, Townsend
    Wagner, Miles
    Evans, Tucker
    Reynoso, Francisco
    Zhao, Tianyu
    [J]. MEDICAL PHYSICS, 2021, : 4472 - 4484
  • [10] FLASH Radiotherapy for the Treatment of Symptomatic Bone Metastases (FAST-01): Protocol for the First Prospective Feasibility Study
    Daugherty, Emily C.
    Mascia, Anthony
    Zhang, Yong
    Lee, Eunsin
    Xiao, Zhiyan
    Sertorio, Mathieu
    Woo, Jennifer
    McCann, Claire
    Russell, Kenneth
    Levine, Lisa
    Sharma, Ricky
    Khuntia, Deepak
    Bradley, Jeffrey
    Simone, Charles B.
    Perentesis, John
    Breneman, John
    [J]. JMIR RESEARCH PROTOCOLS, 2023, 12 (01):