Evaluation of the Alinity hq Automated Hematologic Analyzer and Determination of Reference Ranges for Novel Red Blood Cell Parameters

被引:0
作者
Kim, Seon Young [1 ,2 ]
Kim, Jimyung [1 ]
Kwon, Gye Cheol [1 ]
机构
[1] Chungnam Natl Univ, Dept Lab Med, Coll Med, Chungnam Natl Univ Hosp, Daejeon, South Korea
[2] Chungnam Natl Univ, Coll Med, Canc Res Inst, Daejeon, South Korea
基金
新加坡国家研究基金会;
关键词
automated hematology analyzer; complete blood count; Alinity hq; performance evaluation; red blood cell parameters; PERFORMANCE EVALUATION;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objective. We evaluated the performance of the Alinity hq automated analyzer (Abbott Laboratories, Diagnostics Division, Hematology, Santa Clara, CA, USA). In addition, we determined the reference ranges for the red blood cell (RBC) research parameters. Method. The precision and stability of the instrument were measured for all complete blood count (CBC) parameters. We compared the CBC results between the Alinity hq and the DxH800 (Beckman Coulter, Miami, FL, USA) and the ADVIA 2120 (Siemens Healthcare Diagnostics, Tarrytown, NY, USA). The white blood cell (WBC) differential results were verified by manual differential counts. We determined the reference ranges of RBC research parameters among healthy adults. Results. The Alinity hq analyzer demonstrated good within-run and between-day precision for all CBC parameters. The calculated correlation coefficients (r) indicated that Alinity hq-determined values of WBC, RBC, platelet (PLT) counts, hemoglobin (HGB), hematocrit (HCT), and mean corpuscular volume (MCV) were in very good concordance (r>0.95) when compared with results from the DxH800 and the ADVIA 2120. The Alinity hq WBC differential counts were comparable with the manual differential counts, and the results of neutrophil counts by Alinity hq correlated well. Lymphocyte and monocyte count correlated well in samples without blasts. Conclusions. The Alinity hq presented good analytical performance and showed good correlation compared with other hematology analyzers and manual differential counts.
引用
收藏
页码:653 / 660
页数:8
相关论文
共 14 条
[1]   Effects of specimen haemolysis on complete blood count results by Abbott Alinity hq System [J].
Ercan, Mujgan ;
Akbulut, Emis Deniz ;
Bayraktar, Nihayet ;
Ercan, Serif .
BIOCHEMIA MEDICA, 2021, 31 (03)
[2]   Accurate white blood cell differential by Alinity hq: A comparison with flow cytometry and manual differential [J].
Gambell, Peter ;
Rowley, Grant ;
Pham, Tracy Anh Thu ;
Dang, Theresa Loan ;
Mulumba, Huguette ;
Smith, Larry ;
Lakos, Gabriella .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (02) :288-295
[3]   Complete blood count and cell population data parameters from the Abbott Alinity hq analyzer are useful in differentiating myelodysplastic syndromes from other forms of cytopenia [J].
Hwang, Sang Mee ;
Nam, Youngwon .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (03) :468-476
[4]  
Koepke JA, 2007, H20-A2: Reference Leukocyte (WBC) Differential Count (proportional) and Evaluation of Instrumental Methods: Approved Standard
[5]   Alinity hq platelet count s not impacted by severe microcytosis [J].
Lakos, Gabriella ;
Mukhtar, Zainab ;
Masi, Loredana ;
Valente, Sabatino ;
Papa, Fabrizio .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2022, 36 (04)
[6]   Alinity hq platelet results are equivalent with the international reference method in thrombocytopenic samples [J].
Lifson, Mark A. ;
Wakui, Masatoshi ;
Arai, Tomoko ;
Mitsuhashi, Takayuki ;
Lakos, Gabriella ;
Murata, Mitsuru .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (06) :1357-1362
[7]   Clinical utility of Abbott Alinity hq extended red blood cell parameters in differentiating β-thalassemia trait and iron deficiency anemia [J].
Mukhtar, Zainab ;
Valente, Sabatino ;
Masi, Loredana ;
Lakos, Gabriella ;
Papa, Fabrizio .
INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (06) :1635-1643
[8]   The Association between Parameters of Erythrocytes Morphology and Thrombophilia-related Mutations [J].
Oz, Ozlem ;
Gonel, Ataman .
REVIEWS ON RECENT CLINICAL TRIALS, 2022, 17 (01) :40-45
[9]   Method comparison between Alinity hq and Sysmex XT-4000i in an emergency laboratory [J].
Preda, Cristina E. ;
Oprea, Oana R. ;
Manescu, Ion B. ;
Dobreanu, Minodora .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2023, 83 (01) :31-37
[10]  
Rabinovitch A, 2010, H26-A2: Validation, Verification, and Quality Assurance of Automated Hematology Analyzers: Approved Standard