The Evolution of Microwave-Induced Thermoacoustic Signals Generated During Pulsed Microwave Ablation in Bovine Liver

被引:3
|
作者
Evans, Audrey L. [1 ]
Liu, Ruixi L. [1 ]
Ma, Chu [1 ]
Hagness, Susan C. [1 ]
机构
[1] Univ Wisconsin, Dept Elect & Comp Engn, Madison, WI 53706 USA
来源
IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY | 2023年 / 7卷 / 03期
基金
美国国家科学基金会;
关键词
Electromagnetic heating; Liver; Microwave antennas; Microwave integrated circuits; Microwave FET integrated circuits; Ultrasonic imaging; Transducers; Microwave ablation; microwave-induced thermoacoustic signals; thermal therapy; PURE WATER; FREQUENCY;
D O I
10.1109/JERM.2023.3268553
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The temperature dependence of microwave-induced thermoacoustic signals generated in tissue may be exploited to monitor microwave ablation in real-time. We present an experimental study investigating the evolution of microwave-induced thermoacoustic signals that are generated within an ablation zone during microwave ablation in bovine liver tissue. An X-band interstitial coaxial ablation antenna is used to simultaneously heat liver tissue to temperatures up to 90 ? and excite thermoacoustic signals via the absorption of pulsed microwave energy. Thermoacoustic signals are detected using a single-element ultrasound transducer located at the surface of the tissue. Both fresh and boiled liver tissue samples are used in experiments to decouple the influence of temperature and tissue coagulation on thermoacoustic signal characteristics. We identify two thermoacoustic signal characteristics of interest: arrival time of the pulse at the ultrasound receiver and the energy in the pulse. We find that the time difference of arrival over the course of microwave ablation grows in magnitude due to temperature-dependent speed of sound and tissue shrinkage. Thermoacoustic signal energy generally increases throughout microwave ablation, implying an increasing temperature-dependent thermal expansion coefficient of liver tissue. Of the two characteristics, time difference of arrival shows the most promise as a trackable feature for monitoring microwave ablation in real time.
引用
收藏
页码:273 / 280
页数:8
相关论文
共 49 条
  • [1] Microwave-induced thermoacoustic signal characteristics in a dynamic temperature environment
    Evans, Audrey L.
    Ma, Chu
    Hagness, Susan C.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2022, 8 (03)
  • [2] Microwave-Induced Thermoacoustic Tomography for Subcutaneous Vascular Imaging
    Aliroteh, Miaad
    Nan, Hao
    Arbabian, Amin
    2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [3] Towards real-time monitoring of pulsed microwave ablation using thermoacoustic signals: A study of signal characteristics as a function of salinity
    Evans, Audrey L.
    Ma, Chu
    Hagness, Susan C.
    2022 16TH EUROPEAN CONFERENCE ON ANTENNAS AND PROPAGATION (EUCAP), 2022,
  • [4] Performance Evaluation of Focused Microwave Brain Hyperthermia Guided by Microwave-Induced Thermoacoustic Tomography
    Xi, Zijun
    Wang, Xuanyu
    Ye, Ke
    Wang, Xiong
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2023, 7 (04): : 383 - 391
  • [5] Temperature monitoring utilising thermoacoustic signals during pulsed microwave thermotherapy: A feasibility study
    Lou, Cunguang
    Xing, Da
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2010, 26 (04) : 338 - 346
  • [6] Measurement of Intrahepatic Pressure during Microwave Ablation in an Ex Vivo Bovine Liver Model
    Kim, Hae Jin
    Rhim, Hyunchul
    Lee, Min Woo
    Jeong, Woo Kyoung
    GUT AND LIVER, 2015, 9 (06) : 784 - 790
  • [7] Microwave-Induced Thermoacoustic Imaging of Small Animals Applying Scanning Orthogonal Polarization Excitation
    Wang, Baosheng
    Xiong, Naping
    Sun, Yifei
    Zhang, Lejia
    Li, Chenzhe
    Li, Jianian
    Wang, Zhicheng
    Chen, Ziling
    Zhang, Yifeng
    Wang, Xiong
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2022, 6 (01): : 123 - 130
  • [8] A Microwave-Induced Thermoacoustic Imaging System With Non-Contact Ultrasound Detection
    Singhvi, Ajay
    Boyle, Kevin C.
    Fallahpour, Mojtaba
    Khuri-Yakub, Butrus T.
    Arbabian, Amin
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2019, 66 (10) : 1587 - 1599
  • [9] Microwave-Induced Thermoacoustic Imaging of Subcutaneous Vasculature With Near-Field RF Excitation
    Aliroteh, Miaad S.
    Arbabian, Amin
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2018, 66 (01) : 577 - 588
  • [10] Continuous versus pulsed microwave ablation in the liver: any difference in intraoperative pain scores?
    Filippiadis, Dimitrios
    Mazioti, Argyro
    Velonakis, George
    Tsochantzis, Athanasios
    Tosoratti, Nevio
    Kelekis, Alexis
    Kelekis, Nikolaos
    ANNALS OF GASTROENTEROLOGY, 2021, 34 (01): : 80 - 84