Is the Blood Pressure-Enabled Smartwatch Ready to Drive Precision Medicine? Supporting Findings From a Validation Study

被引:2
作者
Lee, Wan Ling [1 ]
Danaee, Mahmoud [2 ]
Abdullah, Adina [3 ]
Wong, Li Ping [2 ]
机构
[1] Univ Malaya, Fac Med, Dept Nursing Sci, Kuala Lumpur 50603, Malaysia
[2] Univ Malaya, Fac Med, Dept Social & Prevent Med, Kuala Lumpur 50603, Malaysia
[3] Univ Ma laya, Fac Med, Dept Primary Care Med, Kuala Lumpur 50603, Malaysia
关键词
mHealth; Wearables; Smartwatch; Blood pressure; Validation; Precision medicine; Precision health; HYPERTENSION/INTERNATIONAL ORGANIZATION; INSTRUMENTATION/EUROPEAN SOCIETY; UNIVERSAL STANDARD; MEASURING DEVICES; ADVANCEMENT; ASSOCIATION; POPULATION; ADHERENCE; RISK;
D O I
10.14740/cr1569
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The popular wrist-worn wearables recording a variety of health metrics such as blood pressure (BP) in real time could play a potential role to advance precision medicine, but these devices are often insufficiently validated for their performance to enhance confidence in its use across diverse populations. The accuracy of BP-enabled smartwatch is assessed among the multi-ethnic Malaysians, and findings is discussed in comparison with conventional automated upper-arm BP device.Methods: Validation procedures followed the guidelines by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). Quota sampling was employed to obtain eligible patients with normal and abnormal BP as per guideline. The measurements of BP were taken at wrist using HUAWEI WATCH D (test BP); and the readings were assessed against reference BP by the mercury sphygmomanometer. Agreement statistics and linear regression analyses were performed.Results: BP measurements (234 data pairs) from 78 patients that fulfilled AAMI/ESH/ISO protocol were analyzed. The BP readings taken by the HUAWEI WATCH D were comparable to reference BP by sphygmomanometer based on 1) Criterion 1: systolic blood pressure (SBP) =-0.034 (SD 5.24) and diastolic blood pressure (DBP) =-0.65 (SD 4.66) mm Hg; and 2) Criterion 2: SBPs =-0.034 (SD 4.18) and DBPs =-0.65 (SD 3.94) mm Hg. Factors of sociodemographic characteristics, anthropometric measurements, cardiovascular comorbidities, and wrist hair density were not significantly associated with the mean BP differences.Conclusions: HUAWEI WATCH D fulfilled criteria 1 and 2 of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) guidelines. It can be recommended for clinical use across a wider population. The rich data from real-time BP measurements in concurrent with other health-related parameters recorded by the smartwatch wearable offer opportunities to drive precision medicine in tackling therapeutic inertia by personalizing BP control regimen.
引用
收藏
页码:437 / 445
页数:9
相关论文
共 28 条
  • [1] Role of Home Blood Pressure Monitoring in Overcoming Therapeutic Inertia and Improving Hypertension Control A Systematic Review and Meta-Analysis
    Agarwal, Rajiv
    Bills, Jennifer E.
    Hecht, Tyler J. W.
    Light, Robert P.
    [J]. HYPERTENSION, 2011, 57 (01) : 29 - U139
  • [2] Monte Carlo analysis of optical heart rate sensors in commercial wearables: the effect of skin tone and obesity on the photoplethysmography (PPG) signal
    Ajmal
    Boonya-Ananta, Tananant
    Rodriguez, Andres J.
    Du Le, V. N.
    Ramella-Roman, Jessica C.
    [J]. BIOMEDICAL OPTICS EXPRESS, 2021, 12 (12) : 7445 - 7457
  • [3] Impact of Self-Monitoring of Blood Pressure on Processes of Hypertension Care and Long-Term Blood Pressure Control
    Bryant, Kelsey B.
    Sheppard, James P.
    Ruiz-Negron, Natalia
    Kronish, Ian M.
    Fontil, Valy
    King, Jordan B.
    Pletcher, Mark J.
    Bibbins-Domingo, Kirsten
    Moran, Andrew E.
    McManus, Richard J.
    Bellows, Brandon K.
    [J]. JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2020, 9 (15):
  • [4] Wrist-worn and chest-strap wearable devices: Systematic review on accuracy and metrological characteristics
    Cosoli, Gloria
    Spinsante, Susanna
    Scalise, Lorenzo
    [J]. MEASUREMENT, 2020, 159
  • [5] Effectiveness of home blood pressure telemonitoring: a systematic review and meta-analysis of randomised controlled studies
    Duan, Y.
    Xie, Z.
    Dong, F.
    Wu, Z.
    Lin, Z.
    Sun, N.
    Xu, J.
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (07) : 427 - 437
  • [6] Blood pressure measurement using only a smartphone
    Frey, Lorenz
    Menon, Carlo
    Elgendi, Mohamed
    [J]. NPJ DIGITAL MEDICINE, 2022, 5 (01)
  • [7] Institute for Public Health, 2020, National Health and Morbidity Survey (NHMS) 2019: vol. I: non-communicable diseases: risk factors and other health problems
  • [8] Sources of inaccuracy in themeasurement of adult patients' resting blood pressure in clinical settings: a systematic review
    Kallioinen, Noa
    Hill, Andrew
    Horswill, Mark S.
    Ward, Helen E.
    Watson, Marcus O.
    [J]. JOURNAL OF HYPERTENSION, 2017, 35 (03) : 421 - 441
  • [9] Validation of a wrist-type home nocturnal blood pressure monitor in the sitting and supine position according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-9600T
    Kuwabara, Mitsuo
    Harada, Kanako
    Hishiki, Yukiko
    Kario, Kazuomi
    [J]. JOURNAL OF CLINICAL HYPERTENSION, 2019, 21 (04) : 463 - 469
  • [10] Smartphone-Based Blood Pressure Measurement Using Transdermal Optical Imaging Technology
    Luo, Hong
    Yang, Deye
    Barszczyk, Andrew
    Vempala, Naresh
    Wei, Jing
    Wu, Si Jia
    Zheng, Paul Pu
    Fu, Genyue
    Lee, Kang
    Feng, Zhong-Ping
    [J]. CIRCULATION-CARDIOVASCULAR IMAGING, 2019, 12 (08)