Standardized approach for accurate and reliable model development of ion-exchange chromatography based on parameter-by-parameter method and consideration of extra-column effects

被引:7
|
作者
Chen, Yu-Cheng [1 ]
Lu, Hui-Li [2 ]
Wang, Rong-Zhu [2 ]
Sun, Guo [2 ]
Zhang, Xue-Qin [2 ]
Liang, Jing-Qi [2 ]
Jungbauer, Alois [3 ]
Yao, Shan-Jing [1 ]
Lin, Dong-Qiang [1 ,4 ]
机构
[1] Zhejiang Univ, Coll Chem & Biol Engn, Key Lab Biomass Chem Engn, Zhejiang Key Lab Smart Biomat,Minist Educ, Hangzhou, Peoples R China
[2] Gmax Biopharm LLC, Hangzhou, Zhejiang, Peoples R China
[3] Univ Nat Resources & Life Sci, Dept Biotechnol, Vienna, Austria
[4] Zhejiang Univ, Coll Chem & Biol Engn, Hangzhou 310058, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
antibody charge variant; extra-column effect; ion-exchange chromatography; model development; steric mass action model; OPERATING-CONDITIONS; ELUTION CURVES; PREDICTION; PROTEINS;
D O I
10.1002/biot.202300687
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Developing an accurate and reliable model for chromatographic separation that meets regulatory requirements and ensures consistency in model development remains challenging. In order to address this challenge, a standardized approach was proposed in this study with ion-exchange chromatography (IEC). The approach includes the following steps: liquid flow identification, system and column-specific parameters determination and validation, multi-component system identification, protein amount validation, steric mass action parameters determination and evaluation, and validation of the calibrated model's generalization ability. The parameter-by-parameter (PbP) calibration method and the consideration of extra-column effects were integrated to enhance the accuracy of the developed models. The experiments designed for implementing the PbP method (five gradient experiments for model calibration and one stepwise experiment for model validation) not only streamline the experimental workload but also ensure the extrapolation abilities of the model. The effectiveness of the standardized approach is successfully validated through an application about the IEC separation of industrial antibody variants, and satisfactory results were observed with R2 approximate to 0.9 for the majority of calibration and validation experiments. The standardized approach proposed in this work contributes significantly to improve the accuracy and reliability of the developed IEC models. Models developed using this standardized approach are ready to be applied to a broader range of industrial separation systems, and are likely find further applications in model-assisted decision-making of process development. Developing an accurate and reliable model for chromatographic separation that meets regulatory requirements and ensures consistency in model development remains challenging. To address this challenge, a standardized approach was proposed in this study, which was based on the parameter-by-parameter calibration method and the consideration of extra-column effects. The proposed method was successfully validated through an application about the ion-exchange chromatography separation of antibody charge variants. The satisfactory results were observed with R2 approximate to 0.9 for most of the calibration and validation experiments. image
引用
收藏
页数:15
相关论文
共 2 条
  • [1] Parameter-by-parameter method for steric mass action model of ion exchange chromatography: Theoretical considerations and experimental verification
    Chen, Yu-Cheng
    Yao, Shan-Jing
    Lin, Dong-Qiang
    JOURNAL OF CHROMATOGRAPHY A, 2022, 1680
  • [2] Parameter-by-parameter method for steric mass action model of ion exchange chromatography: Simplified estimation for steric shielding factor
    Chen, Yu-Cheng
    Yao, Shan-Jing
    Lin, Dong-Qiang
    JOURNAL OF CHROMATOGRAPHY A, 2023, 1687