Radio-luminescent imaging for rapid, high-resolution eye plaque loading verification

被引:0
作者
Yan, Huagang [1 ]
De Jean, Paul [2 ]
Grafil, Elliot [2 ]
Ashraf, Ramish [3 ]
Niedermayr, Thomas [3 ]
Astrahan, Melvin [4 ]
Mruthyunjaya, Prithvi [5 ]
Beadle, Beth [3 ]
Xing, Lei [3 ]
Liu, Wu [3 ]
机构
[1] Capital Med Univ, Sch Biomed Engn, Beijing, Peoples R China
[2] Luca Med Syst Inc, Palo Alto, CA USA
[3] Stanford Univ, Sch Med, Dept Radiat Oncol, 875 Blake Wilbur Dr,MC 5847, Stanford, CA 94305 USA
[4] Eye Plaque LLC, Los Alamitos, CA USA
[5] Stanford Univ, Dept Ophthalmol, Sch Med, Stanford, CA 94305 USA
关键词
eye plaque brachytherapy; gamma evaluation; Monte Carlo simulation; quality assurance; radioluminescent imaging; I-125; PLAQUE; DOSIMETRY; AAPM; SIMULATION; TRANSPORT; MELANOMA; KERMA; FILM;
D O I
10.1002/mp.16003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Eye plaque brachytherapy is currently an optimal therapy for intraocular cancers. Due to the lack of an effective and practical technique to measure the seed radioactivity distribution, current quality assurance (QA) practice according to the American Association of Physicists in Medicine TG129 only stipulates that the plaque assembly be visually inspected. Consequently, uniform seed activity is routinely adopted to avoid possible loading mistakes of differential seed loading. However, modulated dose delivery, which represents a general trend in radiotherapy to provide more personalized treatment for a given tumor and patient, requires differential activities in the loaded seeds. Purpose In this study, a fast and low-cost radio-luminescent imaging and dose calculating system to verify the seed activity distribution for differential loading was developed. Methods A proof-of-concept system consisting of a thin scintillator sheet coupled to a camera/lens system was constructed. A seed-loaded plaque can be placed directly on the scintillator surface with the radioactive seeds facing the scintillator. The camera system collects the radioluminescent signal generated by the scintillator on its opposite side. The predicted dose distribution in the scintillator's sensitive layer was calculated using a Monte Carlo simulation with the planned plaque loading pattern of I-125 seeds. Quantitative comparisons of the distribution of relative measured signal intensity and that of the relative predicted dose in the sensitive layer were performed by gamma analysis, similar to intensity-modulated radiation therapy QA. Results Data analyses showed high gamma (3%/0.3 mm, global, 20% threshold) passing rates for correct seed loadings and low passing rates with distinguished high gamma value area for incorrect loadings, indicating that possible errors may be detected. The measurement and analysis only required a few extra minutes, significantly shorter than the time to assay the extra verification seeds the physicist already must perform as recommended by TG129. Conclusions Radio-luminescent QA can be used to facilitate and assure the implementation of intensity-modulated, customized plaque loading.
引用
收藏
页码:142 / 151
页数:10
相关论文
共 39 条
[31]   Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array [J].
Sharma, Dayananda Shamurailatpam ;
Mhatre, Vaibav ;
Heigrujam, Malhotra ;
Talapatra, Kaustav ;
Mallik, Suman .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (04) :238-248
[32]   Incorporating patient-specific CT-based ophthalmic anatomy in modeling iodine-125 eye plaque brachytherapy dose distributions [J].
Tien, Christopher J. ;
Astrahan, Melvin A. ;
Kim, Jenna M. ;
Materin, Miguel ;
Chen, Zhe ;
Nath, Ravinder ;
Liu, Wu .
BRACHYTHERAPY, 2017, 16 (05) :1057-1064
[33]   Dosimetry verification in eye brachytherapy using silicon pixelated detectors [J].
Weaver, M. ;
Petasecca, M. ;
Lerch, M. L. F. ;
Cutajar, D. ;
Jakubek, J. ;
Pospisil, S. ;
Rosenfeld, A. B. .
RADIATION MEASUREMENTS, 2011, 46 (12) :2010-2013
[34]   Three-dimensional dosimetry imaging of I-125 plaque for eye cancer treatment [J].
Weaver, M. ;
Green, J. ;
Petasecca, M. ;
Lerch, M. L. F. ;
Cutajar, D. ;
Franklin, D. ;
Jakubek, J. ;
Carolan, M. G. ;
Conway, M. ;
Pospisil, S. ;
Kron, T. ;
Metcalfe, P. ;
Zaider, M. ;
Rosenfeld, A. B. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 633 :S276-S278
[35]   An improved treatment planning and quality assurance process for Collaborative Ocular Melanoma Study eye plaque brachytherapy [J].
Weersink, R. A. ;
Patterson, S. ;
Ballantyne, H. ;
Di Tomasso, A. ;
Borg, J. ;
Vitkin, A. ;
Rink, A. ;
Beiki-Ardakani, A. .
BRACHYTHERAPY, 2019, 18 (05) :658-667
[36]   MONTE-CARLO EVALUATION OF KERMA AT A POINT FOR PHOTON TRANSPORT PROBLEMS [J].
WILLIAMSON, JF .
MEDICAL PHYSICS, 1987, 14 (04) :567-576
[37]  
Yan H, 2019, MED PHYS, V46, pE114
[38]   Monte Carlo dosimetry modeling of focused kV x-ray radiotherapy of eye diseases with potential nanoparticle dose enhancement [J].
Yan, Huagang ;
Ma, Xiangyu ;
Sun, Weiyuan ;
Mendez, Stacy ;
Stryker, Stefan ;
Starr-Baier, Sean ;
Delliturri, Gianpiero ;
Zhu, Dengsong ;
Nath, Ravinder ;
Chen, Zhe ;
Roberts, Kenneth ;
MacDonald, Carolyn A. ;
Liu, Wu .
MEDICAL PHYSICS, 2018, 45 (10) :4720-4733
[39]   Method for verifying the air kerma strength of I-125 plaques for the treatment of ocular melanoma [J].
Zimmermann, L. W. ;
Wilkinson, D. Allan .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (04) :297-305